43 2 BLA 125742 0 08 04 2021 Telecon Information Reques

Pfizer Documents (PHMPT/FDA)

Pfizer Bla Submission

Pfizer 16 Plus Documents

4

Document text

From:  Smith, Michael (CBER)  
Sent:  Wednesday, August 4, 2021 5:27 PM  
To: Harkins Tull, Elisa <[email protected]>; Aghajani Memar, Neda 
<[email protected]>; Devlin, Carmel M <[email protected]>; Rohlfing, Paul 
<[email protected]>  
Cc: Naik, Ramachandra <[email protected]>; Gottscha lk, Laura 
<[email protected]> 
Subject:  RE: [EXTERNAL] STN 125742.0: Comments on LRP template submitted to STN 125742/0.14 on 
7/20 
 
Elisa,  
 
I apologize for not including the attachment, please see the attached document.  
 
 Regards,  
 Mike  
 
 From:  Harkins Tull, Elisa < [email protected]
>  
Sent:  Wednesday, August 4, 2021 5:19 PM  
To: Smith, Michael (CBER) < [email protected] > 
Cc: Agha jani Memar, Neda < [email protected] >; Devlin, Carmel M 
<[email protected]>; Rohlfing, Paul <[email protected] >; Naik, Ramachandra 
<[email protected] >; Gottschalk, Laura < [email protected] > 
Subject:  Re: [EXTERNAL ] STN 125742.0: Comments on LRP template submitted to STN 125742/0.14 on 
7/20 
 
CAUTION:  This email originated from outside of the organization. Do not click links or open attachments unless you 
recognize the sender and know the content is safe.  
 Hi again M ike, 
 
Your message refers to attachments but there are not attachments to the mail. Can you please send 
these when you have a moment? Or alternatively please clarify.  
 Best regards, Elisa  
On Aug 4, 2021, at 4:48 PM, Smith, Michael (CBER) <[email protected]
> wrote:  
  
Elisa,  
FDA-CBER-2021-5683-1024820
  
The review team has the below comments on the lot release protocol (LRP) 
template that was submitted to BLA 125742/0.14 on July 20, 2021.   The review 
team has requested that responses are submitted to the BLA by Wednesday, 
August 11, 2021.   We anticipate responding to Paul’s clarification questions on 
the LRP and samples that were e- mailed us yesterday in the near future.  
  
Throughout document  
Please correct the cc: line to STN 125742 -0/2229/FC  
  
Page 1 of 6  
Please replace with the attached example (Attachment 1 – Electronic 
Protocol Page 1)  
  
Please make sure that the electronic Protocol Number at the bottom of Page 1 matches the number on the eLRP Signature letter.
 
  
Information after the Date of Manufacturing line and before the Storage Temperature is optional.
 
  
Note: When submitting LRPs electronically, please use a letter formatted per Attachment 2 – eLRP Signature letter.   Place this letter before the 
electronic protocol.
 
  
Page 2 of 6  
Components table  
• Please add component description for the LNP  
  
Page 3 of 6  
• RNA Encapsulation and RNA content.  
Please use the RNA content template (Attachment 3) to report the results  
  
• Lipid analysis  
Please provide full  of the lipid components for the sample, 
. Insert the  
for the reference standard above the sample lines and label all the  
(b) (4)
(b) (4)
(b) (4)
(b) (4)
FDA-CBER-2021-5683-1024821
The page on which the  are provided should include a table 
with test date, specifications and results for each lipid content and identity.  
  
Page 4 of 6  
• Table 1 (Continued) Filled Vaccine Quality Control Tests  
Please remove the abbreviations for  
 
 These tests are not performed for the filled vaccine.  
  
• Identity of encoded RNA sequence  
Please use the identity test template (Attachment 4) to report the drug 
product test results  
  
• In vitro expression  
Please use the in vitro expression test template (Attachment 5) to report the drug product test results
 
  
• RNA integrity  
Please provide  that are  
RNA  for each sample replicate and insert the reference 
standard  above this line so that the sample lines are not 
obscured. Please include clear labels for product and  
 that were includ ed in the 
analysis of each  Include the name of the test method, the 
specification, date of test and the result on the same page as the 
 
• Bacterial endotoxin  
Please use the Limulus Amebocyte Lysate Test template (Attachment 6) to repo rt the drug product endotoxin results. 
 
  
• Sterility  
Method: Please add  Method to  
  
Container: Please change 20 mL to  
  
  
Regards,  
(b) (4)
(b) (4)
(b) (4)
(b) (4)
(b) (4)
(b) (4)
(b) (4)
(b) (4)
(b) (4)
(b) (4)
(b) (4)
(b) (4)
(b) (4)
(b) (4)
FDA-CBER-2021-5683-1024822
  
Mike  
  
- Please confirm receipt of this e -mail and let us know if you have any 
questions.  
  
Mike Smith, Ph.D.  
Captain, USPHS  
 
Senior Regulatory Review Officer  
Food and Drug Administration  
Center for Biologics Evaluation & Research  
Office of Vaccines Research & Review  
Division of Vaccines and Related P roducts Applications  
Tel: 301- 796-2640  
[email protected]  
<image001.png>  
 
<image002.jpg>  
   
<image003.jpg>  
   
<image004.jpg>  
   
<image005.jpg>  
   
<image006.jpg>  
 
  
  
THIS MESSAGE IS INTENDED ONLY FOR THE USE OF THE PARTY TO WHOM IT IS 
ADDRESSED AND MAY CONTAIN INFORMATION THAT IS PRIVILEGED, CONFIDENTIAL, AND 
PROTECTED FROM DISCLOSURE UNDER LAW.  If you are not the addressee, or a 
person authorized to deliver the document to the addressee, you are hereby notified that any review, disclosure, dissemination, copying, or other action based on the content of this communication is not authorized.  If you have 
received this document in error, please immediately notify the sender immediately by e- mail or phone.
 
  
  
  
 
FDA-CBER-2021-5683-1024823