38 1 BLA 125742 0 08 02 2021 Telecon Information Reques

Pfizer Documents (PHMPT/FDA)

Pfizer Bla Submission

Pfizer 16 Plus Documents

4

Document text

From:  Smith, Michael (CBER)  
Sent:  Monday, August 2, 2021 4:40 PM  
To: Harkins Tull, Elisa <[email protected]>; Aghajani Memar, Neda 
<[email protected]>; Devlin, Carmel M <[email protected]> 
Cc: Naik, Ramachandra <[email protected]>; Gottschalk, Laura 
<[email protected]>  
Subject:  STN 125742.0: Questions regarding the Validation Report VR -MVR -10077 
 
Elisa,  
 
The review team has the below set of questions for Pfizer regarding the Validation Report (VR -MVR -10077) entitled “Validation Report for a  
 
” that was submitted in STN 125742.0.19 on July 28, 2021.  
 
1. In your C4591001 clinical protocol amendment 17 (submitted to IND 19736 
in amendment 414 on July 20, 2021), you stated that   
 assay or S1 -binding IgG assay will be used for the exploratory 
immunogenicity endpoint.   Please clarify which immunoassay has been 
used for the assessment of binding antibodies for your Phase 2 and 3 studies.   If both assays are used, please provide the validation study report 
for the S1 IgG dLIA assay.   In addition, please provide the SOP of the 
 assay (VR -TM-10309), including a description of critical 
materials (e.g., reference standard, quality control sample (QCS), 
 antigen, , etc.), assay validity criteria, 
sample validity criteria, and interpretation of results.   Please also provide 
the test method VR -TM-10309, and the validation protocol VR -MVP -10077.  
2. We note that assay cr oss-reactivity was not assessed in your validation 
studies. Please provide data for specificity analysis, including inhibition 
experiments comparing the effects of heterologous antigens (e.g.,  antigen 
from other coronaviruses) and homologous antigens (e. g., antigens at 
different concentrations) when spiked into the panel of samples.   Alternatively, you can evaluate the assay cross -reactivity using a 
set of antibodies or serum samples with known reactivity to distinct strains of human seasonal coronaviruse s. 
3. We note that assay robustness was not assessed in your validation studies. 
Please consider evaluating the assay robustness for any variables that may 
(b) (4)
(b) (4)
(b) (4)
(b) (4)
(b) (4)
(b) (4)
(b) (4)
(b) (4)
(b) (4)
(b) (4
FDA-CBER-2021-5683-1024723
potentially affect assay results, for example, stability of the antigen -coated 
beads, assay incubation time and temperature, etc.  
4. Please provide information on the source, preparation instructions, and 
titer assignment for the reference standard and quality control samples 
that contains  in the validation study.   A 
negative QCS below or at the lower limit of quantitation (LLOQ) level should 
be included in routine testing for the  assay.   Please explain how 
you determine the cut -off values for specific  levels in the seronegative 
samples.  
5. Please note that your assa y needs to be validated using incurred samples 
from the vaccine trials for precision and accuracy via  linearity including verification of the LLOQ, ULOQ, and the assay range.   The assay 
should be fully validated prior to testing the Phase 3 clinical samples.  
6. Performance of your assay would be strengthened by inclusion of the WHO 
SARS -CoV-2 International Antibody Standard.   You might inquire from the 
National Institute of Biological Standards and Control (NIBSC) about the 
availability of this rea gent.   We recommend assessing the performance of 
your assay using the International Antibody Standard and convert your results to International Units.  
7. You set the LLOQ to be , which was obtained by  
 
 
  Such LLOQ could serve as a theoretical LLOQ, but data are needed 
to support adequate assay performance at such LLOQ.   We note that the 
precision near the lower assay limit on well concentration level was 
evaluated mostly with data at the  
 (green stars and light green diamonds 
in Figure 2 on Page 17 of the report).   We also note that the lowest incurred 
sample included in the precision evaluation has concentration of  
(Table 12 on Pages 36- 37 of the report), which is far above the theoretical 
LLOQ (i.e., ).  Please explain how the current data support  
adequate precision at the proposed LLOQ of  or provide 
additional data.  
8. We note that, in Table 10 on Page 34 of the report, the  
 do not include the  for 
some samples.   Taking  Sample  as an example, the  
(b) (4)
(b) (4)
(b) (4)
(b) (4)
(b) (4)
(b) (4)
(b) (4)
(b) (4)
(b) (4)
(b) (4)
(b) (4)
(b) (4)
(b) (4)
(b) (4)
(b) (4)
(b) (4)
(b) (4)
(b) (4)
FDA-CBER-2021-5683-1024724
 which do es not include the  
  Please verify the numbers in this table and make 
corrections, if needed.  
9. You stated in Section 5.2.7.2 of the r eport that the mean, standard 
deviation, and  
  However, as shown in Table 8 
on Page 21 of the report, 
 
 
 
 
10.Please s ubmit the data sets used for the  linearity and precision 
analyses in an analyzable format (e.g., spreadsheet).  
11.The x -axis in Figure 4 on page 35 of the report is blurry.   Please resubmit 
the figure in higher quality so that it is readable.  
 
Regards,  
  Mike  
  
- Please confirm receipt of this e -mail  and let us know if you have any 
questions.  
  
Mike Smith, Ph.D.  
Captain, USPHS  
 
Senior Regulatory Review Officer  
Food and Drug Administration  
Center for Biologics Evaluation & Research  
Office of Vaccines Research & Review  
Division of Vaccines and Related Products Applications  
Tel: 301- 796-2640  
[email protected]  
 
 
         
  
  
THIS MESSAGE IS INTENDED ONLY FOR THE USE OF THE PARTY TO WHOM IT IS ADDRESSED AND MAY CONTAIN INFORMATION THAT IS PRIVILEGED, CONFIDENTIAL, AND 
PROTECTED FROM DISCLOSURE UNDER LAW.  If you are not the addressee, or a 
person authorized to deliver the document to the addressee, you are hereby 
(b) (4)
(b) (4)
(b) (4)
(b) (4)
(b) (4)
(b) (4)
(b) (4)
(b) (4)
(b) (4)
(b) (4)
(b) (4)
FDA-CBER-2021-5683-1024725
notified that any review, disclosure, dissemination, copying, or other action 
based on the content of this communication is not authorized.   If you have 
received this document in error, please immediately notify the sender immediately by e- mail or phone.
 
  
  
  
 
FDA-CBER-2021-5683-1024726