SCIENTIFIC 


Operating Instructions 
Thermo Scientific Barnstead GenPure 
Water Purification System 
with xCAD 





50131281 Stand: 11.12 Rights to technical changes reserved 





Thermo Thermo Scientific Barnstead GenPure with xCAD 
te ere Ultrapure water system 


EC-Declaration of Conformity 
in accordance with the EEC machine directive 2006/42/EC, appendix Il A 


We hereby certify that the following described machine in its conception and form put by us into 
circulation is in accordance with all the relevant essential health and safety requirements of the 
EC machinery directive 2006/42/EC as amended and the national laws and regulations adopt- 
ing this directive. 
This declaration is no longer valid if the machine is modified without our consent. 


Manufacturer: Thermo Electron LED GmbH 
Robert-Bosch-Strake 1 
D-63505 Langenselbold 


Description of the machine: 


function: Ultrapure water system 


type: GenPure with xCAD, GenPure UF with xCAD, GenPure UV with xCAD, 
GenPure UV/UF with xCAD, GenPure UV-TOC with xCAD, GenPure UV- 
TOC/UF with xCAD 


article number: bench version: 50131286, 50131250, 50131254, 50131252, 50131296, 
50131298 

wall version: 50131300, 50131302, 50131317, 50131315, 50131321, 
50131323 


The agreement with further valid guidelines/regulations following for the product is ex- 
plained: 


EMC Directive (2004/108/EC) 


Reference to the harmonised standards: 


DIN EN ISO 12100-1 Safety of machinery, Part 1: Basic terminology 
DIN EN ISO 12100-2 Safety of machinery, Part 2: Technical principles 
DIN EN ISO 14121-1 Safety of machinery, Part 1: Risk assessment 
DIN EN 61326-1 


Authorized person for the technical documentation: 


Detlef Opp 
Stockland3 
D-56412 Niederelbert 


Niederelbert, 1. April 2010 


Detlef Opp, Head of Technical documentation 
Signature 


Thermo Thermo Scientific Barnstead GenPure with xCAD 
a Ultrapure water system 


Preface 


Dear Sir or Madam, 


In deciding to purchase a high-purity water system of type GenPure with xCAD you have se- 
lected a high-quality product. 


Thank you for the confidence you have placed in us. 

Before you start to install and operate your pure water system, please carefully read the infor- 
mation given in these Operating Instructions on how to correctly carry out installation and how 
to properly operate the system. 

This is particularly important, as we, the manufacturer, cannot accept liability for any damage 


occurring as a result of improper operation of this system, or from use of it for other than the 
intended purpose. 


Niederelbert, 20.03.2008 


Thermo Thermo Scientific Barnstead GenPure with xCAD 
iad ns Ultrapure water system 


1. Contents 


1- Greg) (=) 6) Scere ener eee oe eee nt A ne ee E te meres erent eer 4 
2. Explanatory notes on the operating instructions ................:::eceeeeesseeeeeeees 6 
3. Transport and packaging vcvcssiccccinecstsnonsaresturcrnceeonesesn vecsuitenzenvcriueduedéces T7 
3.1 Examination On reCeipt ...........cccccecccecccecccceecceeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeettes 7 
3.2- COMPLAINS sensensa a a a a Ea avin e E RE EENE 7 
3.3 Packaging and return SHIDMENE: .xccicsrtaloacecsccte: steeesteteatedsatean sdanteetictdetnaucsassDibestcendidesaseans Bene 7 
4. Safety precautions axcsariiisssccitessscuceswsdartadenkadissanccdalawicdestemcuctanGiiveaiantdens 8 
5. Intended purpose ecpiiecsctccccpaneddd ainetatausenievecdvanewsaenreid anesiebioreeecsdenuecdndans 10 
5:1 Application areas iiron ds ain in ee aie i es 10 
6. Extent of delivery so iiccccomntonwuerecetaasematorntwcaiubinidecsdeddemeconemuccameatumnedes 11 
7. Technical specifications ice siiciecsnevsicianieabisidesanclucietumeciaeeneeas 13 
8. Description of how the systems fUNCTION...............ccccceeeeeeeeeeeeeeeaeeeeeeeeees 17 
8.1 Flow chart, GenPure Standard with xCAD o.oo... ccc cccccceeesceeceseeceeeeeeeeeeeseeeeeeeaseeeeeeenanes 18 
8.2 Flow chart , GenPure UV with XCAD..... ccc cccccccsceeceeseeeeeeeeeeseeeueeeeeeeeeueeeaeeenens 19 
8.3 Flow chart, GenPure UF with xCAD ....... ccc ccc ccesceceseeeeseeeesseeueeseceeeeeeeeeeeseueueeeaseenaess 20 
8.4 Flow chart, GenPure UV/UF with XCAD uu... ccc cc ceccce esse ecseeeceeeeeeseeeseeeseueaeeeeeeeenegs 21 
8.5 Flow chart, GenPure UV-TOC with XCAD uo. ccc ccccccccsseeceseeceeeeueeeeeeseeeeseeeeeueeneenaees 22 
8.6 Flow chart, GenPure UV-TOC/UF with XCAD o.oo ccc cccceceecsseeceseeseeeeeeseeeeeseeueeeeseeeanes 23 
9. Setting up the system ccc sanceesccceassilenndernanetincciaccewessceassuectenouberetietoaeedios: 24 
9.1 Installation Area cccncccacacensveaasntccareauareesarauadehenencgansay aattnnastdedanavasxacaounntessabeaseeacnareoneecesmuaeonts 24 
9.2 OS Ui a aerarii ee a ea A eaa e A E a S ete 25 
9.3 General setting up of the system / bench version XCAD ...........c:ccceeeeeeeeeeeeeeeeeeeeeeeeeeeeeaaas 26 
9.4 Setting up the system / wall version XCAD.............::::::ccceeeeeeeeeeseeeeeeeeeeeseeeeseeeeeeeeeeneeeeences 27 
9.5 Wall mounting GenPure system asa scetassinteetiincencostscigeeddecacheualeidsesiaeateiecastegetiavsadanuncdeaeatvatdee 30 
9.6 Mounting the power pack (voltage supply)........eessssssssereesseseerirrrerrreeesirrrrrnnnsssrnnnnnrnsssseene 31 
9.7 Examples of system Set-UDS ...............ccceeeeeeeeeeee eect eeee cece eeeeeegeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeneeeees 32 
10. Putting the system into OperatiOn.................cccccecseeeeeeeeeeaeeeeeeeeeeeeneeeeeees 34 
11. Operating SClOMISIIS vessss a sacceveviieaverwctccrsicevdiinsssnuicbdusawevesteytcem andiddsaatetee 35 
12: System control oecite ie e rE EERE AEE E EREE EERE 36 
TZA MENU ise aa cai a E E E E eects aad te cor eagaane 37 
12.1.1 Feedwater conductivity: .2der0tdcedencotendtlenstienstacestaelsidadastnedeasdsagbtenticcns peoelaedesveldsdnecere? 37 
12.1.2 Ultra pure water limiting value: ncsucredincicttiobndnsetinaatmretdccunetdbasattetatieanehetnetsiealderendss 37 
12.1.3 UV-Lamp operating time and intensity: ............ccccceeseeeeenceeeeeeeeeeneeeeeeeeeeeeeeeneseeeeeeees 38 
12.1.4 Filter cartridge operating time counter: ..............:ccccceeceeenceeeeeeeeeeeeeeeeeeeeeeeneneeneeeeeeees 38 
12:14:95 RINSING arnncmntcedssteaanscecmebicantesnneenacancatenxantteds aldo aa Aaa ARARAT T ANNESSA 38 
12.1.0 [DISIMPOCUON sccdictsnsacnicxerecvariiis stcunlnccaddicbadue sa cecita ie daeesunka dana CaA Eaa EEA EAEE NENEA 39 


Thermo Thermo Scientific Barnstead GenPure with xCAD 


SCIENTIFIC Ultrapure water system 
12.1:-7 Display of TAGE SIC AGS sain es ca saci tee ttccendcteadaeedededeicanas tenier ii eiki 40 
12.1280 PCIE OU sssri taan aE EEA Ra ea 41 
12.1.9 Unlocking the system ..........ssssesssseeeeesseesesirirttttetttrrrritttteeeertrinttnsssstttnnnnnnnnnatenennnnnnat 41 

12-2 QEM MENU cag ttacgeetenacttaieh tacts cceamiah eerie aed ta vesen enti dati cleoet EAEan aain 43 
12.2.1 Setting the maximum operating temperature: .......... eee eeeeeeeee cette eeeeeeeeeeeeeeeeeeeeeeaaae 43 
12.2.2 Setting the rinsing TMG igccsisicasss cetnvecdcepeas nebaetucverks cana eabocebhiceebeavansayuantnedeseccaanaanevedte 44 
12.2.3 Changing the disinfection time: scccccccv.cirs.0sse-seceeeneav.ssanduc tes bicanadede udanendecndedes duane stones 44 
12.2.4 Setting the interval pump tiMe?............... ccc cceceeeneeeeeeeaeeeaeeaaeeeaeeaaeeaaeeaaaeaaaeaaeeaaasaaaeees 44 
12.2.5 Setting the interval rinse LIMO! as stviccctrctetniincincaredcteatdsieatsneeientasteeardsuonseteeaebidaanncendbs 45 
12.2.6 Setting the real-time clock: ....ussssssiseeesesesesssessssorosrrrrnnnnnrrrrnnnunnunuonnnnnnnnnuenaaeeeeeneeeee 45 
12.2.7 Setting the sending interval: caisscecscccccevsanaisacttecertataraecanedeoseschanavehduk us tedeeerevxarkieasminetss 45 
12.2.8 Selecting the lANQUAGE fiiccicciassidcancccnseneanunnd stdecebaeian Gans scuiebinantexaion santnmdeneetanaviamicents 46 
12:29. SWwitehing WING inini ian a anehe ee Rani daad iani aaa A aana a S4ecicitcaaddgacatenrs 46 
12.2.10 Switching temperature compensation on or Off: .............::ccccceseeeeeeeeeeeeeeeeeeeeeeeeaeeees 46 
12.2.11 Entering the system type and serial NUMDE-?.............ccccceeeeeeeeeeeeeeteeeeeeeeeteeeeenaaaeees 47 
12.2.12 Water dispensing via volume control ............:cccceeeeseeeeneeceeeeseeeeeseeeeceeeeseeneseenenseeees 47 

12.3 Printer output.. s-ren cece cece cece cece cece cece eect eee dane aan EE EE EEE EEEEEEEE EEE EEE NE EA TERE A Eanan i Ra nina EEA 48 
‘12st. Standard (MESSAGES: cre. kaencstasnsscisncdceatiderathanaeumedtaateaag deena tictudtetahdealegteinedeteetaadanactyinens 48 
12.3:2 Code messages a cat Sica i neta ssc a heeds alps decedent eal 48 
12:3-3 Fault IIMS SAS eenia N EE a O E 49 

13; TEMAS ANG -srra n aaa boc oo aia EE aE EE entiey 50 

13.1 Maintenance intervals .............nnsnssssrnssesrnrnnrnnsssussnenrsnneeesocoosssarnstrrrrnrrrnnnnnneurnnnnnunaauseeeenee 51 

13.2 Replacing the filter CANN Ss icsetictcticstinceids cabipecdddsvalaiademeatlatnots tnesssouelesaaldesendioacubacetstes 52 

13:3: DISINTSCHUO Negrie ieee tees eecwte see dea secect cece es eas neous mada E E 53 

14: Waste disposal css ssa ic nieri ea eae e aa AEEA aE EEEa 55 

15. Trouble shooting..............esseoenenneeonnnnneernnnnneernnnnnternnnnrrennnnnerennnnennnnnn nne 56 

16. Replacement parts and consumables ................ccccseeeeeeeeeeeeeeeeeeeaaeeees 58 

16:1 GenPU Gins aa aaa aa aeaa ora a aa i aAa nee meee 58 

16 2- XSAD ee a ene ee era rer ee ree ee rere eer ree 59 

17. Consumables and accessories .........ccceeee cece eeeeeeeeeeeeeeeeeeeeeeeeeaaaaeeeeeeees 60 

18. Terminal ASSIGNING sis i12s orrtaerencinrnscadeetednnecincemernetatedadiauenetcotss 62 

19. Maintenance (ECON cso eectteasetcce ianamprge seeds ate Gaakie cence tetherewncie 63 


Thermo Scientific Barnstead GenPure with xCAD 
Ultrapure water system 


2. Explanatory notes on the operating instructions 


d: 
(SA 
Z\ 
(i) 


= / 


D 


EU Mark of Conformity 


CSA - admission 


Important operating and/or maintenance instructions! Read the oper- 
ating instructions with due care. 


Risk of electric shock! Electrical work on the system is only to be car- 
ried out by qualified personnel. 


General information! Particularly important notes are marked with this 
information sign. 


Protective conductor connection 


Connect the power supply to an electrical socket with a protective 
connection. 


The information provided in these operating instructions is only valid for the system which has 
the serial number which is to be entered on the front page. 


Please enter the serial number* of your GenPure with xCAD 
system in the space provided on the front page. 


* Read the serial number of your ultra pure water system from the type plate. 


For quick and correct service, please include the following information on all inquiries and re- 
placement parts orders which relate to your system: 


- The serial number 


- The article number 


Thermo Thermo Scientific Barnstead GenPure with xCAD 
SETE Ultrapure water system 


3. Transport and packaging 

Ultra pure water systems are carefully checked and packed prior to shipment but, despite this, 
there is always a possibility that damage could occur to them during shipment. 

3.1 Examination on receipt 


> Check the completeness of the delivery against the shipping papers. 


Is the packaging damaged? 


> Check the system for damage. 


3.2 Complaints 
Should the system have been damaged during transport: 
> Immediately contact the post office, railway or forwarding agent*. 


> Keep the packaging, including the outer cardboard box, for a possible inspection 
and/or return shipment. 


3.3 Packaging and return shipment 
Whenever possible use the original packaging and packing materials. 


Should these no longer be available: 


> Pack the system in packing film and then in a strong cardboard box so that it is held 
shock-proof. 


* Complaints are only valid for 6 days (after receipt of the goods). 
After this time, the right to claim for damages expires. 


Thermo Thermo Scientific Barnstead GenPure with xCAD 
ee ee Ultrapure water system 


4. Safety precautions 


For your own safety, please observe the above safety precautions! 


>  GenPure series systems are modern ultra pure water systems that are intended ex- 
clusively for the purification of tap water of drinking quality that has been previously 
treated by reverse osmosis, ion exchange or distillation. 


> Do not install the system and operate it until you have read through the appropriate in- 
formation that is given in these Operating Instructions. 


> Lifting and carrying the ultra pure water system, e.g. to the installation location, should 
be carried out by two people. To lift it, each takes hold of it under the base plate at two 
corners. 


> — The CE-Mark becomes invalidated when constructional changes are made to the sys- 
tem or when products of other manufacturers are installed in it. 


> Protect the system from frost.The temperature at the installation area must be at least 
+2°C. 


> Observe all general requirements and regulations, including current accident preven- 
tion regulations, that are applicable at the ultra pure water system installation area. 


> The feedwater pressure must be at least 0.1 bar and at most 6 bar. Should the feed- 
water pressure be higher, then an additional pressure reducer must be installed. 


> Water purification systems must have a safety device acc. to DIN EN 1717 to protect 
the tap water from contamination. 


> A grounded 100 - 240V/ 50/60Hz socket must be available. 


> The installation area must be equipped with a floor drain with at least DN 50 pipe (40 
mm o.d.). 


> When the system is to be wall-mounted, first check the statics of the wall. It must have 
sufficient load-bearing capacity (for weights, see the technical specifications). 


> The maximum operating temperature is +40°C. 


> When the system will be at a standstill for a longer period (e.g.or during long holidays), 
proceed as follows: 
— Switch the system off. 
— Close the supply of feedwater to the ultrapure water system. 
Damage to the pump will occur should the supply of feedwater be turned off with the 
system still on. The manufacturer does not accept liability should such damage occur. 


> — When planning the installation of the system, ensure that there will be sufficient work- 
ing room for convenient operation of the system and for maintenance such as chang- 
ing the filter cartridge and opening, breaking and checking connections. 


Thermo Thermo Scientific Barnstead GenPure with xCAD 
Ultrapure water system 


> The term of the guarantee is12 months! 


> — Never look directly at a switched-on UV-lamp, because UV light endangers eyesight! 
Never switch the UV-lamp on when is has been taken out of the metal cylinder! 


> Only authorized personnel are to carry out UV-lamp replacement. 


Thermo Thermo Scientific Barnstead GenPure with xCAD 
Se Ultrapure water system 


5. Intended purpose 


The continuous increase in ultra pure water quality requirements, technologies that are increas- 
ingly more sophisticated, lowered detection limits in laboratories and the need of for user- 
friendly systems and complete solutions were the basic for the development of the novel ultra 
pure water systems that we call GenPure with xCAD. 


GenPure with xCAD has been specifically developed for the production of salts-free, organically 
pure, particle-free and sterile filtered ultra pure water. 


To ensure long service lives of the high-quality purification media, this type of ultra pure water 
system must be fed with water from upstream pre-purification steps (reverse osmosis, ion ex- 
change, distillation). 


5.1 Application areas 


- Analytical techniques in laboratories: 


- HPLC ( High Performance Liquid Chromatograhy ) 
-IC ( lon Chromatography ) 

- ICP ( Inductive Coupled Argon Plasma ) 

- AAS ( Atomic Absorption Spectrophotometry ) 

- TOC-Analysis ( Total Organic Carbon ) 

- DNA-Research 

- etc. 


- Preparation of reagents and solutions: 
- Cell culture media 
- Tissue culture media 


- Make-up water for reagents for 
online automatic analyzers 


- Ultra pure water for demanding laboratory rinsing processes 


10 


Thermo Thermo Scientific Barnstead GenPure with xCAD 
SEEE Ultrapure water system 


6. Extent of delivery 


The GenPure with xCAD ultra pure water system is available in the following versions: 


GenPure Standard, article no. 50131286 (Basic system) with xCAD, bench version 
GenPure Standard, article no. 50131300 (Basic system) with xCAD, wall version 


GenPure UF, article no. 50131250 (Basic system + Ultrafiltration module) with xCAD, 
bench version 

GenPure UF, article no. 50131302 (Basic system + Ultrafiltration module) with xCAD, 
wall version 

GenPure UV, article no. 50131254 (Basic system + UV-photooxidation) with xCAD, 
bench version 

GenPure UV, article no. 50131317 (Basic system + UV-photooxidation) with xCAD, 
wall version 


GenPure UV/UF, article no. 50131252 (Basic system + UV-photooxidation + Ultrafiltration 
module) with xCAD, bench version 

GenPure UV/UF, article no. 50131315 (Basic system + UV-photooxidation + Ultrafiltration 
module) with xCAD, wall version 

GenPure UV-TOC, article no. 50131296 (Basic system + UV-photooxidation and TOC meas- 
urement) with xCAD, bench version 

GenPure UV-TOC, article no. 50131321 (Basic system + UV-photooxidation and TOC meas- 
urement) with xCAD, wall version 

GenPure UV-TOC/UF, article no.50131298 (Basic system + UV-photooxidation and TOC meas- 
urement + ultrafiltration) with xCAD, bench version 

GenPure UV-TOC/UF, article no.50131323 (Basic system + UV-photooxidation and TOC meas- 
urement + ultrafiltration) with xCAD, wall version 


(Please check your version against the article number given on the delivery note) 


1x GenPure with xCAD (acc. to the version) Article no. 5013xx 


incl. Assembly kit for bench and wall version, 

consisting of: Article no. 50131283 
(bench version) 
Article no. 50131380 
(wall version) 


Filter cartridge for the ultra pure water system Article no. 09.2005 
Sterile filter capsule, 0.2 um Article no. 09.1003 
PE-Hose 8 x 1mm, 2 m Article no. 18.0036 
PE-Hose 6 x 1mm, 5m Article no. 18.0047 
PE-Hose 4 x 1mm, 5m Article no. 18.0053 
SUB-D Extension cable, 25-pin, 5 m, 2x Article no. 16.0375 
Screw, 4 x 40 mm (only wall version version) Article no. 21.0001 
Plug, 2 x S8 Article no. 21.0035 
Screw hook 2 x 5,2 x 50 mm Article no. 21.0057 
Plug 4 x S6 (only wall version) Article no. 21.0002 
Operating Instructions Article no. 50131281 
Tabletop power pack, 24V DC Article no. 50134196 
Universal adapter Article no. 21.1006 
Universal holder Article no. 21.1007 
Connecting cord (rubber connector to nema plug connector) Article no. 50132200 
Connecting cord (rubber connector to british ST plug connector) Article no. 50132203 
Connecting cord (rubber connector to euro plug connector) Article no. 50132215 





11 


Thermo Thermo Scientific Barnstead GenPure with xCAD 
ile kali Ultrapure water system 





Feedwater connector kit Article no. 25.0075 
consisting of: 

Union nut, R 3/4“ Article no. 14.0003 
Insert, R 1/4“ Article no. 14.0189 
Screwed connector, R 1/4“ Article no. 14.0075 
Gasket, R 3/4" Article no. 21.5008 
PE-Hose, 8 mm o.d., 2m Article no. 18.0036 
Dirt-trap sieve for union nut, R 3/4“ Article no. 14.0390 





12 


Thermo Scientific Barnstead GenPure with xCAD 
Ultrapure water system 


7. Technical specifications 


| Demands on the feedwater | 





Source 


Tap water pretreated by reverse osmosis, ion ex- 
change or distillation. 





Silt density index (SDI) 


max. 1 for all versions. For water that is not pretreated 
by reverse osmosis, a prefilter with a 1 um membrane 
is recommended. 



































Feedwater resistance >0.5 MOxcm 

Free chlorine max. 0.05 ppm 

TOC max. 50 ppb 

Bacteria count < 100 CFU/ml 

Turbidity <1.0 NTU 

Carbon dioxide (CO2) max. 30 ppm 

Silicate max. 2 ppm 

Particles Filtration to 0.2 um is recommended to protect the in- 
ternal filter and, if appropriate, the final filter. 

Temperature 2-35°C 

Pressure 0.1 - 6 bar 


Product water quality 





Standard 


UV- 


UN UF TOC/UF 





Resistance 
(Reference 
temperature 
25°C) 


18.2 





TOC 
Bacteria 


<1 <1 <1 





Bacterial En- 
dotoxines 


< 0.001*| <0.001* < 0.001* 





Particles 
Flow rate 


L/min** 


< 1/ml < 1/ml < 1/ml 





Flow rate with 


L/min 
volume control 























* Depending on feedwater and appropriate disinfection 
** Depending on feedwater pressure 


Dimensions, GenPure 





Height 


615 mm 





Width 


372 mm 





Depth 


330 mm 





Weight: 





GenPure Standard 





GenPure UF 
GenPure UV 





GenPure UV/UF 





GenPure UV-TOC 





GenPure UV-TOC/UF 








13 


Thermo Scientific Barnstead GenPure with xCAD 
Ultrapure water system 


Dimensions, xCAD (bench version) 


approx. 725 mm 

260 mm 

approx. 530 mm 
12 kg 






































approx 








Dimensions, xCAD (wall version) 


approx. 655 mm 

260 mm 

approx. 530 mm 
Weight 
































f 
= 
= » 
= WA N 


@ 





£ 


{ \ 


E 
i 

4 
Q 


pa, 
A Sp 








14 


Thermo Thermo Scientific Barnstead GenPure with xCAD 
AAEE Ultrapure water system 


Cell constants, measuring cells 
Conductivity, feedwater 0.16 cm“ 
Conductivity, after UV-photooxidation 0.01 cm“ 
Conductivity, ultra pure water 0.01 cm“ 














Connectors for water genpure 


Feedwater Hose, 8 mm o.d. 
Rinse water Hose, 8 mm o.d. 
Ultra pure water/outlet Hose, 6 mm o.d. 
Ultra pure water/recirculation Hose, 4 mm o.d. 














Connectors for water, xCAD 
Ultra pure water/inlet Hose, 6 mm o.d. 
Ultra pure water/recirculation Hose, 4 mm o.d. 
Ultra pure water/outlet R 1/4" 
Sterile filter outlet Hose, 8 - 10 mm o.d. 

















Electrical connections / external switched mode power supply 
Input voltage AC 100 — 240 V, 50 — 60 Hz, 5 - 3.8 A 
Output voltage DC 24 V, 3,8 A 
System connection DC 24 V, 80 W 
Serial interface RS 232 
Protection Class Class II (external SMPS certified as Class I) 




















Electrical connections, xCAD 
2x SUB-D socket 25 pin 





Airborne sound emission 





Sound-pressure level 49 db(A) 


Ambient conditions (DIN EN 61010-1 (VDE 0411-1):2011-02) 


Usage Indoor rooms 
Height Up to 2000 m 
Temperature range From 5° C to 40° C 
Relative humidity Maximum relative humidity 80 % at temperatures of 
up to 31° C, linearly decreasing to 50 % relative hu- 
midity at 40° C 
Line-voltage variation Not more than + 10 % of the line voltage 
Transient overvoltages As usually occur in the supply network (overvoltage 
category II acc. to IEC 60364-4-443). 
Note: The rated level of transient overvoltage is the 
withstand impulse voltage acc. to overvoltage cate- 

Il of IEC 60364-4-443 
Ventilation requirements There are no special requirements with regard to ven- 
tilation. 
Degree of pollution 2 





























15 


Thermo Scientific Barnstead GenPure with xCAD 
Ultrapure water system 


Materials of parts that contact water 





Pressure reducer 


NBR 





Pump head 


Nylon with glass fibre 





UV-Lamp 


High purity quartz 





UV-Housing 


Stainless steel 





Filter cartridge 


PP 





UF-Housing 


Polycarbonate 





Rinsing solenoid valve 





Dispensing valve 
Conductivity measuring cell 


POM, stainless steel 





Distributor block 


POM 





Connections 


POM 





Hoses 
Gaskets 


16 





PE 





Thermo Thermo Scientific Barnstead GenPure with xCAD 
SETE Ultrapure water system 


8. Description of how the systems function 


GenPure versions: Standard, UV, UF, UV/UF with xCAD 


Tap water that has been pretreated upstream by reverse osmosis, ion exchange or distillation 
passes through a pressure reducer and into the GenPure ultra pure water system, where the 
conductivity is monitored. A pump forces this feedwater through UV-photooxidation (only with 
GenPure UV and GenPure UV/UF) and the filter cartridge, from where it flows through an ul- 
trafiltration module (only with GenPure UF and GenPure UV/UF), after which the conductivity is 
measured by a special conductivity measuring cell (with temperature compensation). 


When ultra pure water is taken from the system, it flows through a sterile filter before reaching 
the dispensing outlet. During interval operation, the water is circulated around an internal circuit 
at regular intervals. 


GenPure versions: UV-TOC, UV-TOC/UF with xCAD 


Tap water that has been pretreated upstream by reverse osmosis, ion exchange or distillation 
passes through a pressure reducer and into the GenPure ultra pure water system, where the 
conductivity is monitored. A pump forces this feedwater through UV-photooxidation, after which 
the conductivity is again measured to determine the TOC-value. The water then flows through 
the filter cartridge and an ultrafiltration module (only with GenPure UF-TOC/UF), after which the 
conductivity of it is measured by a special conductivity measuring cell (with temperature com- 
pensation). 


The TOC value is calculated by taking the difference between the values measured by the 
measuring cells QIA300 and QI302. 

The measurement range is 0 - 30 ppb. 

When this measurement range is exceeded, the number 99 is shown in the display instead of 
the measured value. 

In Stand-by operation, "___" is shown instead of the measured value. 


When ultra pure water is taken from the system, it flows through a sterile filter before reaching 


the dispensing outlet. During interval operation, the water is circulated around an internal circuit 
at regular intervals. 


17 


8.1 Flow chart, GenPure Standard with xCAD 


peisewasser 
Eingan 


Feedwater inlet ` 


A-F1 
A-F2 
A-P1 
FIS400 
QIA 300 
QIA 301 
TIA 500 


18 





GenPore Standard 


Abfl 





rain 








TIL 


Thermo Scientific Barnstead GenPure with xCAD 
Ultrapure water system 













































































Filter cartridge 

Sterile filter 

Circulation pump 
Digital flowmeter 
Conductivity, feedwater 
Conductivity, ultra pure water 
Temperature sensor 
Pressure reducer 
Check valve 

Solenoid valve 

Rinsing solenoid valve 
Check valve 1 bar 

















8.2 Flow chart , GenPure UV with xCAD 


peisewasser 


Ein si i m> 


Feedwater inlet 


A-F1 
A-F2 
A-P1 
A-UV1 
FIS400 
QIA 300 
QIA 301 
TIA 500 


GenPore 


Thermo Scientific Barnstead GenPure with xCAD 
Ultrapure water system 

















z 
= 
o 
ee 
= 
mn 










































































Filter cartridge 

Sterile filter 

Circulation pump 
UV-Photooxidation 
Digital flowmeter 
Conductivity feedwater 
Conductivity, ultra pure water 
Temperature sensor 
Pressure reducer 
Check valve 

Solenoid valve 

Rinsing solenoid valve 
Check valve 1 bar 














19 





Thermo Scientific Barnstead GenPure with xCAD 
Ultrapure water system 


8.3 Flow chart, GenPure UF with xCAD 


peisewasser 
Eingar 
Feed 





ater inle 


A-F1 
A-F2 
A-F3 
A-P1 
FIS400 
QIA 300 
QIA 301 
TIA 500 


20 


Lh 


GenPure LF 











= 
i 
1 
=o 
| 






























































© o Z -F 




















Filter cartridge 

Sterile filter 
Ultrafiltration module 
Circulation pump 
Digital flowmeter 
Conductivity, feedwater 
Conductivity, ultra pure water 
Temperature sensor 
Pressure reducer 
Check valve 

Solenoid valve 

Rinsing solenoid valve 
Check valve 1 bar 














KZ 


Reinstwasser 





Itrapure water 


rain 





Thermo Thermo Scientific Barnstead GenPure with xCAD 
SETE Ultrapure water system 


8.4 Flow chart, GenPure UV/UF with xCAD 


GenPure F 





























ala L EIS í ` AF3 QIA YTIA 
peisewasser 3 1 ak ou) N v3 
Eingan D ec | L . } X 
I \ 
Feedwater inle v ” i LX 
po A-F Ke 




















































































































pa 


Reinstwasser 
ltrapure water 





rain 


A-F1 Filter cartridge 

A-F2 Sterile filter 

A-F3 Ultrafiltration module 
A-P1 Circulation pump 


A-UV1 UV-Photooxidation 

FIS 400 Digital flowmeter 

QIA 300 Conductivity, feedwater 

QIA 301 Conductivity, ultra pure water 
TIA 500 Temperature sensor 


V1 Pressure reducer 

V2 Check valve 

V3 Solenoid valve 

V4 Rinsing solenoid valve 
V5 Check valve 1 bar 


21 


peisewasser 
Eincan 
Feedwater inle 


A-F1 
A-F2 
A-P1 
A-UV1 
FIS400 
QIA 300 
QIA 301 
QI 302 
QIA 303 
TIA 500 


22 





Thermo Scientific Barnstead GenPure with xCAD 


Ultrapure water system 

















































































































Filter cartridge 

Sterile filter 

Circulation pump 
UV-Photooxidation 

Digital flowmeter 
Conductivity, feedwater 
UV-Intensity 

Conductivity, TOC measurement 
Conductivity, ultra pure water 
Temperature sensor 
Pressure reducer 

Check valve 

Solenoid valve 

Rinsing solenoid valve 
Check valve 1 bar 














Reinstwa 
Itrapure 


arer 





Eingan 
eedwater inle 


2 m 





Thermo Scientific Barnstead GenPure with xCAD 
Ultrapure water system 
































A-F1 
A-F2 
A-F3 
A-P1 
A-UV1 
FIS400 
QIA 300 
QIA 301 
QI 302 
QIA 303 
TIA 500 





ala \TIA 

































































Filter cartridge 

Sterile filter 

Ultrafiltration module 
Circulation pump 
UV-Photooxidation 

Digital flowmeter 
Conductivity, feedwater 
UV-Intensity 

Conductivity, TOC measurement 
Conductivity, ultra pure water 
Temperature sensor 
Pressure reducer 

Check valve 

Solenoid valve 

Rinsing solenoid valve 
Check valve 1 bar 














Reinstwasser 
ltrapure water 


23 





Thermo Thermo Scientific Barnstead GenPure with xCAD 
Se re Ultrapure water system 


9. Setting up the system 


9.1 


Installation area 


Take the following criteria into consideration when selecting the installation area: 


24 


Vv 


Feedwater pressure at least 0,1 bar, at most 6 bar 
Lowest temperature +2°C 
Level standing surface 


When the system is to be wall-mounted: A smooth wall is required. Check the statics 
of the wall, it must be sufficiently loadable (for weight, see Technical specifications) 


Floor drain, DN 50 
Free gravity fall to drain. 


Should no such floor drain be available, then, for safety reasons, a water watcher (ar- 
ticle no. 16.0129) must be installed! 











Free gravity fall to drain must be ensured ! 





Safety socket, 100 - 240V, 50/60Hz 
Sufficient working space around the system (filter replacement, etc.) 
It must be easy to operate and monitor the system 


R 3/4" Tap water connector 


9.2 Connectors 


Genk 


re 


Thermo Scientific Barnstead GenPure with xCAD 
Ultrapure water system 


































































































Connectors, GenPure: 


Connector for 25 pin sockets (system control) 
Connector for 25 pin sockets (system control) 
Feedwater connector, 8 mm o.d. 

Rinse water connector, 8 mm o.d. 

Ultra pure water connector, 6 mm o.d. (to xCAD) 
Ultra pure water connector, 4 mm o.d. (to xCAD) 
Connector for optional printer (art. no.: 09.2207) 
Mains plug/100 - 240V, 50/60Hz power supply 


Connectors, xCAD: 


Connector for 25 pin sockets (system control) 
Connector for 25 pin sockets (system control) 

Ultra pure water connector, 6 mm o.d. (to GenPure) 
Ultra pure water connector, 4 mm o.d (to GenPure) 
Dispensing valve outlet, R 1/4” female thread 
Sterile filter 0,2 um 


20 


Thermo Thermo Scientific Barnstead GenPure with xCAD 


Ultrapure water system 











9.3 General setting up of the system / bench version xCAD 


1) 
2) 


3) 


4) 


5) 


6) 


26 


Proceed as follows to set your GenPure with xCAD (bench version) ultra pure water 
system up ready for use: 


Stand the GenPure system with the xCAD where it is to be installed. 
Remove the cartidge cover from the GenPure system. 
Take the plugs off from the filter cartridge and save them. 


Push the filter cartridge in the guide in the system, plug the quick-connect connectors on so 
that they audibly click into position and put the cartridge cover back on. 


Connect the 8 mm o.d. hose that is provided in the connector kit to feedwater connector (3) 
of the GenPure system. 


Now make a pressureless connection between the 8 mm o.d. hose (4) and the drain. 
The drain to the sewer must be max. Are 1m above the rinsing water connector of the Unit. 


Plug each of the extension cable cables with 25 pin socket in the sockets (1) and (2) of the 
GenPure System and screw them tight. 

Then connect the tubing (4 mm and 6 mm) by plugging the connectors (5) and (6) of the 
system. 


Thermo Thermo Scientific Barnstead GenPure with xCAD 
SAE Ultrapure water system 


7) Lead the 25 pin plugs/extension cables and 4 mm and 6 mm o.d. hoses through the wall 
and connect the 25 pin plug (1) from the GenPure with the socket (1) on the xCAD and the 
25 pin plug (2) from the GenPure with the socket (2) on the xCAD. 
Caution: Do not interchanging the connections of the 25 pin connector, otherwise no ab- 
straction of water possible! 

8) Fix the 6 mm und 4 mm o.d. hoses in the hose connectors (5) and 6) of the xCAD. 


9) If you intend to use the sterile filter (10) that is standardly supplied, screw it in the R 1/4" 
female thread outlet of the dispensing valve (9) of the xCAD. 


10) Connect the mains plug (8) of the voltage supply to an earthed 100-240V/50/60Hz socket. 
11) Open the feedwater supply. 


12) Check all connections for leaks. 


9.4 Setting up the system / wall version xCAD 


Proceed as follows to set your GenPure with xCAD (wall version) ultra pure water 
system up ready for use: 


View from below/front 


xCAD mit montiertem Wandhalter 













































































xCAD with mounted wall mount Wandhalter 
wall mount 
- E 
N Schrauben 4x12 mm 
7 screws 4x12 mm 
132 
_ 12, | 10 
N o 2h 8h 
d4 
p 5 f 
Ne = a 
Ol 0O 
































1) Release the four screws (4x12 mm) marked above from the xCAD underside and 
remove the wall mount from it. 


27 


Thermo Thermo Scientific Barnstead GenPure with xCAD 


2) 


3) 
4) 


5) 


6) 


7) 


8) 


28 


Ultrapure water system 


Installing the wall mount 


Caution: The included screws and dowels are only intended to fix the wall mount 
to a concrete wall or a solid masonry! 


The two 25 pin plugs/extension cables and the 4 mm o.d. and 6 mm o.d. hoses can be led 
laterally down the wall or be led out through the wall at the back. There are three possibili- 
ties for cut-outs for the cable and hoses (see the next diagram). 


Note: Only one of these cut-outs is required. 


Hold the wall mount at the wanted position on the wall and mark the four boreholes 
for the fixing of the wall mount. 


Wand 
wall 
Yypyy min.18 
_ YXGF rT 
ZZ * 
ZZ Z l 
Uy Uy, 
AY Ao “Yi 
Yd AY 
VK Ay wo 
AN Wij, O 
yf VI 
Vi VAY 
Za: 
U : 
Wandhalter moglicher Ausschnitt in der Wand, für Kabel- 
und Schlauchdurchführung nach hinten 
wall mount 


possible wall cut- outs for cable 
and hose taken out at the back 
Bore the holes for the 6 mm plugs and make the necessary cut-out. 
Fix the wall mount in position with the plugs and screws provided. 
Lead the 25 pin plugs/extension cables and 4 mm and 6 mm o.d. hoses through the wall 


to the wall mount and connect the 25 pin plug (1) from the GenPure with the socket (1) on 
the xCAD and the 25 pin plug (2) from the GenPure with the socket (2) on the xCAD. 


Caution: Do not interchanging the connections of the 25-pin connector, otherwise no ab- 
straction of water possible! 


Fix the 6 mm und 4 mm o.d. hoses in the hose connectors (5) and (6) of the xCAD. 


Now carefully position the xCAD on the wall mount and screw it tight to the wall 
mount using the four 4x12 mm screws. 


Place the GenPure system at the planned position. 


Thermo Thermo Scientific Barnstead GenPure with xCAD 
AAEE Ultrapure water system 


9) Take the cartridge cover off of the GenPure system 
10) Take the plugs off from the filter cartridge and save them. 


11) Push the filter cartridge in the guide in the system, plug the quick-connect connectors on so 
that they audibly click into position and put the cartridge cover back on. 


12) Connect the 8 mm o.d. hose that is provided in the connector kit to the feedwater connector 
(3) of the GenPure system. 


13) Now use the 8 mm o.d. hose to make a pressureless connection between the system 
(connector 4) and the drain. The drain to the sewer must be max. Are 1m 
above the rinsing water connector of the Unit. 


14) Plug the two extension cables with 25 pin plugs in the sockets (1) and (2) of the GenPure 
system and screw them tight. 


15) If you intend to use the sterile filter (10) that is standardly supplied, screw it in the R 1/4" 
female thread outlet of the dispensing valve (9) of the xCAD. 


16) Open the feedwater supply. 


17) Check all connections for leaks. 


29 


Thermo Thermo Scientific Barnstead GenPure with xCAD 
eee Ultrapure water system 


9.5 Wall mounting GenPure system 


Ansicht von hinten 


Le acl ide 




















= 
| 











ohruncen für 
andmontace 





J Holes for wall 
mountin 








ii 






































Proceed as follows to mount your GenPure system to a wall: 
> Use a twist frill (8 mm or 5/16 inch) to make the two holes in the wall that are required as 
shown in the diagram above. 


> Plug the nylon S8 dowels that are supplied in the assembly kit in the holes. Screw the 
5.2 x 50 mm screw hooks that are also supplied in the assembly kit in the dowels. 


> Lift the GenPure system (2 people are required for this) and hang the back side of it on 
the screw hooks. 


30 


Thermo Thermo Scientific Barnstead GenPure with xCAD 
SIETE Ultrapure water system 


9.6 Mounting the power pack (voltage supply) 


ack view power supp niversal adapter 
























































































































































> Whenever possible, mount the power pack on the wall to the left or right of the ultra 
pure water system where it is freely accessible. 


> Stick the universal holder which is supplied in the assembly kit to the back of the 
power pack as shown in the above Figure. 


> Stick the universal adapter to a smooth wall surface or screw it to the wall using the 
dowels and screws supplied in the assembly kit. 


> When the universal holder and universal adapter have been fitted, hang the power 
pack in. 


> Plug the connecting cable (appliance cable) in the power pack socket. 


> Connect the power pack to the ultra pure water system (Connectors, GenPure, 4-pin 
power supply connector, pos. 8) and to an earthed 100 - 240V, 50/60Hz socket. 


> The system is now ready for use. 


31 


Thermo Scientific Barnstead GenPure with xCAD 


Thermo 
Ultrapure water system 


9.7 Examples of system set-ups 


Connection to a pretreatment: 


lonenaustuscher DI 1 
nicht im Lieferumfang enthalten 





lon exchanger Il 
not included in extent of delivery 

















a 
i 


S 






































































































































Connecting hose 


Connection to a Pacific TII system with storage tank: 





Pacific TIITII mit Tank und Vorbehandlun 
nicht im Lieferumfang enthalten 
Pacific TH/TI with tank and pretreatment 
not included in extent of deliver 





















































ag 






















































































Anschlussschlauch 
Connecting hose — 
T 
il 
rain 





32 


Thermo Thermo Scientific Barnstead GenPure with xCAD 
SETENE Ultrapure water system 


Connection to a storage tank: 


orratstank nicht im 
Lieferumfang enthalten 
Storrage tank not included 
in extent of deliver 





















































Anschl chlauch 
Connecting hose i - 











Caution: When connected to a storage tank, the system must be switched to 
“nonstop“ operation for ultra pure water dispensing. 


33 


Thermo Thermo Scientific Barnstead GenPure with xCAD 
i lists Ultrapure water system 


10. Putting the system into operation 





Allow the system to warm up to, or to cool down to, room temperature 
before starting it up for the first time. 














Check that all hose connections have been made as specified in the 
"Setting up the system" section. 











I Switch the system on by pressing this button. After a compulsory rinse the sys- 
tem switches to the last operating mode. 





To vent air from the system, switch the system 3 times successively to 

"Rinsing" in the menu and collect and discard approx. 5 litres of water 

each time. The ultra pure water limiting value can be gone below during 
this procedure. 





MONSIG Use this “NONSTOP“-key to switch the system to the “Nonstop“ operating 
dlià mode. 


You can set the system back to “Interval“ operation when the system has 


brought the ultra pure water to the quality you require. 
INTERVAL 


34 


Thermo Scientific Barnstead GenPure with xCAD 
Ultrapure water system 


11. Operating elements 








Intervall — ppb 
18,2 MOxcm 
18,5°C UV/TC 














NONSTOP 





Entei 


Menu, 


INTERVAL 


Switches the system on or off 


Switches "Nonstop" operation on or, in the menu, 
increases a value on display 


Confirms the value shown in a menu 
point 


Switches the menu to the next menu 
point 


Switches "Interval" operation on or, in the menu, 
decreases a value on display 


Switches the UV-lamp on or, in the menu, 
allows you to select the position in a number 
that you wish to change 


35 


Thermo Thermo Scientific Barnstead GenPure with xCAD 
i kali Ultrapure water system 


12. System control 


General information 


A press on the ON/OFF-key causes the system to run in the /nterval operating mode (see Inter- 
val mode). 

Interval operation then runs automatically every 0.5 hours for the set length of time to ensure 
the ultra pure water quality. 

When the UV-lamp is on, then the "UV" text message is displayed. When temperature compen- 
sation of measured values is active, then the "TC" message is displayed. In addition, the meas- 
ured values for the ultra pure water conductivity and the temperature are displayed. 

Should a fault exist, then a fault message is given out across the potential-free output and is 
shown in line 4 of the display. Should several faults occur simultaneously, then they are alter- 
nately displayed. 


When the Nonstop-key is now pressed, the system switches to the Nonstop mode, the pump is 
started and the solenoid rinsing valve opens for the set rinsing time. Nonstop operation can be 
stopped by pressing the Interval-key. After a wait of 2 hours, the system switches itself back to 
the Interval mode. 


A press on the UV-key brings "UV" to display. The UV-lamp is only switched on when the sys- 
tem is in the Nonstop mode, however. The UV-lamp switches itself off when the Nonstop mode 
is ended (2 h). When Nonstop operation is manually stopped by a press on the Interval-key, 
then the UV-lamp is not switched off until it has been burning for 0.5 hours. 

During the time that the UV-lamp is burning, the intensity of the UV-light in monitored. The dis- 
play of the UV-intensity is made in the menu. The TOC-value is also displayed. 


User-menu 


All of the measured values, operating times and limiting values that are relevant for the user can 
be set and read in this menu. 

A press on the menu-key brings you into this menu. Each further press on the menu-key moves 
you on from one menu prompt to the next. 

Settings can be changed with the arrow keys. When you confirm a value by pressing the Enter- 
key, you are taken to the next menu prompt. Settings can only be made when the system con- 
trol has been previously unlocked (see section12.1.8). 


To simplify changing settings, a press on the UV-key allows you to select a certain number in 
the numerical value that you wish to change. The arrow-keys can then be used to enter the 
wanted number from 0 to 9 at the selected position. 


With all three measuring cells for conductivity and with the temperature sensor, a minimum and 
a maximum measured value are fixed in the programme. Should measured values be below or 
above these values respectively, then it must be assumed that there is a cable break. In such a 
case, the appropriate fault message "Measuring cell LF1", "Measuring cell LF2", "Measuring cell 
LF3" or "Measuring cell Temp." is displayed in line 4. 


36 


Thermo Thermo Scientific Barnstead GenPure with xCAD 
SIETE Ultrapure water system 


12.1 Menu 


12.1.1 Feedwater conductivity: 


A single press on the menu-key allows the feedwater conductivity to be read and the limit- 
ing value of the feedwater conductivity to be set. The fault message that flashes in the 
fourth line of the display when the feedwater limiting value is exceeded is "Limit value 
feed". 


Measuring range, feedwater: 10.0- 0.010 MOxcm 
Setting range, limiting value: 0.1- 50.0 pS/cm 
Basic setting: 0.5 MOxcm 


With settings above 50 uS/cm, the limiting value is switched off and the word "Off" is 
shown in the display. 


The display shows: 


Feedwater 
1.25 MOxcm 


Limit value feed 
2.0 uS/cm 





12.1.2 Ultra pure water limiting value: 


Two presses on the menu-key in this menu allow the fault display for the pure water limit- 
ing value and the pure water limiting value to be set. As soon as the fault display is 
switched on, the fault will be displayed both in Stand-by mode and in Production mode. 
When the fault display is switched off, the fault is only displayed in Production mode. 

The “Lim. val.pure w.” message is displayed when the limiting value is exceeded. 


Ultrapure water measuring range: 0.100 MQxcm 
Limiting value setting range: 0.055- 5.000 pS/cm 
Basic setting: 10.0 MOxcm 

Basic setting, fault suppression: On 


With settings above 5.000 uS/cm, the limiting value is switched off and the word "Off" is 
shown in the display. 


The display shows: 


Fault message 
Stand- by On 


Lim. val.pure w. 
0.100 uS/cm 





37 


Thermo Thermo Scientific Barnstead GenPure with xCAD 
i ir kali les Ultrapure water system 


12.1.3 UV-Lamp operating time and intensity: 


In this menu, the operating hours of the UV-lamp are displayed and the input for the UV- 
sensor is evaluated. 

When the maximum operating time is reached, then the "UV-time" fault message is trig- 
gered. The UV-sensor measures the actual intensity of the UV-lamp. The display shows 
the % of this compared to the maximum value. 


The display shows: 


UV Time 
0000 h 


UV intensity 
90% 





12.1.4 Filter cartridge operating time counter: 


The operating time counter for the filter cartridge is set in this menu by a fourth press on 
the menu-key and the entry of a valid serial number for the filter cartridge.. 





The display shows: 
Ser. No.: 
Cartridge 
Press Enter 
omer 
12.1.5 Rinsing 


38 


A fifth press on the menu key in this menu allows rinsing to be carried out if this is neces- 
sary. Triggering is made by confirming the requirement with a press on the Enter-key. The 
pump is started and the rinsing solenoid valve is opened for the rinsing time period set in 
the OEM-menu. During rinsing, the count-down of the rinsing time is displayed 

When rinsing has finished, the system returns to the last operating mode (Interval or Non- 
stop). 


The display shows: 


Rinse ? 
Press enter 





Thermo Thermo Scientific Barnstead GenPure with xCAD 
TaS Ultrapure water system 


During Rinsing the display shows: 


Rinse 
30 sec. 


12.1.6 Disinfection 


A sixth press on the menu-key gives you the possibility of carrying out disinfection. When 
this menu-prompt is confirmed with Enter, the command Disinfection cartidge. Install one“ 
is shown. After confirming this with Enter, disinfection is carried out. The pump runs for the 
disinfection time that has been set in the OEM-menu. When the half of the disinfection 
time has passed, the rinsing solenoid valve opens and stays open up to the end of the dis- 
infection process. The "New filter cartridge. Install one" message is given when the disin- 
fection process has finished 

On confirming this with Enter, the system returns to the last operating mode. 

During disinfection, the count-down of the disinfection time is shown in the display 


The display shows: 


Disinfection 
Press enter 





After confirnation with Enter, the display shows: 


Disinfection 
cartridge! 
Install one 

Press enter 





During disinfection, the display shows: 


Disinfection 
25 min. 





39 


Thermo Thermo Scientific Barnstead GenPure with xCAD 
ee ee Ultrapure water system 


After disinfection, the display shows: 


New 
filterset 


Press enter 





12.1.7 Display of fault storage: 


A seventh press on the menu-key calls the fault storage. Confirmation with Enter allows 
you to look through the faults in the fault storage. 

Should several faults have occured, then two faults, each with date and time, are shown in 
the display. Appropriate pressing of the arrow-keys brings previous and subsequent faults 
to display. 

A press on the menu-key or the Enter-key returns the display to the last operating mode. 


The display shows: 


Error history 
Press enter 





Show of faults in the fault storage: 


14.03.04 14.30 
Limit value feed 


14.03.04 15.30 
Lim. val.pure w. 





40 


Thermo Thermo Scientific Barnstead GenPure with xCAD 
Ultrapure water system 


12.1.8 Print-out 


A eight press on the menu-key bring you to a menu in which you can hand-request a print- 
out. 


The display shows: 


Print out? 
Press enter 


0000 





12.1.9 Unlocking the system 


A nine press on the menu-key in this menu brings you to the “Code” menu. To prevent un- 
authorized access to the settings in the system control, changes to the settings can only 
be carried out when a correct code from the Assignment Table that follows is entered and 
confirmed with the Enter-key. 

The unlocking remains active for 5 minutes. 

Each access via the code is issued to the printer (RS 232) complete with date, time and 
abbreviated code number (“Code 0001” corresponds to code 150, “Code 0002” to code 
250 etc.). 


The display shows: 


Code 
Press enter 


0000 








Code numbers can be assigned to individual persons in the Assignment Table on 
the following page. 

Please remove this page from the Operating Instructions and store it where it is 
safe from unauthorized viewing. 











41 


Thermo Thermo Scientific Barnstead GenPure with xCAD 


42 


Ultrapure water system 


Assignment Table for code numbers 
that allow the system to be unlocked 


Printer 



































Code no. Person 
output 
150 0001 
250 0002 
350 0003 
450 0004 
550 0005 
650 0006 
750 0007 
850 0008 
| 950 0009 | 


Thermo Thermo Scientific Barnstead GenPure with xCAD 
SIETA Ultrapure water system 


12.2 OEM-Menu: 


Basic settings and limiting values can be changed in this menu. 
To make changes in the OEM menu, the system control must previously be unlocked (see 
section “Menu” Unlocking the system). 


Calling the OEM menu: 

Simultaneous pressing of the Interval-key and the Nonstop-key calls the OEM menu. Fol- 
lowing this, the prompt "OEM menu Press Enter" appears. When this is confirmed with the 
Enter-key, the first menu prompt can be worked on. To simplify changing settings, press 
the UV-key to select the number in a numerical value which you want to change. Now use 
the arrow keys to enter the wanted number from 0 to 9 at that selected position. 

A press on the menu-key takes you to the next menu prompt. 


The OEM menu call display shows: 


OEM Menu 


12.2.1 Setting the maximum operating temperature: 


The maximum temperature which the system is to be allowed to reach can be set in this 
menu. When this temperature is exceeded, the "max. Temperature" fault message is trig- 
gered. 

It is shown in the 4th line of the display. 


Basic setting: 50 °C 
Setting range: 1-50°C 


The display shows: 


OEM Menu 
Max. temp. 


35 °C 





43 


Thermo Thermo Scientific Barnstead GenPure with xCAD 
AAEN EEES Ultrapure water system 


12.2.2 Setting the rinsing time: 


Basic setting: 30 sec. 
Setting range: 10 - 60 sec. 


The display shows: 


OEM Menu 
Rinse time 


30 sec. 





12.2.3 Changing the disinfection time: 
Basic setting: 30 min. 


Setting range: 15 -90 min. 


The display shows: 


OEM Menu 
Disinfect. time 
30 min. 





12.2.4 Setting the interval pump time: 


Basic setting: 5 min. 
Setting range: 1 -30 min. 


The display shows: 


OEM Menu 
Pump interval 


05 min. 





44 


Thermo Thermo Scientific Barnstead GenPure with xCAD 
SETE Ultrapure water system 


12.2.5 Setting the interval rinse time: 


Basic setting: 0.5 sec. 
Setting range: 0.1 - 2 sec. 


The display shows: 


OEM Menu 
Rinse interval 


0.5 sek. 





12.2.6 Setting the real-time clock: 


Basic setting: The actual date 
Setting range: 1 -12 Month, 1 - 31 day, 0 - 24 h, 0 - 60 min. 
The display shows: 


OEM Menu 
Day 30 Month 12 


Year 2008 
Hour 12 Min.30 





12.2.7 Setting the sending interval: 


In this menu, the sending interval is set for transmissions of measured values and fault 
messages to the RS 232 interface. 


Basic setting: 1 hour 
Setting range: 0.5 - 12 hours 


The display shows: 


OEM Menu 
Send interval 


08:00 h : min 





45 


Thermo Thermo Scientific Barnstead GenPure with xCAD 
i rials Ultrapure water system 


12.2.8 Selecting the language: 


Basic setting: English 
Setting range: Englisch, French, German 
The display shows: 


OEM Menu 
Language 


English 





12.2.9 Switching units: 
Basic setting: Resistance MOxcm 


Setting range: Resistance MOxcm, 
Specific electrical resistance MO cm 


The display shows: 


OEM Menu 
uS/cm - MO cm 


MOxcm 





12.2.10 Switching temperature compensation on or off: 


Basic setting: on 
Setting range: on, off 
The display shows: 


OEM Menu 
Temp. comp. 


On 





46 


Thermo Thermo Scientific Barnstead GenPure with xCAD 
Ultrapure water system 


12.2.11 Entering the system type and serial number: 


This menu prompt is for the entry of the system type and the serial number of it. Both en- 
tries are given in the headline of each print-out. Settings are made with the arrow keys. 
The system type and the serial number can also be briefly seen when the on/off-key is 
pressed (on booting). 


The serial number has six characters including a slash and can be entered with the arrow 
keys just as all other entries. 


The display shows: 


OEM Menu 
Genpure UV UF 


S.No.: 9999/08 





12.2.12 Water dispensing via volume control 
As soon as the Nonstop-mode is selected, then a litre volume is shown in line 2 of the dis- 
play. This is the volume of ultra pure water that was last dispensed. 
A single press on the push button on the grip allows the litre volume to be changed within 
the range from 0.01 bis 65.5 litres by means of the arrow-keys. The UV-key can be used 
to position the cursor at a particular number that you wish to change. 
A second press on the push button causes dispensing of the volume of water that has 
been set. The litre volume shown in the display is the actual volume that has been dis- 
pensed. Dispensing stops as soon as the set volume has been reached. 
Dispensing can be stopped at any time by a further press on the press button. 
Manual dispensing can also be made of any volumes within the set volume by two 


presses on the press button to start dispensing and a further press to stop it. 


The display shows: 


Nonstop 5 ppb 
18.2 MOxcm 1.00L 


21.3°C UV/TC 





47 


Thermo Thermo Scientific Barnstead GenPure with xCAD 
eee eee Ultrapure water system 


12.3 Printer output 


Various parameters are recorded by the printer. A differentiation is made between three types of 
messages: 


- Standard messages 
- Code messages 
- Fault messages 


12.3.1 Standard messages: 


Here a record of all measured values is printed out in dependence on the sending interval. 
A complete data set is also printed out during NONSTOP-operation. 
Print-out: 


e.g.: 03.04.08 15:15 

GenPure Standard 
S.No. 3988/08 

Interv. TC on UV off 
LF1= 18.2 MOxcm 
LF2= 10.0 MOxcm 
LF3= 0.000 MOxcm 
Temp.= 16.8 °C 
TOC= 0 ppb 
UV Intens.= 0% 


The standard record includes all measured values. With systems that are not equipped 
with TOC measurement and UV-intensity measurement, 0 is given as measured value! 


12.3.2 Code messages: 


Whenever a code number is entered in the system control and confirmed with the Enter-key, 
then the code input is immediately printed out. 

For code identification refer to the “Assignment Table for code numbers that allow the sys- 
tem to be unlocked”. 


Print-out: 
e.g.: 03.04.08 15:17 
GenPure Standard 


S.No. 3988/08 
Code 0002 


48 


Thermo Thermo Scientific Barnstead GenPure with xCAD 
SETE Ultrapure water system 


12.3.3 Fault messages: 


When a fault message is shown in the display, e.g. for the ultra pure water limiting value, 
then the fault message is printed out at the end of the sending interval. 


Print-out: 
e.g.: 03.04.08 17:09 
GenPure Standard 


S.No. 3988/08 
Lim val.pure w. 


49 


Thermo Thermo Scientific Barnstead GenPure with xCAD 
See S Ultrapure water system 


13. Maintenance 


Regular maintenance of your system maintains the value of it. We recommend that you close a 
service contract with your service company. You then have the certainty of a high operational 
safety and reliability. 


NOTE! 


To ensure that your system will work reliably for a long time, it must be checked, serviced and 
cared for at regular time intervals in accordance with these Operating Instructions. 

For this reason, the Operating Instructions must be readily available to operating and mainte- 
nance staff at all times, and be carefully followed. 


Any maintenance work should become necessary during the validity of the guarantee, this is 
only to be carried out by a service professional which is expressly authorized to do such work. 


The staff who are trained on and operate the system are committed to carry out daily/weekly 
checks. 

The maintenance work specified on the maintenance record supplied with these Operating In- 
structions should to be carried out weekly. 


CAUTION! 


The cleaning and disinfection of your system is performed for reasons of hygiene and has no 
effect on the technical condition of the system. 

The system must only be cleaned and disinfected when algae or slime are detected inside it, or 
at least once yearly. 


the system off at the mains switch to make it currentless and ensuring that it will not be 
inadvertently turned back on. Such work is only to be carried out by qualified electri- 
cians 


Checks or maintenance work on electrical equipment are only to be made after turning 


50 


Thermo Thermo Scientific Barnstead GenPure with xCAD 
SEE Ultrapure water system 


13.1 Maintenance intervals 


Consumable materials are to be replaced at the intervals given in the following Table or when 
there is a drop in performance: 





























Material Flow chart no. | Article no. Interval* 

Filter cartridge F1 09.2005 12 Months 
Sterilefiltercapsule F2 09.1003 12 Months 
Ultrafiltration membrane F3 50133980 24 Months 





*Please keep in mind that the life of your consumables is directly dependent on the quality of 
the feed water and the amount of water used daily. The interval is contingent on the feedwater 
quality so that a shorter one may be necessary. 


51 


Thermo Thermo Scientific Barnstead GenPure with xCAD 
i ili cles Ultrapure water system 


13.2 Replacing the filter cartridge 


(i) 





The filter cartridge is to be replaced when the maximum limiting value 
that you have set for the ultra pure water is exceeded or when the "New 
Filter cartridge" message is shown in the display. 





Proceed as follows to replace the filter cartridge: 


52 


1. 


2. 


10. 


11. 


12. 


13. 


Shut off the supply of feedwater. 


Press the press button on the xCAD for water dispensing until no more water 
emerges and push it again to stop dispensing. 


Switch the system off. 
Remove the cartridge cover. 


Take off the quick-conects from the feed and ultra pure water sides of the filter car- 
tridge and close the connectors with the stoppers you have saved. 


Draw the spent filter cartridge out from the guide and push the new filter cartridge in. 
Remove the stoppers from the new filter cartridge and save them. 

Fit the quick connects to the filter cartridge so that they audibly click into position. 
Replace the cartridge cover. 

Open the supply of feedwater. 

Switch the system on. 

Dispense and discard at least 5 litres of water. 

To set the operating time counter of the filter cartridge back to zero, enter the serial 
number that is given on the filter cartridge as is described in "Filter cartridge operating 


time counter" section. 


Caution: To do this, the system must be unlocked, as is decribed in the "Unlocking 
the system" section. 





Switch the system to the "Rinsing" operating mode to vent air out of it. 











Thermo Thermo Scientific Barnstead GenPure with xCAD 
SETAS Ultrapure water system 


13.3 Disinfection 





Disinfection should be regularly carried out, at the latest when the filter 
cartridge is replaced. 











To disinfect the system you require a Disinfection cartridge, article no. 09.2201. 


Use cleaning solutions as follows: 
MICRO-Chlor Granulate, 1 box, article no. 09.2202 (Europe only) 
Cleaning Solution, 1 syringe, article no. CMX 25 (US-market only). 





with Micro-Chlor disinfectant to avoid possible health hazards! 





À Please observe the information given in the safety data sheet supplied 


Proceed as follows to disinfect your system: 
1. Switch the GenPure system off. 


2. Shut off the supply of feedwater to the system and open the dispensing valve so that 
pressure in the system is completely released. 


3. Remove the filter cartridge (as under "Changing the filter cartridge" in the Operating 
Instructions for the system). 


4. _ Unscrew the stopper from the disinfectant cartridge, fill the cartridge with water then 
empty the contents of a box respectively a syringe of the cleaning solution into the wa- 
ter. 

Important! For effective disinfection the cartridge must be completely filled 
with water. 

5. | Screw the stopper back on the disinfectant cartridge and connect the cartridge in the 
system (as under "Changing the filter cartridge" in the Operating Instructions for the 
system). 

6.  Re-open the feedwater supply. 


7. | Switch the system on and select the "Disinfection" prompt in the menu. The disinfec- 
tion programme is finished after approx. 30 minutes. 


8. Switch the system off. 
9. Shut off the supply of feedwater to the system. 


10. Remove the disinfectant cartridge (as under "Changing the filter cartridge" in the Op- 
erating Instructions for the system). 


11. | Connect the new filter cartridge in the system (as under "Changing the filter cartridge" 
in the Operating Instructions for the system). 


Before dispensing water from the system, let water run out for approx. 
15 minutes. The system is then ready for use. 


53 





Thermo Thermo Scientific Barnstead GenPure with xCAD 
Ultrapure water system 

















=>) lil 
_ = 
5| T = 
esintektion al [| E | = 
kart che j — 
Th IE SS ¥ 
esintection N | 
Cartridge | | 
|i | 
| A 
(a) 









































Well F 




















54 


Thermo Thermo Scientific Barnstead GenPure with xCAD 
SEE Ultrapure water system 


14. Waste disposal 


When the packaging is no longer needed it can be disposed of as household waste. 

Systems are in conformity with EEC Guideline 2002/95/EC 

The system is not to be thrown away as household waste but must be properly disposed of. It 
can be returned to the manufacturer for safe disposal according to EEC Guideline 2002/96/EC. 
We therefore request our customers in Germany and other member States in the European 
Economic Area to contact our local service centre or our headquarters: 

Thermo Electron LED GmbH 

Stockland 3 

D-56412 Niederelbert, Germany 

WEEE-Reg.-no.: DE 12471402 


In countries outside of the European Economic Area, please contact your local authorities or 
waste disposal company. 


55 


15. Trouble shooting 


The system does not start 


Thermo Scientific Barnstead GenPure with xCAD 


No supply of power 


Ultrapure water system 


Provide power 





Dispensing not possible 


Feedwater tap is closed 
Feedwater and rinse water 
connections mixed up 
Feedwater pressure 

< 0.1 bar 


Open feedwater tap 
Correct the connections 


Increase feedwater pressure 





Resistance < 18.2 MOxcm 


System has a UF-module 
(higher conductivity 
possible) 

lon exchange capacity 
exhausted 


Replace filter cartridge 
with a new one 





System control no longer 
reacts 


Improper operation 


Unplug the mains plug for 
5 seconds 





Water flows out 


Leaky hose connection 


Feedwater pressure > 6 bar 


Check and seal the hose 
connection 
Install a pressure reducer 





Dispensed water flow rate is 
too low 


UF-Module blocked 
Pre-pressure too low 
Internal pressure too low 


Replace UF-module 
Increase pre-pressure 
Readjust pressure reducer 





Wrong time or date 


Time zone 
Summer/winter time 


Reset time and date 





Wrong language 


Wrong language set 


Correct language setting 





Fault message: 
"Limit value feed" 


56 





Feedwater conductivity 
too high 
Limiting value set too low 





Check the pretreatment 


Check and suit the limiting 
value 





Fault message: 
"Lim. val.pure w." 


Thermo Scientific Barnstead GenPure with xCAD 
Ultrapure water system 


Filter cartridge exhausted 
Limiting value set too low 
System has a UF-module 
(higher conductivity 
possible) 


Replace with new filter 
cartridge (art.-no.: 09.2005) 
Check and suit the limiting 
value 





Fault message: 
"UV-time" 


UV-Lamp operating time 
has been exceeded 


Replace the UV-lamp 
(art.-no. 09.2002) and reset 
the operating time counter 
(see 12.1.4) 





Fault message: 
"UV-intensity" 


UV-Lamp intensity no 
longer sufficient 
UV-Sensor dirtied 
Limiting value set too low 


Replace and measure in 
a new UV-lamp 

Clean UV-sensor 

Check and suit the limiting 
value 





Fault message: 
"max. Temperature" 


The temperature in the 
system is too high 

Interval pump time too long 
Limiting value set too low 


Feedwater temperature 
too high 


Reduce temperature by 
running water off 

Reduce interval pump time 
Check and suit the limiting 
value 

Reduce feedwater 
temperature 





Fault message: 
"Measuring cell LF1" 


Fault message: 
"Measuring cell LF2" 


Measuring cell cable break 
System control defect 
Conductivity of ultra pure 
water ouside measuring 
range 


Measuring cell cable break 
System control defect 
Feedwater conductivity 
ouside measuring range 


Replace measuring cell 
Replace system control 
see "Conductivity 

< 18.2 MOxcm" 


Replace measuring cell 
Replace system control 
see “Limit value. feed“ 





Fault message: 
"Measuring cell LF3" 


Fault message: 


"Measuring cell Temp. 


Measuring cell cable break 
System control defect 


Measuring cell cable break 
System control defect 


Replace measuring cell 
Replace system control 


Replace measuring cell 
Replace system control 





Fault message: 
"New Filter cartridge" 





Filter cartridge operating 
time has expired 





Replace with new filter 
cartridge (art.-no.: 09.2005) 





57 


Thermo Scientific Barnstead GenPure with xCAD 
Ultrapure water system 


16. Replacement parts and consumables 


16.1 GenPure 











































































































































































































































































































































































































































































































































































































































































































































































































10 f - © 
OR | 12 
T i 11 J 
: | There | i 14 
LU | | — 3 
| 15 T 
| 
= ME 13 
D s 
| 5 
Hh) 4 
We) 6 L 
I a | & 
l 
l TM A ai 9 É 
No. | R+I no. | Article desigantion Article no. 
1 A-F1 Filter cartridge 09.2005 
2 A-UV1 Replacement UV-lamp 09.2002 
3 A-P1 Pressure booster pump 19.0050 
4 |A-F3 Ultrafiltration module (optional) 50133980 
5 |V4 Rinsing solenoid valve 50131190 
6 V1 Pressure reducer 50133985 
7 FIS400 | Flowmeter 15.0100 
8 QIA300 | Conductivity measuring cell, feedwater 16.0126 
9 Adapter board 16.0364 
10 Microprocessor-system control, interface 50131346 
11 Booster for UV-lamp (optional) 22.0088 
12 |QIA303 | Conductivity measuring cell, ultra pure water 50133992 
TIA500 | Conductivity measuring cell, temperature sensor 
13 | V2 Check valve 15.0009 
14 |QIA301 | UV-Intensity sensor (optional) 16.0222 
15 |V5 Check valve 15.0019 
16 Fuseholder for glas tube fuse, 5x20mm 50133979 
Glas tube fuse, 5x20mm, 3.15 A, slow 50131758 
Table top power pack, 24V DC (not showns) 50134196 

















Ui) 


58 








We ask for your understanding that the guarantee that we give becomes 
invalidated when replacement parts, accessories or consumables from 
other manufacturers are used in or for the system, as we have no influ- 
ence on their appropriateness, composition or quality. 





























Thermo Scientific Barnstead GenPure with xCAD 


Ultrapure water system 














16.2 xCAD 
19 
18 
| 
| 
| 
| 
| 
| 
Z anal 
I? ( | 
e= 
No. | R+I no. | Article designation Article no. 
17 Microprocessor-system control 26.0025 
18 |A-F2 Sterile filter capsule, 0,2 um 09.1003 
19 Press button 16.0370 
20 |V3 Solenoid valve 15.0101 
Extension cable SUB-D, 25 pin, GenPure/xCAD 16.0375 

















59 


Thermo Thermo Scientific Barnstead GenPure with xCAD 
eee Ultrapure water system 


17. Consumables and accessories 





























Article designation Article no. 

lon exchanger system DI 1500 02.1500 
Disinfection agent, MICRO-Chlor (pack of 12 cans, Europe only) 09.2202 
Cleaning Solution, 1 syringe (US-market only) CMX25 
Disinfection cartridge 09.2201 
Printer 09.2207 
Qualification documents IQ, OQ IOQDOCE50133918 








The address to contact when your system requires service: 


Overview of Thermo Scientific International Sales Organization 
Postal address USA: 

Thermo Scientific 

275 Aiken Road 

Asheville, NC 28804 


USA 
Enquiries from USA/ Canada 
Sales: +1 866 984 3766 
Service: +1 800 438 4851 
Enquiries from Latin America 
Sales: +1 866 984 3766 
Service: +1 866 984 3766 
Enquiries from Asia 
China 
Sales: +86 10 8419 3588 
Service: Toll free 8008105118 
Support Mobile 4006505118 or +86 10 8419 3588 
India 
Sales: +91 22 6716 2200 
Service: Toll free 1 800 22 8374 or +91 22 6716 2200 
Japan 
Sales: +81 45 453 9220 
Service: +81 45 453 9224 
Enquiries from the Rest of Asia/ Australia/ New Zealand 
Sales: +852 2885 4613 
Service: +65 6872 9720 
Enquiries from Countries not listed / Rest of EMEA 
Sales: +49 6184 90 6940 or +33 2 2803 2000 
Service: +49 6184 90 6940 
Enquiries from Europe 
Austria 
Sales: +43 1801 40 0 
Service: +43 1801 40 0 


60 


Thermo Thermo Scientific Barnstead GenPure with xCAD 


Belgium 

Sales: +32 53 73 4241 
Service: +32 53 73 4241 
Finland/ Nordic/ Baltic countries 
Sales: +358 9 329 100 
Service: +358 9 329 100 
France 

Sales: +33 2 2803 2180 
Service: +33 825 800 119 
Germany: 


Postal Address Germany: 
Thermo Electron LED GmbH 
Robert-Bosch-Strake 1 

D - 63505 Langenselbold 


Phone 

Sales: Toll free 0800 1 536 376 
or +49 6184 90 6940 

Service: Toll free 0800 1 112110 
or +49 6184 90 6940 

E-Mail: [email protected] 

Italy 

Sales: +39 02 95059 341 

Service: +39 02 95059 250 

Netherlands 

Sales: +31 76 579 5555 

Service: +31 76 579 5639 

Russa/ CIS 

Sales: +7 812 703 4215 

Service: +7 812 703 4215 

Spain/ Portugal 

Sales: +34 93 223 0918 

Service: +34 93 223 0918 

Switzerland 

Sales: +41 44 454 1212 

Service: +41 44 454 1212 

UK/ Ireland 

Service: +44 870 609 9203 

Sales: +44 870 609 9203 


Ultrapure water system 


61 


Thermo Thermo Scientific Barnstead GenPure with xCAD 
Se Ultrapure water system 





























































































































f i i 


aannnnni 




















3 { oO l 


baL A 


8 ‘(MMM muan | 





























+24 
+48 

















+ w 


































































































































































































































































































































































































































































































































































































o ‘a 
ls : | 
= 
a5) = SE . 
a l r 
zł 
E = 
s : : 
PIO l I 
© Sje © v slels p ai 
| 5 
be ; a 
i Be fle]: | = altel I | 
a o ae 
i HIRI PEL: 
3 ai] | g 
















































































S 
o 

| 
‘mera 
| oS 











62 


Thermo Thermo Scientific Barnstead GenPure with xCAD 
AAEE Ultrapure water system 


19. Maintenance record 


(Correct keeping of this maintenance record is a condition of the guarantee) 


Customer address: Location: 


System type: 








Serial no.: 








Year made: 


Feedwater | Ultra pure Ultra pure UV- UV-lamp 
resistance | water water intensity operating 
resistance | volume flow time, 


[MOxcm] [MOxcm] [L/h] [%] [h] 












































Last change |Last Last change | Remarks Signature 
of filter cleaning / of 
cartridge disinfection | pretreatment 



































Any false entry is considered to constitute a falsification of documents. 


The following points are to be observed to safeguard the quality of the system: 


> 1x Weekly, enter measured values. 


The pretreatment must be checked and replaced at regular intervals to ensure the 
optimal ultra pure water quality. 


63 


Thermo Thermo Scientific Barnstead GenPure with xCAD 
Se re Ultrapure water system 


Contact I nformation Thermo Scientific 


North America: 
USA/Canada +1 866 984 3766 (866-9-THERMO) 


Europe: 

Austria +43 1 801 40 0, Belgium +32 53 73 42 41, France +33 2 2803 2180, 
Germany national toll free 08001-536 376, 

Germany international +49 6184 90 6940, Italy +39 02 95059 448, 
Netherlands +31 76 579 55 55, Nordic/Baltic/CIS countries +358 9 329 10200, 
Russia +7 812 703 42 15, Spain/Portugal +34 93 223 09 18, 

Switzerland +41 44 454 12 12, UK/Ireland +44 870 609 9203 


Asia: 

Australia +61 39757 4300, China +86 21 6865 4588 or +86 10 8419 3588, 
India toll free 1800 22 8374, India +91 22 6716 2200, 

Japan +81 45 453 9220, New Zealand +64 9 980 6700, 

Other Asian countries +852 2885 4613 


Countries not listed: +49 6184 90 6940 


© 2011 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fish- 
er Scientific Inc. and its subsidiaries. Specifications, terms and pricing 

are subject to change. Not all products are available in all countries. Please consult your local sales 
representative for details. 


64