125742 S1 M5 5314 vr mvr 10083

Pfizer Documents (PHMPT/FDA)

Pfizer Bla Submission

Pfizer 16 Plus Documents

22

Document text

PF-07302048: Method Validation ofthe SARS -CoV-2 mNeonGreen Virus Microneutralization Assay
VR-MVR-10083, Ver. 1.0
PFIZER CONFIDENTIAL
Page 1Title:Method Validation of the SARS-CoV-2 mNeonGreen Virus Microneutralization 
Assay
Study Number: N/A
Parent Compound Number(s): PF-07302048
Alternative Compound Identifiers: N/A
Pfizer Vaccine Research and Development
401 N. Middletown Rd.
Pearl River, NY
090177e196376ca2\Approved\Approved On: 09-Feb-2021 02:38 (GMT)
FDA-CBER-2021-5683-0779463
PF-07302048: Method Validation ofthe SARS -CoV-2 mNeonGreen Virus Microneutralization Assay
VR-MVR-10083, Ver. 1.0
PFIZER CONFIDENTIAL
Page 2Title:Method Validation of the SARS -CoV-2 mNeonGreen Virus Microneutralization 
Assay
PRINCIPAL INVESTIGATORS: , Director
Ingrid L. Scully, Senior Director
, Director
CONTRIBUTING SCIENTIST S: , Senior Associate Scientist
Tyler Garretson, Principal Scientist
Kyle Jacobs, Associate Scientist
, Associate Scientist
, Senior Associate Scientist
George Phel an, Associate Scientist
, Scientist
, Senior Associate Scientist
, Associate Scientist
PREPARED BY:
APPROVED BY:
Tyler Garretson
Principal Scientist , Clinical and Diagnostic Assay Deve lopment
Ingrid L. Scully
Senior Director , Clinical and Diagnostic Assay  Development
Charles Tan
Senior Director, Earl y Clinical Development Biostatistics
David Cooper 
Executive Director, High
-throughput Clinical Immunoassay s & Diagnostics
Michael Pride
Executive DirectorBiology, Clinical and Diagnostic Assay  Development
090177e196376ca2\Approved\Approved On: 09-Feb-2021 02:38 (GMT)
(b) (6)
(b) (6)
(b) (6)
(b) (6)
(b) (6)
(b) (6)
(b) (6)
(b) (6)
(b) (6)
(b) (6)
(b) (6), (b) (4)
FDA-CBER-2021-5683-0779464
PF-07302048: Method Validation ofthe SARS -CoV-2 mNeonGreen Virus Microneutralization Assay
VR-MVR-10083, Ver. 1.0
PFIZER CONFIDENTIAL
Page 3Title:Method Validation of the SARS -CoV-2 mNeonGreen Virus Microneutralization 
Assay
SYNOPSIS
This reportdescribesthe validation of the SARS-CoV-2 mNeonGreen Virus
Microneutralization Assay  (SARS-CoV-2 mNG NT) used for the detection of serum 
antibodies capable of neutralizing SARS- CoV-2. The assay  readout of 50% virus 
neutralization titer was validated per protocol VR-MVP-10074.
Based on dilutional linearity  and assay  precision, the lower limit of quantitation (LLOQ) 
was determined to b e a titer equal to 41 and the upper limit of quantitation (ULOQ) was 
determined to be a titer equal to 3,187.Samples with titers greater than the ULOQ may
be pre-diluted before testing to yield titers within thevalidated assay  range.
Intermediate precis ion assessment demonstrated overall assay  variability  of 26.5% RSD.
Serum samples are run in replicate in the SARS -CoV-2 mNG NT.  Based on the assay  
performance, it was determined that the replicate titer ratio for a sample must be less than 
or equal to 2.55 (sample extravariability  rule).
090177e196376ca2\Approved\Approved On: 09-Feb-2021 02:38 (GMT)
FDA-CBER-2021-5683-0779465
PF-07302048: Method Validation ofthe SARS -CoV-2 mNeonGreen Virus Microneutralization Assay
VR-MVR-10083, Ver. 1.0
PFIZER CONFIDENTIAL
Page 4TABLE OF CONTENTS
SYNOPSI S................................ ................................ ................................ ................................ .3
LIST OF TABLES ................................ ................................ ................................ ..................... 5
LIST OF FIGURES ................................ ................................ ................................ ................... 5
1. OBJECTI VES................................ ................................ ................................ ........................ 6
2. INTRODUCTION ................................ ................................ ................................ ................. 6
3. GLOSSARY ................................ ................................ ................................ .......................... 7
4. MATERIAL S AND MET HODS................................ ................................ ........................... 7
4.1. Critical Reagents ................................ ................................ ................................ .......7
4.2. Quality  Control Samples ................................ ................................ ........................... 8
4.3. Negative Serum Diluent................................ ................................ ............................ 8
4.4. Validation Serum Sample Panels................................ ................................ ..............8
5. EXPERI MENTAL DESI GN................................ ................................ ............................... 10
6. STATI STICAL ANALY SIS................................ ................................ ............................... 10
6.1. Dilutional L inearity................................ ................................ ................................ .11
6.2. Precision................................ ................................ ................................ .................. 11
6.3. Limits of Quantitation ................................ ................................ ............................. 12
6.4. Assay  Intermediate Precision ................................ ................................ .................. 12
6.5. Limit of Detection ................................ ................................ ................................ ...13
6.6. Extravariability  of Replicates................................ ................................ .................. 13
7. RESUL TS AND DI SCUSSION................................ ................................ .......................... 13
7.1. Dilutional L inearity................................ ................................ ................................ .13
7.2. Precision ................................ ................................ ................................ .................. 14
7.3. Limits of Quantitation ................................ ................................ ............................. 15
7.4. Assay  Intermediate Precision ................................ ................................ .................. 16
7.5. Limit of Detection ................................ ................................ ................................ ...16
7.6. Extravariability  of Replicates................................ ................................ .................. 16
8. CONCLUSION ................................ ................................ ................................ .................... 17
9. DEVIATIONS ................................ ................................ ................................ ..................... 17
10. REFERENCES ................................ ................................ ................................ .................. 17
11. SUPPORTIVE TABLE S................................ ................................ ................................ ...19
090177e196376ca2\Approved\Approved On: 09-Feb-2021 02:38 (GMT)
FDA-CBER-2021-5683-0779466
PF-07302048: Method Validation ofthe SARS -CoV-2 mNeonGreen Virus Microneutralization Assay
VR-MVR-10083, Ver. 1.0
PFIZER CONFIDENTIAL
Page 511.1. Precision by  Individual Sample ................................ ................................ .............19
12. SUPPORTIVE FIGUR ES................................ ................................ ................................ ..20
12.1. Assa y Design Schematic ................................ ................................ ....................... 20
12.2. Dilutional L inearity Plot by  Individual Sample ................................ .................... 21
LIST OF TABLES
Table1. Terms and Definitions................................ ................................ ................ 7
Table2. Critical Reagents for the SARS- CoV-2 mNG NT Validation .................... 8
Table3. Quality Control Sample Specification Limits.............................................8
Table4. Validation Serum Sample Panels ................................ ................................ 9
Table5. Example Validation Run Schedule ................................ ........................... 10
Table6. Final Assay  Range................................ ................................ .................... 15
Table7.  of Samples Within the Quantitation Range................................ .....16
Table8. Assay Performance Near the L OD................................ ........................... 16
LIST OF FIGURES
Figure1. Dilutional L inearity Results Plot ................................ .............................. 14
Figure2. Precision Results Plot ................................ ................................ ...............15
090177e196376ca2\Approved\Approved On: 09-Feb-2021 02:38 (GMT)
(b) (4)
FDA-CBER-2021-5683-0779467
PF-07302048: Method Validation ofthe SARS -CoV-2 mNeonGreen Virus Microneutralization Assay
VR-MVR-10083, Ver. 1.0
PFIZER CONFIDENTIAL
Page 6Title:Validation Report for the SARS -CoV-2 mNeonGreen Virus Microneutralization 
Assay
Study Number: N/A
Functional Area: Vaccine Research and Development
Test Facility:Hackensack Meridi an Health 
Nutley, NJ 07110
Study/Testing Initiation Date:01-Dec-2020
Study/Testing Completion Date: 20-Jan-2021
1.OBJECTIVES
This report describe sthe validation results for the SARS-CoV-2 mNeonGreen Virus
Microneutralization Assay  (SARS-CoV-2 mNG NT) used for the detection of serum
antibodies capable of neutralizing SARS- CoV-2.  The assay  readout of 50% virus 
neutralization titer was validated per protocol VR-MVP-10074.1  This validation provides
documented evidence that the SARS-CoV-2 mNG NT is suitable for its intended use when 
performed in accordance with standard operating procedures b y qualified personnel .
2.INTRODUCTION
The SARS- CoV-2 mNG NT is a biofunctional assay  that measures neutralizing antibodies 
against SARS -CoV-2.  The SARS- CoV-2 mNG virus is derived from the USA_WA1/2020 
strain that had been rescued by  reverse genetics and engineered to contain a mNeonGreen 
(mNG) reporter gene in open reading frame 7 of the viral ge nome that produces green 
fluorescence upon productive infection of cells.2This reporter virus generates similar plaque 
morphologies and indistinguishable growth curves from wild- type virus.3
This assay  is described in the test method VR-TM-10298.4  Briefly,serially diluted test 
serum samples are mixed with SARS -CoV-2 mNG virus in a 96 -well plate to allow 
virus-specific antibodies to bind to the virus.  This serum-virus mixture is then transferred 
onto aVero cell monolayer and incubated overnight to allow f or infection by  non-neutralized 
virus.  Productive viral infection is detected b y enumerating green -fluorescent viral foci 
using a cell -imaging reader.  The total number of cells per well is calculated by  enumerating
Vero cell nuclei stained blue with Hoechst 33342.  An infection ratio is then calculated for 
each well, whereb y the total number of virus infected (green) cells is divided by the total 
number of cells present (blue nuclei).  A sample titer is definedas the reciprocal serum 
dilution at which a specific percentage of the virus is neutralized, eg, 50%, 80% or 90% 
(termed “Titer Determining Value”, TDV) .
090177e196376ca2\Approved\Approved On: 09-Feb-2021 02:38 (GMT)
FDA-CBER-2021-5683-0779468
PF-07302048: Method Validation ofthe SARS -CoV-2 mNeonGreen Virus Microneutralization Assay
VR-MVR-10083, Ver. 1.0
PFIZER CONFIDENTIAL
Page 73.GLOSSARY
Table1.Terms and Definitions
Term Definition
Assay range Range of neutralization titers that can be measured in the assay with acceptable 
dilutional linearity and precision.
COVID-19 Coronavirus Disease 2019
DL Dilutional linearity
GMT Geometric Mean Titer
LLOQ Lower Limit of Quantitation
LOD Limit of Detection
MDP Master Dilution Plate
NHS_dep Normal Human Serum -Antibody Depleted
NT Microneutralization Assay
QCS Quality Control Sample
Replicate An independent determination of an assay result ; the geometric mean of tw o replicate 
titers is the reportable result for a sam ple tested in the SARS -CoV-2 mNG NT.
RSD Relative Standard Deviation
SARS-CoV-2 Severe acute respiratory syndrome coronavirus 2; the etiologic agent of COVID -19
SARS-CoV-2 mNG NT 96-well manual microneutralization assay for the detection of functional antibodies to 
SARS-CoV-2 using the mNeonGreen reporter virus.
SAS Programming language and integrated software solution -set proprietary to SAS ®(the 
company) that enables the coding of the various tasks associated w ith handling 
datasets.
SOP Standard Operating Procedure
TDV Titer Determining Value; the threshold value in percent of measured viral green 
particle counts that is used to report sample titers.  Titers may be reported at 50%, 
 90% TDV.
Titer Inverse of serum dilution required to neutralize a specific percentage of the input 
virus (eg 50% TDV).
TM Test Method
ULOQ Upper Limit of Quantitation
Vero African green monkey kidney epithelial cell line
4.MATERIALS AND METHOD S
General material supplies, reagents and equipment used are listed in SOPs VR-TM-10298,4
VR-SOP-LC-11299,5VR-SOP-LC-11294,6and VR-SOP-LC-11287.7
4.1.Critical Reagents
Critical reagents used for the validation assay s are listed in Table2.
090177e196376ca2\Approved\Approved On: 09-Feb-2021 02:38 (GMT)
(b) (4)
(b) (4)
FDA-CBER-2021-5683-0779469
PF-07302048: Method Validation ofthe SARS -CoV-2 mNeonGreen Virus Microneutralization Assay
VR-MVR-10083, Ver. 1.0
PFIZER CONFIDENTIAL
Page 8Table2.Critical Reagents for the SARS-CoV-2 mNG NT Validation
Reagent Name/Lot#
VirusStock SARS-CoV-2-mNG Lot 5
P2 Vero E6
Cell Line Vero cells; ATCC CCL81, 
used from passage 134 to 155
4.2.Quality Control Samples
Quality control samples (QCS) were generated from pools of  
  QCS lower and upper titer limits 
were established prior to the start of validation as docum ented in VR -RGR-RQ-107818and
arelisted in Table3.
Table3.Quality Control Sample Specification Limits
QCS Lower Specification Limit Upper Specification Limit
4.3.Negative Serum Diluent
 
 
was used as a negative 
serum diluent for dilutional linearity  testing during this validation .
4.4.Validation Serum Sample Panels
Panels comprised of serum samples from either  
 
 were selected to be used in the evaluation of assay  dilutional linearity  
and precision.  Convalescent serum was obtained from  
  Negative samples 
were collected from donors in 2013 (pre- COVID-19).  All sera were heat- inactivated for 
30min at 56°C prior to testing.
090177e196376ca2\Approved\Approved On: 09-Feb-2021 02:38 (GMT)
 
(b) (4)
(b) (4)
(b) (4)
(b) (4)
(b) (4)
(b) (4)
(b) (4)
(b) (4)
(b) (4)
(b) (4)
FDA-CBER-2021-5683-0779470
PF-07302048: Method Validation ofthe SARS -CoV-2 mNeonGreen Virus Microneutralization Assay
VR-MVR-10083, Ver. 1.0
PFIZER CONFIDENTIAL
Page 9Table4.Validation Serum Sample Panels
Intended Purpose Sample ID Serum Source
Dilutional Linearity Panel
Precision Panel
LOD Assessment
090177e196376ca2\Approved\Approved On: 09-Feb-2021 02:38 (GMT)
(b) (4)
FDA-CBER-2021-5683-0779471
PF-07302048: Method Validation ofthe SARS -CoV-2 mNeonGreen Virus Microneutralization Assay
VR-MVR-10083, Ver. 1.0
PFIZER CONFIDENTIAL
Page 105. EXPERIMENTAL DESIGN
Validation of the SARS-CoV-2 mNG NT was performed as described in the validation 
protocol, VR -MVP-10074.1  The assay readout of 50% virus neutralization titer(50% TDV) 
was validated;  90% neutralization datasets were collected for exploratory  purposes 
only (VR-VTR-107429).  Dilutional linearity  and precision were evaluated in separate assay  
runs according to the schedule shown in Tab le5. The factorial design includes  assay runs, 
completed on  different day s, utilizing  different sets of sample master dilution plate 
(MDP) preparations, performed by  different analy sts inorder to quantify the contribution 
of these factors to the intermediate precision.  Sample MDPs are prepared in deep -well 
96-well plates and each are of sufficient volume to stamp out sample assay  plates on each of 
the assay days in the experimental des ign (refer to Supportive Figure12.1for the assay  
design schematic).
For the dilutional linearity assessment,  serum samples were tested  
independentl y diluted 
For the precision assessment, serum samples were tested (n= ) and 
analyzed along with the  dilutional linearity  samples (n=8).
For theevaluation of the assay  near the LOD,  presumed negative serum 
samples were tested .
The assay  schedule for validation testing is shown in Table5.
Table5.Example Validation Run Schedule
Assay MDP SetaDay Analyst
6.STATISTICAL ANALYSIS
All analyses of titer data described in the statistical analy sis sections are based on the 
reportable geometric mean titer (GMT).  The reportable titer is the  
  Unless indicated otherwise, the word titer will be 
understood to be reportable GMTat 50%TDV.  All ranges a nd/or limits determined based 
on the statistical analy ses will be applied for all non-clinical and clinical testing moving 
forward.
090177e196376ca2\Approved\Approved On: 09-Feb-2021 02:38 (GMT)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
FDA-CBER-2021-5683-0779472
PF-07302048: Method Validation ofthe SARS -CoV-2 mNeonGreen Virus Microneutralization Assay
VR-MVR-10083, Ver. 1.0
PFIZER CONFIDENTIAL
Page 11All statistical analy ses, including , were performed 
using SAS ®version 9.4.
6.1.Dilutional Linearity
Each dilutional linearity  sample wastested at dilutions  and at dilutions of 
 in assay runs.  For each assay run, a 
dilution-adjusted titerwascalculated for each of the dilutions
.  
10
6.2.Precision
Sample precision refers to the closeness of 2or more measurements of a sample to each other
and was reported as %RSD. The precision samples were tested  in  independent 
runs(refer toTable5).  The evaluation of precision consisted of using the titers of the 
samples from the precision panel combined with the  samples prepared for 
dilutional linearity .
090177e196376ca2\Approved\Approved On: 09-Feb-2021 02:38 (GMT)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
FDA-CBER-2021-5683-0779473
3)0HWKRG9DOLGDWLRQRIWKH6$56&R9P1HRQ*UHHQ9LUXV 0LFURQHXWUDOL]DWLRQ$VVD\
950959HU
3),=(5&21),'(17,$/
3DJH /LPLWV RI 4XDQWLWDWLRQ
7KHORZHUDQGXSSHUOLPLWVRITXDQWLWDWLRQ//24DQG8/24UHVSHFWLYHO\ZHUHGHILQHGE\
WKHUDQJHRIWLWHUVWKDWKDYHDFFHSWDEOHGLOXWLRQDOOLQHDULW\D QGSUHFLVLRQ 7KHPRVW
FRQVHUYDWLYHYDOXHVIURPWKHORZHUDQGXSSHUWLWHUOLPLWVIURP GLOXWLRQDOOLQHDULW\DQG
SUHFLVLRQDVGHVFULEHGLQ 6HFWLRQ DQG6HFWLRQ UHVSHFWLYHO\ZHUHXVHGWRGHWHUPLQH
WKHDVVD\UDQJH 7KHDVVD\UDQJHLVERXQGHGE\WKH//24DQG8/24 
 $VVD\ ,QWHUPHGLDWH 3UHFLVLRQ
7RHYDOXDWHWKHLQWHUPHGLDWHSUHFLVLRQRIWKHDVVD\ ZDVSHU IRUPHGXVLQJWKHPRGHO
EHORZWKDW 2QO\WLWHUVZLWKLQWKHDVVD\UDQJH
OLPLWVRITXDQWLWDWLRQZHUHXVHGIRUWKLVDQDO\VLV
7KHHVWLPDWHGRYHUDOODVVD\SUHFLVLRQZDVEDVHGRQ XVLQJWK HIROORZLQJPRGHOZKLFK090177e196376ca2\Approved\Approved On: 09-Feb-2021 02:38 (GMT)
(b) (4)
(b) (4)
(b) (4)
(b) (4)
(b) (4)
(b) (4)
FDA-CBER-2021-5683-0779474
PF-07302048: Method Validation ofthe SARS -CoV-2 mNeonGreen Virus Microneutralization Assay
VR-MVR-10083, Ver. 1.0
PFIZER CONFIDENTIAL
Page 136.5.Limit of Detection
The limit of detection (LOD) was set to the low est possible dilution testedin the assay  (titer 
of 20).  The LOD was used during assay  development in lieu of an established LLOQ. Since
the LLOQ was determined through the validation experiments described here , the LLOQ will 
be used in place of the LOD.However, t o demonstrate assay  performance nearthe LOD, 8
presumed negative samples were tested  in each of the  assay  runs, for a total of up 
to reportable titer s.  The data i s presented as a descriptive measure of assay  performance .
6.6.Extravariability of Replicates
As defined in the data review SOP, VR -SOP-LC-11293,11reportable titers are the geometric 
means of the  
7. RESULTS AND DISCUSSI ON
Assay validation runs were performed by  qualified anal ysts from01-Dec-2020through 
20-Jan-2021,inclusive.  The 9 assay runs were performed to assess dilutional linearity , 
precision, assay performance near the LOD,and to determine LLOQ and ULOQ .  The raw 
titer data for the dilutional linearity  and precision experiments are listed in attachments
VR-MVR-10083-ATT0112and VR-MVR-10083-ATT02,13respectively .  LOD titer data are 
listedin VR-MVR-10083-ATT03.14  Documentation of anal yst training on the validation 
protocol is provided in VR -MVR-10083-ATT04.15
7.1. Dilutional Linearity
A sample shows dilutional linearity  when its titer changes in proportion to its dilution.  Each 
sample replicate was evaluated  and at dilutions of  
090177e196376ca2\Approved\Approved On: 09-Feb-2021 02:38 (GMT)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
(b) (4)
FDA-CBER-2021-5683-0779475
PF-07302048: Method Validation ofthe SARS -CoV-2 mNeonGreen Virus Microneutralization Assay
VR-MVR-10083, Ver. 1.0
PFIZER CONFIDENTIAL
Page 14For each of the dilutional linearity  samples,  
) 
(Figure1).  Supportive Figure 12.2shows 
this data for all individual samples from the runs plotted together.
Figure1. Dilutional Linearity Results Plot
The dilutional linearity data displayed is based on samples, each assayed  dilutions, tested in 
independent runs.  The dilutional linearity 
7.2.Precision
The assay  precision was calculated  
. Precision results for individual 
samples are display ed in Supportive Table 11.1.
 Acceptable precision  was demonstrated between titers of 41 and 
3,187 (Figure2).
090177e196376ca2\Approved\Approved On: 09-Feb-2021 02:38 (GMT)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
(b)(4
(b)(4)
FDA-CBER-2021-5683-0779476
PF-07302048: Method Validation ofthe SARS -CoV-2 mNeonGreen Virus Microneutralization Assay
VR-MVR-10083, Ver. 1.0
PFIZER CONFIDENTIAL
Page 15Figure2.Precision Results Plot
Theprecision data displayed is based on precision samples, tested twi ce in  independent runs ,and  
, tested once in  independent runs.   The precision low er and upper titer 
limitsare41 and 3,187 , respectively.
7.3.Limits of Quantitation
 
 
Table6displays the final 
assay rangebounded b y the LLOQ and ULOQ.
Table6.Final Assay Range
Dilutional Linearity Range Precision
RangeFinal Assay Range
Lower Upper
41 3,187
090177e196376ca2\Approved\Approved On: 09-Feb-2021 02:38 (GMT)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
(b)(4
(b)(4
FDA-CBER-2021-5683-0779477
PF-07302048: Method Validation ofthe SARS -CoV-2 mNeonGreen Virus Microneutralization Assay
VR-MVR-10083, Ver. 1.0
PFIZER CONFIDENTIAL
Page 167.4. Assay Intermediate Precision
The intermediate precision of the assay was evaluated using the  
as described in Section 6.4and theresults are summarized in Table7.   
  
The total %RSD is the estimated intermediate precision and the  
Table7.  S amples 
%RSD
26.5
7.5.Limit of De tection
The limit of detection (LOD) was set to the lowest possible dilution tested inthe assay (titer 
of 20).  To demonstrate assay  performance nearthe LOD, negative samples were tested
 in each of the  assay  runs, for a total of up to titersor a maximum number of 
 titers per sample .  The number of reportable titers for each sample as well as the percent 
that the sample returned a negative titer is listed in Table8.
Table8. Assay Performance Near the LOD
Sample N of Resultsa% Negativeb
100
100
100
100
100
100
100
100
a.There were  titers measured out of the  titers possible .  Sixteen titers were not resolved due to failed 
assay suitability requirements on the plate.
b.A sample is considered negative if its titer is less than the LOD of 20 (ie, the low est measurable titer).
7.6.Extravariability of Replicat es
An upper limit for acceptable replicate titer ratios was computed from titer pairs by  the 
method defined in Section 6.6.  This ratio limit was 2.55, indicating good agreement between 
titers of replicate pairs .
090177e196376ca2\Approved\Approved On: 09-Feb-2021 02:38 (GMT)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
FDA-CBER-2021-5683-0779478
PF-07302048: Method Validation ofthe SARS -CoV-2 mNeonGreen Virus Microneutralization Assay
VR-MVR-10083, Ver. 1.0
PFIZER CONFIDENTIAL
Page 178.CONCLUSION
The results documented in this validation report for the SARS-CoV-2 mNG NT provide 
evidence that the assay  is validated and suitable for its intended use in testing clinical, 
epidemiological ,and non-clinical study  samples.  The assay  demonstrated dilutional linearity  
and precision that met predefined acceptance criteria.  The LLOQ for the assay  is a titer of 
41 and the ULOQ a titer of 3,187 .  Samples with titers greater than the ULOQ may be 
pre-dilutedin assay bufferbefore testing to yield titers within thevalidated assay range.
Theperformance of the assay  near the LOD is acceptable.
9.DEVIATIONS
During statistical anal ysis of the validation dataset, it was observed that SAS was 
inappropriatel y assigninga final titer to samples with only  one valid replicate titer. This 
occurred 4 times in the validation and in each case the assay result was overridden to the 
These assay  results were then appropriatel y excluded from the final 
data set and validation analysis.  This deviation was recorded in Laboratory  Deviation 
Report VR -LDR-12674.The SAS data sy stem is being updated for future clinical 
testing.
Section 5.4 of the validation protocol (VR -MVP-100741) incorrectl y stated that the 
 diluent would be obtained from , however it was
preparedat Pfizer.
Section 7.7 of the validation protocol (VR -MVP-100741) stated that a r eplicate titer ratio 
limit of  be applied to all replicate titer pairs during validation .However, a more 
conservative limit of was applied after preliminary examination of the validation data.
10.REFERENCES
1.VR-MVP-10074: Validation Protocol for the SARS- CoV-2 mNeonGreen Virus 
Microneutralization Assay .
2.Xie X, Muruato A , Lokugamage KG et al, An Infectious cDNA Clone of SARS -CoV-2. 
Cell Host &  Microbe 2020, 27 (5): 841-848.
3.Muruato AE, Fontes -Garfias CR, Ren P et al, A hi gh-throughput neutralizing antibody  
assay for COVID- 19 diagnosis and vaccine evaluation. Nature Communications 2020, 
11 (1):4059.
4.VR-TM-10298: Manual 96- well Microneutralization Assay  for the Detection of 
Functional Antibodies to SARS- CoV-2 in Test Serum.
5.VR-SOP-LC-11299: Reagent Preparation and Vero and Vero E6 (V- E6) Thawing, 
Passaging, Harvesting and Plating.
6.VR-SOP-LC-11294: Preparation and Assessment of SARS -CoV-2-mNGStocks.
090177e196376ca2\Approved\Approved On: 09-Feb-2021 02:38 (GMT)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
FDA-CBER-2021-5683-0779479
PF-07302048: Method Validation ofthe SARS -CoV-2 mNeonGreen Virus Microneutralization Assay
VR-MVR-10083, Ver. 1.0
PFIZER CONFIDENTIAL
Page 187.VR-SOP-LC-11287: Preparation of Samples for SARS -CoV-2 Neutralization Assay  
Using the  Workstation.
8.VR-RGR-RQ-10781:Quality Control Sample Specification L imits for  
 in SARS -CoV-2-mNG Microneutralization Assay .
9.VR-VTR-10742: Exploratory Anal ysis of the SARS -CoV-2 mNG NT Validation Dataset 
using  90% Virus Neutralization Titer Readouts.
10.  
  
11.VR-SOP-LC-11293: Procedure for Data Review of the SARS -CoV-2-mNGManual 
96-well Microneutralization Assay .
12.VR-MVR-10083-ATT01: Dilutional L inearity Data Listing for VR -MVR-10083.
13.VR-MVR-10083-ATT02: Precision Data Listing for VR -MVR-10083.
14.VR-MVR-10083-ATT03: LOD Data Listing for VR -MVR-10083.
15.VR-MVR-10083-ATT04: Training Documentation VR- MVP-10074 for 
VR-MVR-10083.
090177e196376ca2\Approved\Approved On: 09-Feb-2021 02:38 (GMT)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
FDA-CBER-2021-5683-0779480
PF-07302048: Method Validation ofthe SARS -CoV-2 mNeonGreen Virus Microneutralization Assay
VR-MVR-10083, Ver. 1.0
PFIZER CONFIDENTIAL
Page 1911.SUPPORTIVE TABLES
11.1. Precision by Individual Sample
Data Source Sample Dilution N GMT %RSD
Precision
Dilutional
Linearity
090177e196376ca2\Approved\Approved On: 09-Feb-2021 02:38 (GMT)
(b)(4)
(b)(4)
FDA-CBER-2021-5683-0779481
PF-07302048: Method Validation ofthe SARS -CoV-2 mNeonGreen Virus Microneutralization Assay
VR-MVR-10083, Ver. 1.0
PFIZER CONFIDENTIAL
Page 2012.SUPPORTIVE FIGURES
12.1. Assay Design Schematic
090177e196376ca2\Approved\Approved On: 09-Feb-2021 02:38 (GMT)
(b)(4)
FDA-CBER-2021-5683-0779482
PF-07302048: Method Validation ofthe SARS -CoV-2 mNeonGreen Virus Microneutralization Assay
VR-MVR-10083, Ver. 1.0
PFIZER CONFIDENTIAL
Page 2112.2. Dilutional Linearity Plotby Individual Sample
090177e196376ca2\Approved\Approved On: 09-Feb-2021 02:38 (GMT)
(b)(4)
FDA-CBER-2021-5683-0779483
Document Approval Record
Document Name:	


Document Title:
   
  


  !

"   #
Signed By: Date(GMT) Signing Capacity
!#$ %& 
'
(
	 	)*
+*		 ' ,,-

'
(
	 	)*	*)
  ,,-
/
$ !
 
'
(
	 	)**01 ' ,,-
2$ 
 
'
(
	 	)*)*0) &
 ,,-

'
(
	 	)*5*1  ,,-

 $ 2#
 
'
(
	 	)*)1*)) ' ,,-

'
(
	 	0*00* 6#   !
 ,,-
,
$ 7- 
1'
(
	 
**	 ' ,,-
090177e196376ca2\Approved\Approved On: 09-Feb-2021 02:38 (GMT)
(b) (6)
(b) (6)
(b) (6)
FDA-CBER-2021-5683-0779484