Infusomat® fm - Frank`s Hospital Workshop

Survival, Water, Medical Field Manuals

Military Manuals

Document text

Infusomat? fm 


Service Manual 


c2) 
[amm 
C 
um, 
= 
as] 


Li 
Leed 
Lee 
f — 
SS 
} — 
— 
e 
| = 
— 
— 


e = 
In Use Tanî dr 


E 


—, 


L 














Version 2.2 English 


B BRAUN 


SHARING EXPERTISE 


This Service Manual is valid for 


This Service Manual is available under 
the following part number: 


Languages of this Manual 


The complete Service Manual contains 
the following pages: 


Voltages 200 V to 240 V: Ord. No. 


Infusomat® fm, Deman... 871 9420 
Intusomat® frm, FRENCH. cu sob tree em 871 9527 
Infusomat® fm, Dutchi« ui peor 871 9535 
Infusomat? fm, Traian... 871 9560 
Infusomat? fm, Danteh,,,,, ee eaes 871 9438 
Infusomat? fm, Norwegian.....................…… 871 9497 
Infusomat® fm, Swedish ................llsss.... 871 9500 
Infusomat® Tm; Finnish .........................- 871 9519 
Infusomat? fm, Spanish. «sav desasmertr e aet Gat 871 9457 
Infusomat® fm, Portuguese......................… 871 9462 
Infusomat? fm, English (BSI...................... 871 9446 
Infusomat® fm, English zu et mE ERE Ea tee nma 871 9543 
Infusomat? fm, Turkish ............. Lll... 871 9578 
Infusomat® fm, Czech, 871 9586 
Infusomat® fm, Polish oue Sted dodo ioe wire Sucked eG ES 871 9594 


Voltages 100 V to 120 V: 


Infusomat® fm, English (BSI)...................... 871 9411 
Intusomat® fm, Dutch + sasciacseeRke ei me iE RR 871 9470 
Infusomat? fm, Spanish... kk kk kn 871 9489 
Infusomat® fm, Castellano .....................….- 871 9551 
Designation Part No. 
Service Manual Infusomat? fm, English ........... 8713 9122 
Designation Part No. 
Service Manual Infusomat? fm, German ........... 8713 9121 


Page 0-1 to page 0-10 
Page 1-1 to page 1-12 
Page 2-1 to page 2-10 
Page 3-1 to page 3-12 
Page 4-1 to page 4-26 
Page 5-1 to page 5-2 
Page 6-1 to page 6-2 
Page 7-1 to page 7-2 
Page 8-1 to page 8-6 
Page 9-1 to page 9-2 
Page 10-1 to page 10-4 
Page A-1 to page A-2 


Infusomat® fm 2.2 gb 


Important Preliminary Remarks 


Contact Persons 


System Overview 


Software 


Service Program 


Unit Elements 


Infusomat? fm 2.2 gb 


Table of Contents 


Service Work 

Technical Safety Checks 
Current Versions 

Revision Service 

Quality Management 

Checks and Repair 

Notes on ESD 

Spare Parts and Test Equipment 
Setting Off 

List of Abbreviations 


Technical Training 
Entry for Technical Training 
Ordering of Spare Parts and Test Equipment 
Service Hotline 
Return of Spare Parts and Test Equipment 
Safety Officer 
(8 30 MPG) 
Translation 


Physical Construction 
Operation Flow Chart 
Function 

Voltage Supply 
Signal table 

Mains Operation 
Battery Operation 
Alarm Circuit 

Pump Unit 
Computer Interface 
Braun fluid manager system (fm system) 
Internal Assignment 
Accessories 


Software Update 

Approved Software Versions 
Error Messages and Alarms 
Software Default Values 


Structure of the Service Program 

Additional Functions with 
Plugged in Service Connector 

Start / Quit the Service Program 

Unit Data 

History Data 

Alignment 

Unit Modifications 

Calibration 


Mains Fuses 
Battery 
Door Lock 
Pump Cover 
Housing 


Page 
Page 
Page 
Page 
Page 
Page 
Page 
Page 
Page 
Page 


Page 
Page 
Page 
Page 
Page 


Page 
Page 


Page 
Page 
Page 
Page 
Page 
Page 
Page 
Page 
Page 
Page 
Page 
Page 
Page 


Page 
Page 
Page 
Page 


Page 


Page 
Page 
Page 
Page 
Page 
Page 
Page 


Page 
Page 
Page 
Page 
Page 


OC OO CO OC QO OS OS OO O 
I 
DO WN 0505050101 C1 01 


0-9 
0-9 
CG 
Î 
[5 
1-4 
1-5 
ee 
Îev 
127 
1-8 
1-9 
1-10 
1-10 
1-11 
Fi 
ai 
2:3 
2-9 
244 
3-2 
jeg 
3-4 
3-5 
3-6 
27 
3-10 
4-1 
4-1 
4-2 
4-3 
4-4 
0-3 


Table of Contents 


Checks after Repair 
Maintenance 
Technical Safety Check TSC 


Procedural Instructions for Inspection 


Test Equipment and Special Tools 


Spare Parts List 


Appendix 


Handle 
Microprocessor Board 
Analog Board 

Rear Panel 

Drop Sensor Socket 
Rear Panel Board 

fm Recessed Plug 
Potential Equalization Bolt 
Recessed Mains Plug 
Mains Transformer 
Reed Sensor Set 
Front Frame 

Pump Unit 

Pressure Sensor 

Air Inline Sensor 
Alarm Buzzer 

Air Insulation Board 
Frame with Seal Plate 
Door Frame 
Membrane Keyboard with Support Plate 
Display Board 

LCD Module 


1. Visual Inspection 
2. Electrical Safety 
4. Functional Inspection 


Test Equipment and Special Tools 


Unit Elements 
Software Update 
Colours 
Miscellaneous 
Accessories 


Revision Service Documentation 
Current Information 


Page 
Page 
Page 
Page 
Page 
Page 
Page 
Page 
Page 
Page 
Page 
Page 
Page 
Page 
Page 
Page 
Page 
Page 
Page 
Page 
Page 
Page 


Page 
Page 
Page 
Page 
Page 
Page 
Page 
Page 
Page 
Page 
Page 
Page 


Page 
Page 


10-1 
10= 2 
10-2 
10 -2 
10-4 


Infusomat® fm 2.2 gb 


Important Preliminary Remarks 


Service Work 


Technical Safety Checks 


Current Versions 


Revision Service 


Responsibility of the Manufacturer 


Infusomat? fm 2.2 gb 


The present manual is for your information only. The possession of 
this manual does not authorize the performance of service work. 
Service tasks may only be executed by persons, who 


- have received appropriate training on the system from 
B. Braun 


- are included in the revision service 


- possess the necessary test equipment and mechanical aids, 
and 


- fulfill the personal requirements (training and knowledge). 


The user is obliged to perform or to have performed the Technical 
Safety Checks on those medial products for which these checks 
have been prescribed by the manufacturer and to carry them out 
according to the indications of the manufacturer as well as the 
generally approved technical standards while adhering to the pe- 
riods stated (S 6 MP BetreibV). 


B. Braun also recommends training on the Technical Safety 
Checks, or to perform at least the steps indicated in the current 
version of the manual, as: 


- the TSC requires that the instructions in the manuals are ob- 
served 


= the manuals are a reference for measurements 


- depending on the unit type, the Service Program must be 
called which may lead to a dangerous unit condition in case 
of inappropriate operation. Furthermore, a special service 
connector may be necessary. 


This manual version corresponds to the state when the manual 
was written. B Braun reserves the right to make technical modifi- 
cations. The state of the revision is indicated by the index number 
in the footer of every page. 


The possession of this manual does not automatically mean inclu- 
sion in the revision service. You will be included in the revision 
service after: 


- technical training by B. Braun Melsungen or 


- a written order placed with the sales department of B. Braun 
(fee required). 


The manufacturer, person who assembles, installs or imports the 
device can only be held responsible for safety, reliability and per- 
formance if 


- mounting, enhancements, new settings, changes or repairs 
are carried out by duly authorized persons, 


Important Preliminary Remarks 


Quality Management 


Checks and Repair 


Notes on ESD 





























- the electrical installation in the corresponding room meets 
the requirements of the VDE 0107, VDE 0100 part 710 or 
IEC 60364-7-710 and the national standards, 


- the device is used in accordance with the instructions for use 
and the Service Manual, 


- the Technical Safety Checks are performed at regular inter- 
vals, 


- a current manual which corresponds to the revision state is 
used when carrying out maintenance, repair and service, 


- the service technician takes part in the revision service, 


-  thetechnician has participated in a technical training course 
for the specific B. Braun unit. 


B. Braun is certified in accordance with DIN EN ISO 9001 and 
ISO 13485. This certification also includes maintenance and serv- 
Ice. 


The unit has the CE label. The CE label confirms that the device 
corresponds to the "Directive of the Council for Medical Products 
93/42/EC" of June 14, 1993. 


Training may only be performed by B. Braun. The possession of the 
manual does not authorize the performance of repairs. The in- 
structions on electrostatic sensitive components (ESD standards) 
must be observed. 


After repair a device check or diagnosis is to be carried out. 


Semiconductors can be destroyed by electrostatic discharge. Es- 
pecially MOS components can be damaged by interference from 
electrostatic fields, even without discharge via contact. This type 
of damage is not immediately recognizable. Unit malfunctions 
can even occur after a longer period of operation. 


Each workstation must be equipped according to the recommen- 
dations with the necessary static protective measures, if ESD 
components or boards are handled. 


Each workstation must be equipped with a conductive table sur- 
face. The conductive surface, the soldering iron or the soldering 
stations must be grounded via protective resistors. 


Chairs must be of antistatic design. The floor or floor mats should 
be of electrically conductive material. 


Personnel must wear conductive wristbands which are connected 
to a central ground potential via protective resistors, e.g. the 
ground contact of a wall outlet. Furthermore it is recommended 
that personnel wear cotton clothing and electrically conductive 
shoes to prevent electrostatic charge. 


Infusomat® fm 2.2 gb 


Spare Parts and Test Equipment 


Setting Off 


Infusomat? fm 2.2 gb 


Important Preliminary Remarks 


Only use original spare parts from the manufacturer. Do not 
tamper with assembly groups which can only be exchanged com- 
pletely. The spare parts required are listed in Section 9. 

Service personnel are responsible for the calibration of their test 
equipment. Original test equipment can be calibrated at the 
works of B. Braun. Further information is available upon request. 


Additional notes and warnings are set off as follows: 


Note 


Is used for additional or special notes concerning information and 
working steps. 


CAUTION 


Is used for working steps which may result in damage to the unit, 
system or to a connected device. 


WARNING 


IS USED FOR WORKING STEPS WHICH MAY RESULT IN PERSONAL 
INJURY. 


References to chapters are shown as follows 
(see "Setting Off" » pg. O - 8) 


References to figures and tables are shown as follows 
Fig.: 2 - 3or Table 2 - 1 


References to item numbers in figures are shown as follows 
(Fig.: 1 - 1 / Item 1) 

In this case "Fig.: 1 - 1" is the figure number and "Item 1" the item 
number within the figure. 


When the Service Manual is stored as pdf-file, these references 
are displayed green. Click with the mouse button on a reference 
to jump to the corresponding source. 


Menu commands are described as: 
Menu File 


Important Preliminary Remarks 


List of Abbreviations Åbbreviations which are not generally known, but are used in this 
manual, are listed below. 


CC Computer Controlled 
CLR Clear 
Dig Digit 
ESD Electrostatic Discharge 
UA Unit Alarm 
FuP Function Microprocessor 
KuP Control Microprocessor 
LCD Liquid Crystal Display 
OIL Original Infusomat Line 
PCA Patient Controlled Analgesia 
TSC Technical Safety 

Check 
TEMP Temperature 


0-8 Infusomat® fm 2.2 gb 


Technical Training 


Entry for Technical Training 


Ordering of Spare Parts and Test Equipment 


Service Hotline 


Return of Spare Parts and Test Equipment 


Safety Officer 
(S 30 MPG) 


Translation 


Infusomat? fm 2.2 gb 


Contact Persons 


Via local representative. 


Application for a technical training course must be made via the 
responsible representative. 


Please contact your local B. Braun subsidary. 


International Technicians (Intercompany) 
Nadja Machal 
Fax: +49 5661 / 75 -47 89 
e-mail: [email protected] 


Karl Tippel, Tanja Kórdel 
Phone: +49 5661 / 71 - 35 25 


Fax: 449 5661 / 71 - 35 26 
e-mail: — [email protected] 
e-mail: — tanja.koerdel 2 bbraun.com 


B. Braun Melsungen AG 
Schwarzenberger Weg 73-79 
Wareneingang Werk C 

34 212 Melsungen 

Germany 


Dr. Ludwig Schütz 
e-mail: [email protected] 


PAS GmbH, Brückner GmbH, Germany 


211 Contact Persons 


For your notes: 


0- 10 Infusomat? fm 2.2 gb 


System Overview 1 


Physical Construction The Infusomat fm is a compact volumetric peristaltic infusion 
pump. 
Standard delivery rate range 1 to 999 ml/h 
The unit is operated via a membrane keyboard. It is equipped with 
an LED display (light emitting diode display) for delivery rate dis- 
play and an LCD display (liquid crystal display) for the operating 
support of the user. Four control LEDs display alarms, battery- and 
mains operation and the running of the infusion pump. 


Flow inhibitor 
Drop sensor 


Handle | ecu Peristaltic pump 


+ 


d 
| 


Door lock 


DIANET 


LETTERE 


LED display 


HI iz; 
TIT Tin 


Hut 
tt 





Control LEDs 


KIA 


Keyboard 





LCD displa 
a Optical interface 


Softkeys ar | fm recessed plug 


Potential equalization 


Staff call 


Air inline sensor 


Adapter for infusion line 


Drop sensor socket 
(depending on the unit number) 


Mains connection 


Unit overview Infusomat fm from unit 000001 on 





Infusomat® fm 2.2 gb 


System Overview 


Operation Flow Chart 


Switch on 


Service connector 3) 
plugged … 1) 


Standard operation Special functions ... 


Standby function 


Set delivery rate Volume preselection Time preselection i 
Drug selection 





CC mode 
Delivery calculation 





Switch-off pressure 


Start infusion 





Stop infusion Drop control 


Change delivery rate ... 


Battery capacity 





Recall info 


Data lock 


KOR mode 2) 


Contrast setting 





! See Service Program diagram 
? Only when preselected time or preselected volume has expired 


? Please pay attention to activation in Service Program 
Fig.:1-2 


See instructions for use for detailed information. 


152 Infusomat? fm 2.2 gb 


Function 


~ 








Text Display < 


d 











K eypad 
Query 











` 


K eypad 
X/Y Matrix 


I/O J 


> 
> 





Function 
Processor 
535 











` 


Keypad 
Query 




















» 


A 


System Overview 


Two independent software-controlled microprocessor systems 
control and monitor the hardware. On the basis of their functions, 
they are defined respectively as a control and a function proces- 
sor. Both systems work with independent clock frequencies and 
have access to different program and data memories. All safety- 
relevant functions are handled by both microprocessors and the 
results are counter checked (CF- and FC-latch). 


The input via the keyboard is fed to both processors. Additionally 
the acknowledgement signal of the ON/OFF key is fed to the 
mains power supply logic (voltage E/A-TAS). The function proces- 
sor has also access to this logic via E/A-INT. 


Description of the voltage signals (see „Signal table" => 
p. 1 - 5). 








e DIANET 











Control Signal 





Drop D 











. Sensor , 
Test > - 





LFS EI Air Detection 








» P f 
TSA Drop Detection Së 


Air ) 
. Sensor ` 











PKS 1/2 Position Detection 














FC-Latch 


CF-Latch 





Y 











Rate Dispaly 
Alarm, Operation, a. 
Battery 


L 








LED Delivery 




















Softkey | 
, Illumination | 


d 








Fig.: 1 - 3 


Infusomat® fm 2.2 gb 





Display a 


Control 











Control 
Processor 
HC11 











PKS 1 Speed Control | 
(Double Light Barrier) 4 — 
— 














MS y 
MOTIN _»! Motor Output Stage | — P 





Pump 








Test e e Staff Call 


e Plug 






































ALA-UB — | 
Test 5V Monitoring . NC Battery 
| 7,2N 




















RES-F 





RES-K Reset Generator — Selection 




















45V 12V-16V ` fm System 
: 11V-16V | A External 
Operating Voltage 


Generation 


UBS 6-11V 
( Mains 


EA-INT 














> ON/OFF Logic 











EA-TAS 


System Overview 


Voltage Supply 


The voltage supply can be generated directly from mains (170V to 
264V~ or 85V to 132V-.), an external 12 V supply connected to a 
fluid manager system or the DIANET recessed plug or as an inter- 
nal supply via the internal 7.2 V NiCd battery. The battery type 
(long- or short time battery) is identified by the coded battery 
plug. 

A voltage of 12V to 32V is available after transformation and rec- 
tification. The voltage regulator limits voltage to max. 12V. This 
voltage is fed to the battery charge circuit and the unit supply via 
isolation diodes. This is also valid for an external 12V voltage from 
the DIANET or an fm plug. The FET V22 switches between external 
and internal voltage supply. The transistor V29 works as an ON/ 
OFF switch for the operating voltages UPS, UMOT and +5V. The 
+5V supplies the complete electronics including the double chan- 
nel microprocessor system. A window comparator constantly 
monitors the +5V for undervoltage or overvoltage. The function is 
checked during switch-on. The operating voltage UPS supplies the 
stepper motor and UMOT the stepper motor drive. 


The transistor V21 switches the operating voltage UMOT. In case 
of an alarm the motor is switched off by V21. Additionally the 
switching function of the transistor is checked during the switch- 
on test. 


The circuit has two separate assembly groups with separate sup- 
ply voltages UBA and UBB. The ON/OFF circuit UBA is a retrigger- 
able delay switch-off. A follow-up charging circuit drives the 
transistor V29. V18 is additionally active in battery operation. 


The alarm logic UBB is an RS latch. This is set when the unit is run- 
ning and activates the alarm circuit. The alarm buzzer and driver 
are also driven by UBB. 


The ON/OFF circuit is activated and the voltage supply is switched 
on by pressing the ON/OFF key. The alarm latch is reset simulta- 
neously. A function test of the voltage monitoring, motor circuit 
and alarm activation is performed. The voltage supply is main- 
tained by cyclic self-holding pulses fed to the logic. The alarm 
latch is also activated. 


Infusomat® fm 2.2 gb 


Mains 


170-240 V 


fm Syste 
12-16 V. 


Signal table 


Signal 











Switching 
Þ e e Regulator 
3h | 1-100% 



































e > 
F1 V4 


/^. Switch Current |. 


XM 

























































































u— 
Battery | 6-1,2V 














Meaning 





ON/OFF Logic 








System Overview 





EN UBS 





EE UMOT Drive 





Linear Regulator 


d 

















y 








Over- and Undervoltage | 
Monitoring z 

















- HH FA TAS 
NEM ALA AK 





Meaning 





+5V 
5V-HT 
5V-LT 
AK-| 
AK-LAD 
AK-Test 
ALA-RED 
ALA-UB 
CS 

DI 

DO 
E/A-INT 
E/A-ST 


Table 1 - 1 


Voltage supply electronic 
Overvoltage test 
Undervoltage test 

Battery charge and discharge current 
Battery capacity ON/OFF 
Battery test 

Alarm reduction 
Operating voltage alarm 
Chip Select 

Data Input 

Data Output 

ON/OFF by microprocessor 
ON/OFF status 


Signal table (Part 1 of 2) 


Infusomat? fm 2.2 gb 


Staff call relay control channel 
Staff call relay test 

Delivery display clock 

Delivery display data 

Delivery display enable 
Delivery display test output 
Delivery display test synchronization 
Power on Reset 

Reset function channel 

Reset control channel 

Return to send (DIANET) 
Receive data 


Tube adapter 1 


Signal 
E/A-TAS 
EDB 
EXTN 
EXTP 
EXTPP 
FL-S 
LFCL 
LFDA 
LFS 
LFSEL 
LFT 
MISO 
MOSI 
MOTEIN 
MS 
NA 
P-ENA 
PA 
PHO 
PH1 
PH2 
PH3 
PKS 
PKS1 
PKS2 
PKSS 
PRS 
PRS-F 


Table 1 - 1 


System Overview 


Meaning 

ON/OFF key 

Electronic pressure control 
External 12V supply (-) 
External 12V supply (4) 
External 12V supply (FM) 
Tube adapter OIL-FM 

Air sensor clock 

Air sensor data 

Air sensor signal 

Air sensor selection 

Air sensor reset 

Serial data output interface 
Serial data input interface 
Motor ON 

Motor control 

Mains display 

Port Enable 

Potential equalization 
Phase 0 

Phase 1 

Phase 2 

Phase 3 

Pump cover sensor 

Pump head sensor 1 

Pump head sensor 2 
Pump head sensor control 


Staff call relay control 


Staff call relay function channel 


Signal table (Part 2 of 2) 


Signal 
SA2 
SCK 
SL-S 
TD-AO 
TD-A1 
TD-A2 
TD-A3 
TD-B 
TD-E 
TD-KL 
TD-R/W 
TD4 
TD5 
TD6 
TD7 
TSA 
TSCL 
TSE 
TSR 
TSS 

Tx 

UB 
UBA 
UPS 
UPS-M 
UMOT 
UMOT-M 
URTC 


Meaning 

Tube adapter 2 

Serial data lock 

Tube adapter OIL 

Text Display Address 0 

Text Display Address 1 

Text Display Address 2 

Text Display Address 3 

Text display illumination 

Text Display Enable 

Text display contrast latch 
Text Display Read/Write 

Text Display Data 4 

Text Display Data 5 

Text Display Data 6 

Text Display Data 7 

Drop sensor output 

Drop sensor clock 

Drop sensor receiver 

Drop sensor regulation 

Drop sensor control 

Transmit data 

Operating voltage 6-12V 
Supply voltage for alarm, On/Off logic, RTC 
Switched operating voltage UB 
UPS measuring line 

Supply voltage of motor drive 
UMOT measuring line 


Supply voltage clock module 


Infusomat® fm 2.2 gb 


Mains Operation 


Battery Operation 


Alarm Circuit 


Infusomat? fm 2.2 gb 


System Overview 


When the unit is connected to mains the unit supply voltage is 
switched on for the duration of the switch-off delay time. If the 
microprocessor recognizes a sufficient mains voltage for charg- 
ing, the voltage supply is maintained. In this case only a battery 
balance is carried out, because a key was not pressed. The mains 
control LED is switched on, all other displays are off. 


The unit is switched off when the ON/OFF key is pressed for at 
least 2 seconds. Thereby the self-holding is triggered and the 
alarm latch is reset with a delay. After another 8 seconds the unit 
is switched off, because the pulses are missing. If the Infusomat 
fm is switched off in mains operation with the ON/OFF key, the in- 
ternal mains voltage is still present. All displays are switched off 
and battery charging is monitored. 


In mains operation battery function is checked during the switch- 
on test. Therefor the charge- and discharge current are measured 
and battery charging is interrupted for the duration of measure- 
ment. 


The battery function is monitored by the following data: charge 
current, discharge current and time, and self-discharge time. The 
electronic detects a short time or a long time battery by measur- 
ing the charge current. 


In battery operation the battery function is checked during 
switch-on test. The theoretical load condition is read from the 
clock module of the battery. Then the battery is connected to UPS 
and the voltage is measured. If the minimum requirements are not 
reached a battery alarm is activated. 


The control microprocessor drives the red alarm LED. 

A function check is performed during the switch-on test. 

The alarm buzzer is activated in case of: 

- an operating alarm from control processor (ALA-AK signal) 
- a unit alarm from both processors 

Alarm generation: 


The user checks this function during the switch-on test. The au- 
dible alarm volume in double stage operation is approx. 50 dBA. If 
no alarm acknowledgement or unit handling is performed, the 
maximum volume (approx. 65 dBA) is activated after 2 minutes. 


A single or double stage alarm can be selected in the Service Pro- 
gram. The volume is immediately 65 dBA in single stage mode. 


System Overview 


Pump Unit 


The staff call is activated by the processor system (PRS signal). A 
function check is performed with a second relay contact (PRT sig- 
nal). A static or dynamic staff call can be selected. 


The pump head is driven by a stepper motor. Each full step of the 
motor is realized with 5 microsteps. The motor is driven by an FET 
output stage. The function processor controls the motor via the 
MS signal. A slot disc which is mounted on the pump head axle is 
scanned by two light barriers (PKS1 and PKS2 signal). Thereby the 
control microprocessor monitors direction of rotation and speed 
of the pump head. 


The pump head position is also determined with the PKS2 signal. 
The motor can therefore be accelerated during the withdrawal 
phase. Thus a nearly pluse-free flow is realized in the lower deliv- 
ery range («100 ml/h). The total pump head cycles and running 
time are available in the Service Program under history data. 


Mechanical Pressure Control: 


The Infusomat fm has a linear peristaltic pump. This pump has 12 
slides which are driven by a camshaft. 


When the pump cover is closed, the pump tube is squeezed (oc- 
clusion) by at least one of the slides, independent of the pump 
head position. The complete pump unit is mounted behind the 
front panel in the frame. Hinges and locking bow for the pump 
cover are led through the front panel. The pump cover is automat- 
ically closed when the operating unit door is closed. The slides are 
pressed against the pump cover by a spring system in the pump 
unit. Thereby a delivery pressure is realized and mechanically lim- 
ited by the springs. 


If the pressure limit is exceeded there is no volume delivery. The 
drop sensor activates an alarm. If one of the springs fails, the 
spring system will ensure that an unsafe condition cannot occur 
(free flow). The two remaining springs ensure an appropriately 
high occlusion pressure. 


Infusomat® fm 2.2 gb 


Computer Interface 


Infusomat? fm 2.2 gb 


System Overview 


Electronic Pressure Control: 


The electronic pressure sensor is mounted on the output side of 
the pump. Å spring pressure loaded slide is seated on the infusion 
line. An increase of pressure in the infusion line leads to a deflec- 
tion of the coil core via the pressure slide. The depth of immersion 
Is measured inductively. When a preset pressure threshold is 
reached the pump drive is switched off, and an alarm is activated. 
The electronic pressure control is a single channel circuit. In case 
of a failure, the mechanically limited maximum pressure can be 
reached. 


Motor Switch-Off by Both Processors: 


Function processor: MS signal to switch off the motor drive. - 
Control processor: MOTEIN signal to switch off the drive of the 
motor operating voltage. 


The DIANET is designed as a SV TTL interface until unit No. 17147. 
From unit No. 17148 on an RS 232 interface is present. Retrofit- 
ting is possible by exchanging the rear panel (see ,,Rear Panel 
Board" Þ> p. 4 - 11). The RS 232 interface is marked by a label 
on the rear side of the Infusomat fm. 


Only the following accessories must be applied for the ohmic in- 
sulation between the Infusomat fm and a PC. 


Designation Ord. No. 

For the 5V TIL InterTaCesa va soos Ee aa wêk kê TES 0871 9322 
Interface converter DIANET sc 

Interface [INE m 0871 1640 


Infucab fm, for RS232 interface 


Interface cable fm without ohmic insulation ....... 0871 1658 
(Only applicable for servicing without unit connected to a 
patient.) 


System Overview 


Braun fluid manager system (fm system) The Infusomat fm can be operated as a stand-alone unit or inte- 
grated in an intensive care unit, e.g. the B. Braun fluid manager 
system. It is integrated by simply snapping the unit into the sys- 
tem. 


Mains supply and data communication are automatically con- 
nected. Thereby data acquisition and transmission to higher com- 
puter system levels are possible. 


Internal Assignment 



























































































































































































































































































































































































































































































































































































































































Protective Earth Conductore 





LCD Pressure 
Eê Sensor Board 
+5V 
PKSS 
PKS1 Pump Speed Electronic 
PKS2 Monitoring Occlusion Pressure 
au 
CS 
SK 
DI " 
e emory 
Ney DO Pump Data 
LED mٍ 
EA-TAS SAT 
SCK = Tubing Adapter 
i RDE SA2 
Display MOS 3 
RDTS 
Board BDTA PKS Pump Cover Sensor 
TD4 L 
TD5 
TD6 
TD7 
TD-RW 
MISO 
TD-AO 
TD-A1 A | 
TD-A2 
TDA nalogue 
-P-ENA Board 
PHO 
MS Motor PH1 Stepper 
PH? 
Output Stage PHS Motor 
TSCL i tee 
KE VO TSR Drop Detection +5V Drop Sensor 
SE Keypad 
Illumination LFDCL Transmitter 
LFCL 
LFDA Ai ; i 
ir Detection Air 
LESEL S| Insulation 
LFT Receiver oard 
+5V i | I | 
2 +5V + - 
X > 
T Air Sensor 
R m : 
Eë Plug RTS REST Mains 
Rear NA an Power Supply ] Buzzer 
12 V External Panel Board UBS-M 
PRS UMOT-M 
PRI ALA-AK 
PA ALA-RED 
e = 
E/A-LAD _ Battery 
AK-I T 
NA : 
' Battery Selection 
Staff Call ria 
Recessed Plug L 
fm System Gr 
PA 
D AN EXTP 
Z © EXTN 
Netz 1 " 
Mains Voltage li | li Netz 2 
e e 


Fig.:1-5 


1- 10 Infusomat® fm 2.2 gb 


Accessories 


Infusomat? fm 2.2 gb 


System Overview 


General 

Designation Ord. No. 
Mounting clip for drop chamber "TK 2000" ........ 3477 3223 
Mounting clip for drop chamber "Intrafix air" ...... 3477 3215 
Dröpþ:sensor,:Complêe tê; «ia xl kk na wee aoe DU ESQ 3450 578A 
SOTE Stal ERE 0870 1644 
Drop chamber NOEs WWW KK RR E cate 3477 3088 
Mains lead (200-240V~)...... 0... RY ee 3450 2718 
Mains lead (100-120V~) ........ 00.0 cee V ee 3450 5423 
Holder for mains plug (option)................... 3450 5474 
Universal clamp, complete `... 3450 5857 
Pole clamp (universal clamp, rotating) ............ 3450 9054 


1- 11 


System Overview 


For your notes: 


1- 12 Infusomat® fm 2.2 gb 


Software Update 


Position 12 34 5 6 7 89 
Digit I F M B 0 0 0 0:2 
Revision level 


Hardware identification 


Software group 


Infusomat fm 


Approved Software Versions 


Infusomat® fm 2.2 gb 





Software 7 


Designation Ord. No. 
Update kit IFMB00002 or IFMBO1002 ............. 3450 6306 


The higher digit always replaces the lower digit for the revision 
level, e.g. IFMA00003 replaces IFMA00002. 


Units with an old software version, e.g. IFMA00003 can be updat- 
ed to the new software version IFMBO0002. 


When the software group changes (IFMAO0002) the unit func- 
tions are changed, too. Therefore unit users must be informed (e.g. 
instruct the user and exchange the instructions for use - software 
coding, e.g. IFMBO0002 is on the cover page of the instructions for 
use.) 


Note 


Mark the unit after having updated the software! The new soft- 
ware version must be clearly recognizable. 


Only update from old to new software versions, never in reverse 
order (e.g. never update from IFMBO0002 to IFMBOOOO01 or 
IFMAO0003 |). 


All units used in one ward should have the same software status 
and basic setup to avoid operator mistakes. 


Note 


Software updates must be reported to B. Braun for registration. 
Observe the notes of the update program and the supplements. 


IFMA00002 


- Basic software 


IFMA00003 


- Improved resistance to interference against electrostatic 
charges. 


- The opto staff call is corrected after switch-off. 


- Extended display of unit alarms. The unit alarm no. is dis- 
played on the LCD display. 


IFMAO0003 (17.05.95) 

Elimination of unit alarms (modification of the hc-11 software): 
- For frequent starts and KOR (KVO) operations. 

- For switch-on with empty battery. 


Software 


IFMBOOO01 or IFMB01001 
Expanded functions: 


- Optimization of staff call modes 
(see instructions for use - staff call lines). 


- Extension of the interface functions for recommended data 
(see DIANET interface description). 


- Display of battery capacity, ward identification and software 
version when the unit is switched off. 


- Activation of the display light in battery operation when a key 
IS pressed. 


- Decimal function (can be activated in Service Program). 
Preselect delivery rate from 1 to 99.9 ml/h with increments of 
0.1 ml/h. 


- Display of the calculated rate from volume and time with one 
decimal. Thus the rounding error is significantly reduced. 


- Additional running control on the LCD. 


- Additional alarm display. 
Delivery display flashes with RATE and AAA. 


- Correction of the total air alarm (accumulation of 
microbubbles) from 0.7ml to 1.5 ml air per hour. 


- 10 minutes alarm tone suppression (can be activated in the 
Service Program). 


- Special functions can be deactivated in the Service Program. 


- Service Program menu point Calibration Pump Head (code 
500) is replaced by Calibration Pressure Sensor (code 500) 
and Calibration Scale Factor (code 510). 


- Service Program menu point Air Inline Sensor (code 300) now 
includes only the display of the measured values and the test 
values. 


- Stored operating alarms can be deleted in the Service Pro- 
gram. 


Software IFMBO00002 or IFMBO1002 

Expanded Functions 

- The min./max. delivery rate can be set. 

- The air rate alarm can be set between 0.5 ml/h and 3.5 ml/h. 

- Display of the operating hour counter in the SM battery ca- 
pacity 

Error Elimination 


- Opto staff call 


Infusomat® fm 2.2 gb 


Error Messages and Alarms 


Code 
01 
02 
03 
04 
05 
06 
07 
08 
09 
OA 
OB 
0C 
0D 


Table 2-1 


Text* 

Wéi 

externes RAM defekt 
T1. Netzteiltest 

T2 Ueberspannungtest 
T3. Motorabschalttest 
IA Unterspannungstest 
EA Taste klemmt 
Luftfalle defekt 

LED Test Fehler ! 

LED Test Fehler ! 

LED Test Fehler ! 

LED Test Fehler ! 

LED Test Fehler ! 

LED Test Fehler ! 


(Part 1 of 2) 


Infusomat® fm 2.2 gb 


Description 


alarm reason not clear 


Software 


WARNING 


NEVER OPERATE THE UNIT WITH PLUGGED IN SERVICE CONNEC- 
TOR ON THE PATIENT. WHEN THE UNIT WAS OPERATED WITH 
PLUGGED IN SERVICE CONNECTOR IT IS TO BE SWITCHED OFF 
ONCE BEFORE ANY FURTHER USE. 


Alarms of the function processor 80c535 are displayed on the LCD 
display. Alarms of the control processor 68hc11 are displayed on 
the LED display. The alarms help to troubleshoot unit malfunc- 
tions. As not all malfunctions can be considered, unit malfunc- 
tions with different messages or even no messages can be 
displayed on the LCD/LED display. 


Software IFMA00002, IFMAO0003: 


Detected unit alarms are displayed as 

"I Unit Defective !!" on the LCD display in the selected language. 
If the unit LED has malfunctions "LED Test Error !" will be dis- 
played. Additionally the error number (from IFMAO003 on, for 
IFMAO0002 the number is not displayed) is displayed on the LCD 
display if the service connector is not plugged in. If the service 
connector is plugged in, a German text is displayed instead of the 
alarm number. 


1. LCD display function processor 80c535: 


defective external RAM 8k x 8 

power supply test: 5V_HT=0; 5V_LT=1; MOTEIN=1: UMOT-M unequal 1 
power supply test: 5V_HT=1; 5V_LT=1; MOTEIN=1: UMOT-M unequal 0 
power supply test: 5V_HT=0; 5V_LT=1; MOTEIN=0: UMOT-M unequal 0 
power supply test: 5V_HT=0; 5V_LT=0; MOTEIN=1: UMOT-M unequal 0! 


ON/OFF key was pressed longer than 14 sec 


defective air sensor (calibration value?) 


defective LED display 
defective LED display 
defective LED display 
defective LED display 
defective LED display 
defective LED display 


Code 
OE 
OF 
10 
11 
12 
13 
14 
15 
16 
17 
18 
19 
1A 
1B 
1C 


Table 2 - 1 


Code 
F01 
F02 
F04 
F08 
F10 
F20 
F40 
F80 


Table 2 - 2 


Software 


Text" 

LED Test Fehler ! 

535 ROM defekt 
Modulzeitkon. XX 

MS hat Frequenz 

pku Erfassung 535 ! 
Tastatur defekt 

Mode unterschiedlich 
Mode geaendert 
Schlauchadapter fehlt 
Version ungleich 
c535 Timeout XX 
testbit defekt 
personalruf defekt! 
personalruf defekt? 
personalruf defekt? 


(Part 2 of 2) 


Description 

defective LED display 

defective program 

defective program flow 

MS has frequency without drive 

pump head cycle not plausible 

different keypad gaps between 80c535 and 68hc11 
mode in 68hc11 different from 80c535 

changed operating mode 

defective tube adapter 

different program versions between 68hc11 and 80c535 
defective program flow 

testbitzenable out of switch-on limits 

535 active PR, hc11 not active PR: PR=active 

535 not active PR, hc11 active PR: PR=active 

535 not active PR, hc11 not active PR: PR=not active 


* Text will be only displayed if the service connector is plugged in. 
2. LED display control processor 68hc11: 


Fxx/Exx flashes alternately on the LED display. Fxx is the error 
code. Exx can be neglected. If errors occur simultaneously, 
the alarms are displayed by hexadecimal addition, e.g. defec- 
tive battery FO1 and defective time F10 = error code F11. 


Description 

defective battery 

defective pump head 
defective ROM memory 
defective EEPROM memory 
defective time 

defective active 

defective active reset 


defective switch 


Infusomat? fm 2.2 gb 


Code Text" 
100 
101 
102 
103 
104 
105 
106 
107 
108 
109 
110 
111 
112 
113 
114 
115 
116 
117 
118 
119 
120 
121 
122 
123 
124 
125 
126 
127 
128 


Table 2 - 3 


Infusomat® fm 2.2 gb 


Software 


Software IFMBOOOXX 


Detected alarms are displayed on the LCD display as "Unit Alarms" 
in the selected language. If the unit LED display has malfunctions 
"LED Test Error !" will be displayed. Additionally the error number 
is displayed on the LCD display. 


1. LCD display function processor 80c535: 


Description 

defective RAM memory U13 

UMOT cannot be switched on 

UMOT still switched on despite overvoltage 

UMOT still switched on despite MOTEINZO 

UMOT still switched on despite undervoltage 

ON/OFF key pressed longer than 14 sec 

defective air sensor (calibration value ? ) 

defective LED display 

defective LED display 

defective LED display 

defective LED display - RDTS 

defective LED display 

defective LED display 

defective LED display 

defective program memory U10, U11, U15 

defective program flow 

MS has frequency without drive 

different number of pump head cycles 

different keypad gaps between 80c535 and 68hc11 

mode different in 68hc11 

changed operating mode 

defective tube adapter 

different program versions between 68hc11 and 80c535 
defective local program flow 

testbit=0 out of switch-on test 

PR is not active though 80c535 active PR and 68hc11 active PR 
PR is active though 80c535 not active PR and 68hc11 active PR 
text not loaded / defective program memory U10, U11, U15 


text does not match with program / defective program memory U10, U11, U15 


* Text will be only displayed if the service connector is plugged in. 


Code 
F01 
F02 
F03 
F04 
F05 
F06 
F07 
F08 
F09 
F10 
F11 
F12 
F13 


Table 2 - 4 


Software 


2. LED display control processor 68hc11: 


Fxx is displayed on the LED display with flashing dots. Fxx is 
the error code. 


Description 

dummy for test 

battery not present / missing battery current 
defective RAM memory U17 

defective program memory U21, U22, U31 
defective program memory U21, U22, U31 
calibration data error from EEPROM U1 
pump head cycle not plausible 

failure / inaccuracy of system clock 
failure 100msec system clock 

reset during active operation 

changed mode or 80hc535 different 

no dynamic pressure sensor signal (EDB) 


different status during start 


Infusomat? fm 2.2 gb 


Software 


Alarm Causes 
1. Drop alarm/pressure alarm 
- Empty infusion bottle? 
Connect new bottle. 


- Closed roller clamp/flow? 
Close roller clamp, stop infusion, open roller clamp. There 
must be no continuous dripping. If necessary insert new 
infusion line. 

- Occlusion? 
Lay line without any kinks and check integrity of 
complete infusion line. 

- Drop chamber clouded? 
Shake to remove. Drop sensor not inserted/connected? 
Insert/connect drop sensor. 

- Defective drop sensor? 
If necessary exchange drop sensor. 


2. Ar alarm 
- Ar in system? 
Insert line correctly. Vent and reset fluid level in the drop 
chamber. 


3. Standby alarm 


- Alarm after set PAUSE has expired? 
Switch to standby with SM, stop PAUSE with OFF or 
extend PAUSE with ON. 


4. Battery alarm 


- Battery alarm or battery prealarm? 
Switch off immediately, connect to mains or to 12V. 
Charge battery. 


5. KOR alarm 
- Preselect new delivery rate and start again. 
6. Further alarms/displays 


- Pump cover open? 
Close door, set rate. 


- ` No rate? 
Enter new value. Value correction? 


Infusomat? fm 2.2 gb 2-7 


Software 


Infusomat® fm 2.2 gb 


Software Default Values 


Standard function 


* 


Special Functions 


User Data 


Calibration Data 


Unit Specific Data 


Infusomat® fm 2.2 gb 


Menu Item 
User language 
Alarm type 
Staff call 
Ward identification 
Drug 0 
Drug 1 ... 9 
Operating alarms 
Minimum delivery rate” 
Maximum delivery rate” 
Maximum air rate” 
Maximum bubble size” 
Standby 
Drug selection 
CC mode 
Switch-off pressure 
Drop control 
Battery capacity 
Data lock 
Contrast 
Decimal function" 
Switch-off pressure 
Contrast 
CC address 
Drug 
Data lock 
Standby time 
Drop control 
Air inline sensor calibration value 
Scale factor OIL 
Contrast only if necessary 
DIANET type no. 
Unit No. 
Operating hours 
Battery hours 


Number of pump head cycles 


Unit No.: 


Default 
depending on Art. No. 
double-stage 
static without OFF alarm 
"Ward Identification" 
blank 
drug 1 ... 9 
0 
1.0 ml/h 
999.0 ml/h 
1.5 ml/h 
0.3 ml 
activated 
activated 
activated 
activated 
activated 
activated 
activated 
activated 
deactivated 
high 
optimum contrast 
1 
0 
Off 
30 min 
On 
182mV 
54 


Software 


Customer Setting 


must not be changed 


set to maximum without overmodulation 


depending on unit 
depending on unit 
depending on unit 
depending on unit 


depending on unit 


* only from software IFMBO1001 or IFMBOOO! on 
™ from software IFMB00002, IFMB01002 on 


Software 


For your notes: 


2- 10 Infusomat® fm 2.2 gb 


Service Program 


Structure of the Service Program 


Service connector 
plugged 





Switch on 


Short display: 
software version, 
user language 


Standard operation … 
(see operating 


Special function 
(see operating 
flow diagram) 


Service Program 
activated 
Group: Group: Group: Group: Group: 
Unit data History data Alignment Unit modification Calibration 


Software version Operating hours Air inline sensor Service language Pump head 
100 200 300 400 500 * 
Drug Battery hours in h Pressure control User language 
110 210 310 410 


flow diagram) 





Air inline sensor 





E NN bu udi 
Ward identification Pump head Alarm tone Pressure sensor 
120 cycles 420 500 ** 
220 
Infusion line Staff call Scale factor 
130 Operating alarms 430 510" 
230 
Serial number Special function Air inline sensor 
140 440 ** 520 ** 
DIANET type no. Decimal function 
150 450 ** 
Delivery rate 
* until software IFMAO0003 460 ^ 
** from software IFMBOOO001 / 01001 on ; 
Air alarm 
** from software IFMBO0002 / 01002 on 470 *** 
Fig.: 3-1 


Infusomat? fm 2.2 gb 


Service Program 


Additional Functions with 
Plugged in Service Connector 


WARNING 


NEVER OPERATE THE UNIT WITH PLUGGED IN SERVICE CONNEC- 
TOR ON THE PATIENT. WHEN THE UNIT WAS OPERATED WITH 
PLUGGED IN SERVICE CONNECTOR IT IS TO BE SWITCHED OFF 
ONCE BEFORE ANY FURTHER USE. 


Software Version and User Language 


1. Plug service connector on staff call socket at the rear of the 
unit. 


2. Switch on unit and keep the ON/OFF button pressed 
(for max. 15 s). 


3. The software version, date, and user language are displayed 
in the LCD display. 


4. Theunitis switched on when the ON/OFF button is released. 
5. "*appears in the LCD display if the service connector is 


plugged. 
The following conditions are activated: 


- Operating alarms are muted. 


- All special functions are accessible (including the disabled 
ones). 


- Special functions are slightly modified. (Example: SM battery 
capacity has keys for 0 min/5 min). 


Contrast Setting 


1. Select "Contrast Setting" with SM key. The softkey symbols 
"clock" and "ml/h" are flashing. 


2. Set display contrast with the (+) or (-) key. 


3. Return to main menu with END. 


Mechanical Pressure Control 


Switch off the electronic pressure monitoring to check the me- 
chanical pressure control. 


1. Select "Switch-off pressure" with SM key. 
2. Then select "mechanical". 


3. Return to main menu with END. 


Infusomat® fm 2.2 gb 


Start / Quit the Service Program 





Service Program Version 
Group Description 
Function Description 


END GR+ 





Infusomat? fm 2.2 gb 


FU+ 


OK 


Service Program 


Activate the Service Program 


1. 


Plug service connector on staff call socket at the rear of the 
unit. - ** appears in the display. 


Select "Service Program" with the SM key. When the Service 
Program is activated the red alarm LED flashes. The LED dis- 
plays the code number of the selected group and function. 


Soft key functions 


END Jumps to the initial function 

GR+ Selects group 

FU+ Selects function in the 
activated group 

OK Activates the selected function or if 
necessary skips to sub-functions 
with NEXT 


Note 


When the service connector is plugged in the audible alarm is de- 


activated and only active during the switch-on test. 


Quit the Service Program 


1. 


Press END in the main menu. - Å data storage query is acti- 
vated: "Save changes? YES/NO". 

Modified values are only saved with the YES key. 

YES/NO terminates the Service Program. Press END to jump to 
the last function. 


Switch off the Infusomat fm and remove service connector. 


Note 


Disconnect the Infusomat fm from mains for at least 30 seconds. 


Then the unit can be switched on again. 


Service Program 


Unit Data 





Drug No. 


Character 
END (6 A }) 








Software Version Function 100 
1. Select sub-functions with NEXT. 


2. The current software version is displayed in the LCD display: 
- User program version with date 
- Language with text version. Further languages with (+) 
- Service Program version 
- Service language with text version 


3. Return to initial function with END. 


Drug Name Function 110 
Memory for maximum 10 drugs and 20 characters per name. 

1. Display stored drug names with NEXT key. 

2. Delete displayed entry with CLR. 


3. Press YES to modify a drug name: 
Move cursor to character with NEXT. 
Select new character from line 3 with «« or »». 


4. Repeat procedure for each character. 


5. Return to initial function with END. 


Ward ldentification Function 120 


Enter and display of a ward specific unit identification. Permanent 
display if the unit is connected to mains and switched off. 


1. Delete displayed entry with CLR. Press YES to enter modifica- 
tions: 
Move cursor to character with NEXT. 
Select new character from line 3 with << or >>. 


2. Repeat procedure for each character. 


3. Return to initial function with END. 


Infusion Line Function 130 
(Tube Adapter) 


The fitting position of the tube adapter is displayed. 
Check that display and position of the adapter correspond 
after having exchanged the Reed sensors. 

1. OKactivates the display of the adapter position. 

2. Return to initial function with END. 


Serial Number Function 140 


The displayed serial number must correspond with the number on 
the unit type plate, as this number is used in CC mode. 


1. YES activates the entry mode. Enter via the numeric key- 
board. 


Infusomat® fm 2.2 gb 


History Data 


Bit 154. H2 M Mal DÊS RR 6. Ont SK ge) 
Code 1 0 0 0 0 0 0 O0 0 0 0 O0 O0 0 1 0 


| | 


Example for alarm "Pump cover open" 


Battery empty (battery alarm) 


Pump cover open 
Drop alarm 

Air alarm 

Pressure alarm 
Expired standby time 
CC alarm (interface) 
KOR end alarm 

free (insignificant) 


Operating alarm 


Infusomat? fm 2.2 gb 





Service Program 


2. YES stores the changed or new number. 


3. Return to initial function with END. 


DIANET Type Number Function 150 


The displayed serial number must correspond with the number on 
the unit type plate, as this number is used in CC mode. 


1. YES activates the entry mode. Enter via the numeric key- 
board. 


2. YES stores the changed or new number. 


3. Return to initial function with END. 


Operating Hour Counter Function 200 
1. OK activates the display. 
2. Return to initial function with END. 


Battery Operating Hours Function 210 
1. OK activates the display. 
2. Return to initial function with END. 


Pump Head Cycles Function 220 
Display of the pump head cycles (delivered volume). 

1. OK activates the display. 

2. Return to initial function with END. 


Operating Alarms Function 230 
The last 20 operating alarms can be recalled. 


They are displayed as 16 bit binary codes and each bit position can 
be set from O to 1. 


1. OKactivates the alarm display. 
2. Display operating alarms 01 to 20 with the (+) and (-) key. 


3. Delete the operating alarms with CLR (from software 
IFMBO1001 or IFMBOOO! on). 


4. Return to initial function with END. 


Alignment 


Service Program 


Air Inline Sensor Function 300 
After exchange check function of the air inline sensor. 


See TSC list for permissible check values. 


For software IFMAO0002 and IFMAO0003 
1. Press OK. "Observe Service Manual" is displayed. 
2. Acknowledge with OK. 


3. Insert an infusion line filled with air and check the maximum 
permissible air value. 


4. Insert an infusion line filled with fluid and check the mini- 
mum permissible water value. 


5. Return to initial function with END. The alignment which is 
possible in this software version must not be performed. 


From software IFMBOOO01 or IFMBO1001 on 


6. Press OK. The received signal amplitude is displayed as a 
measured value. 
(The test value with NEXT is not important). 


7. Insert an infusion line filled with air and check the maximum 
permissible air value. 


8. Insert an infusion line filled with fluid and check the mini- 
mum permissible water value. 


9. Return to initial function with END. 


Pressure Control Function 310 
Test equipment: 4 mm gauge 
(see „Test Equipment and Special Tools" = p. 9 - 1) 
1. Push in bottom slide of the finger pump. 
Press OK twice. 
Open unit door. 
Note the OIL value (actual value). 


Insert 4 mm gauge and close the unit door. 


oe SH. ie. GER de 


The new OIL value is displayed. It must be 5 to 15 digits high- 
er than the first value. 


7. Return to initial function with END. 


If the 5 to 15 digits are not reached, the pressure sensor unit must 
be mechanically adjusted (see „Pressure Sensor" = p. 4 - 18). 


Infusomat® fm 2.2 gb 


Service Program 


Unit Modifications Service Language Function 400 
English or German can be selected. 


1. OKactivates the function. 


2. Select language with NEXT. 

3. Acknowledge with YES. 

4. Return to initial function with END. 

User Language Function 410 


Four user languages per language group are available (depending 
on software). 


1. OKactivates the function. 


2. Select language with NEXT. 
The language no. and text version are displayed. 


3. Acknowledge with YES. 
4. Return to initial function with END. 


Alarm Tone Function 420 


Different alarm modes can be selected: 


Software IFMAO0002 and IFMAO0003 
- Single stage: continuous tone 65 dBA. 


- Double stage: alarm with reduced volume (50 dBA) for the 
first 2 minutes, then 65 dBA. 


From software IFMBOOO01 or IFMBO1001 on 
- An additional "10 minutes off alarm" can be selected. 


- In this mode the audible alarm is activated with a delay of 10 
minutes. 


-  Theactivation of the "10 minutes off alarm" is only permissi- 
ble if the staff call is connected and the Infusomat fm has an 
attention label (label drawing no. M007100000F04]). 


1. OK activates the function. 

Select alarm tone with NEXT. 
Acknowledge with YES. 

Return to initial function with END. 


T* ÊR a 


Infusomat? fm 2.2 gb 3- 7 


Service Program 


Unit OFF 


Initialization Test | 


Optical 
Alarm 


Alarm ON 
Alarm OFF 


Audible 
Alarm 


Staff Call 


Alarm ON 
Alarm OFF 


Unit ON Unit OFF 


Operation Alarm 


Alarm OFF 


| 2 min. 


Dynamic with OFF-Alarm (permanent contact) | 


. Static with OFF-Alarm (permanent contact) | 


Dynamic without OFF-Alarm 


` Dynamic with OFF-Alarm 


. Static without OFF-Alarm 





Staff Call 
Different staff call modes can be selected: 


Software IFMA00002 and IFMAO00003 


Function 430 


- Dynamic with OFF alarm (permanent contact). 
- Static with OFF alarm (permanent contact). 
From software IFMB00001 or IFMBO1001 on 

- Dynamic without OFF alarm 

- Dynamic with OFF alarm. 


- Static without OFF alarm 


The self-test of the staff call relay during switching on has been 
cancelled. For further details see staff call line in the instructions 
for use. 


1. OKactivates the function. 

2. Select staff call type with NEXT. 

3. Acknowledge with YES. 

4. Return to initial function with END. 


Special Functions Function 440 ** 


Only available from software version IFMB 00001 / IFMB 01001 
On. 


Special functions can be activated in the Service Program, which 
are then displayed on the user interface. Deactivated special 
functions will not be displayed. The SM softkey will not be dis- 
played in standard operation when all special functions are disa- 
bled. - Special functions to be selected: 


- Standby function 

- Drug selection 

- CC mode 

-  Switch-off pressure 
- Drop control 

- Battery capacity 

- Data lock 


- Contrast setting 


1. OKactivates the function. 

2. Select special functions with NEXT. 

3. Activate / deactivate the respective function with YES/NO. 
* software version IFMAO0002 and IFMAO0003 


** from software IFMBOO0001 or IFMBO1001 on 
*** from software IFMBO0002 on 


Infusomat? fm 2.2 gb 


Infusomat? fm 2.2 gb 


Service Program 


4. Return to main menu with END. 


Decimal Function Function 450 ** 


Only available from software version IFMB 00001 / IFMB 01001 
on. 


1. OK activates the function. 


2. Activate / deactivate the decimal function with NEXT. 

3. Acknowledge with YES. 

4. Return to main menu with END. 

Delivery Rate min/max Function 460” 


The maximum and minimum delivery rates can be set. Range of 
adjustment of the delivery rate: 0.1 to 999.0 ml/h 


1. OK activates the function. 


2. Select min./max delivery rate with NEXT. 

3. Acknowledge with YES. 

4. Enter value with a numeric key. 

5. Acknowledge with OK. 

6. Return to initial function with END. 

Air Alarm Function 470*** 


The air inline sensor sensitivity of the air rate alarm in ml/h (total 
air alarm) and of the maximum air bubble in ml (single bubble) 
can be adjusted. 

Setting range air rate: 0.5 to 3.5 ml/h 

Setting range air bubble: 0.01 to 0.3 ml/h 


1. OK activates the function. 

Select air rate (ml/h) or air bubble (ml) with NEXT. 
Acknowledge with YES. 

Enter value with a numeric key. 

Acknowledge with OK. 

Return to initial function with END. 


ØRE SER. ee USO d 


* software version IFMAO0002 and IFMA00003 
** from software IFMBOO001 or IFMBO1001 on 
**% from software IFMBOO002 on 


Service Program 


Calibration 


3- 10 


Note 


All safety relevant parameters are set by the manufacturer. If 
these parameters are changed, a new calibration must be per- 
formed with calibrated test equipment. 


Note 


Only enter the scale factor values and pressure sensor values for 
OIL (Original Infusomat Line). Values for OIL-FM (Original Infuso- 
mat Line FM) are not applicable. This infusion line is no longer 
avallable. 


Pump Head Function 500" 
1. Software IFMA: 


Software module "Calibration Pump Head" for the entry of 
scale factor and pressure sensor data. 


2. From software IFMBOOO01 or IFMBO1001 on: 


The entry groups pump head and pressure sensor are separate 
functions to avoid operating faults. 


- A Calibration pressure sensor: function 500 


- Calibration scale factor: function 510. 


Pressure Sensor Data Function 500** 


Calibration (see „Pressure Sensor" = p. 4 - 18). 


Scale Factor Function 500* or 510** 


The scale factor can be set in the limits 40 to 99 digits. Every digit 
step is equivalent to a 0.5 % modification of the delivery rate. An 
increase of the scale factor reduces the pump speed, and a de- 
crease increases the pump speed. 


1. OKactivates the function. 

2. The value can be changed with the entry keyboard. 

3. Acknowledge with YES. 

4. Return to initial function with END. 

5. Quit the Service Program and save changes with YES. 
6 


Switch on unit and check delivery rate (see „General Pres- 
sure Measurement" c» p. 8 - 3). If necessary repeat the de- 
livery rate measurement. 


* software version IFMAO0002 and IFMAO0003 
** from software IFMBOO001 or IFMBO1001 on 
*** from software IFMBO0002 on 


Infusomat® fm 2.2 gb 


Service Program 


Air Inline Sensor Function 510” or 520” 


Alignment or check of the air inline sensor threshold value (see 
„Air Inline Sensor" x» p. 4 - 20). 


1. OK activates the function. 

Press OK again to call in the air inline sensor value. 
The value can be changed with the entry keyboard. 
Acknowledge with YES. 

AIR INLINE SENSOR IS SET acknowledges the entry. 
Return to initial function with END. 


N. e GP dr» ES Iw 


Quit the Service Program and save changes with YES. 


* software version IFMAO0002 and IFMAO0003 
** from software IFMBOO001 or IFMBO1001 on 
** from software IFMBOO002 on 


Infusomat? fm 2.2 gb 3- 11 


Service Program 


For your notes: 


3- 12 Infusomat? fm 2.2 gb 


4.1 Mains Fuses 


4.2 Battery 


Battery 


Infusomat® fm 2.2 gb 





Unit Elements 4 


Designation Ord. No. 

Fuse T 0.16 A for 200-240 V (10 pcs.) ............. 3477 2847 

Fuse T 0.315 A for 100-120 V (10 pcs.) ............ 3477 0534 

Fase BOISE ee a wa wa ui ê owns ol anê HOD goed ûnê Re bai 3450 5652 
Note 


Only use recommended fuses. 


Exchange 


1. Press the expansion clamps at the fuse holder on the recessed 
mains plug with a screw driver in direction of the arrows and 
pull out fuse holder. 


2. Replace blown fuses and press in fuse holder. 


Check 


Electrical safety, functional check. 


Applicable types: Long time battery 1.8 Ah, short time battery 0.5 
Ah 


Designation Ord. No. 
Battery incl. connector plug, 1.8 Ah / 7.2 V......... 3450 6357 
Cover for battery compartment .................. 3450 5504 
Exchange 


Tools: Pointed pliers 

1. Switch off unit and disconnect from mains. 
Open cover of the battery compartment. 
Remove battery from the battery compartment. 


2 
3 
4. Pull off battery plug with pointed pliers. 
5. Assembly is done in reverse order. 

6 


After having exchanged the battery connect Infusomat fm to 
mains (thereby the charge and discharge current are bal- 
anced). 


7. Recharge battery (16 h). 


Check 


Perform switch-on test in battery operation and check the battery 
running time, if necessary. 


4.3 


Unit Elements 


Door Lock 


Door lock 





Designation Ord. No. 
Door lock complete with push button ............. 3450 5601 
Spring holder for door lock Kh. kk kk 3450 5440 
Fixture for door JOCKS ender denas uci edic act EE 3477 2790 
Exchange 


1. Open door and disassemble countersunk screw under the 
door lock button. 


2. Press fixture for the door lock forward and remove fixture. 


3. Pressspring holder down with a screwdriver through the bot- 
tom hole and unlatch. 


4. Pull door lock forward and lift door lock. 
5. Remove door lock in an upward direction and exchange. 


6. Assembly is done in reverse order. 


Check 


Pump unit check (only mechanical pressure control). 


Infusomat® fm 2.2 gb 


4.4 Pump Cover 





Infusomat® fm 2.2 gb 


Front frame 





Unit Elements 


Designation Ord. No. 
Pump cover with lock uada ae kK KK RR K> 3450 5717 
Blind plug 7.1 mm (10 pcs.) ..................... 3477 3207 
Torsion spring in lever / pump cover (5 pcs.) ....... 3477 3363 
Torsion spring for pump cover (5 pcs.) ............ 3477 3355 
Lever (pump COVER) agudo kirên wa kêy bts 3477 4092 
Hinge pin for pump'COVer z ciues KK KK KK KK KK KK ER 3477 3967 
Hinge pin for pump cover lever..................… 3450 5725 
Exchange 


Tools: Pin punch 1.8mm, pin punch 6mm, 4mm gauge 


1. Open door and remove hinge pin with pin punch (1.8 mm) 
from below. Do not lose torsion spring. 


2. Disassemble pump cover. 

3. Insert the torsion spring in new pump cover and assemble. 

4. Press hinge pin from the top in the holder with a pin punch. 

5. Check pressure sensor with 4mm gauge in the Service Pro- 
gram, and calibrate, if necessary. 

Check 


Electrical safety, pump unit check. 


4.5 


Unit Elements 


Housing 


S M4x12 





Designation Ord. No. 
Housing labelling 
German ........... see rns 3450 5610 
French uoces E a al d er deuce a a anû lak a da 3450 5946 
DUC EE 3450 5954 
ltalian ............ceee RR RII 3450 5962 
anc Mee ———— Hr 3450 5970 
sec PT""-————— 3450 5989 
Danish s.d RETE RETE mas 3450 5997 
Norwegian `. 3450 6101 
Swedish ........ kk RR 3450 6110 
salı CHE 3450 6128 
POrtUGUCSE dts a ye au HRK pax eee Kinê ê êna abê 3450 6136 
deld METTRE 3450 6144 
eli M PTT TCR 3450 6152 
Cäastellang,,,,,,, naan 3450 6160 
WOMENS) EEN 3450 6179 
Label for short instructions for use ............... 3450 8651 
Foot stand complete with rubber feet............. 3450 5415 
Rubber feet (20 pes.) as eg ENEE dx 3477 3096 
Exchange 


1. Remove battery (see ,,Battery" > p. 4 - 1). 


2. Turn unit upside down and remove the 4 screws from the foot 
stands. Remove foot stands. 


3. Place unit upright, remove safety seal from the rear panel, 
and break the tamper-proof cap. 


4. Remove countersunk screw and serrated lock washer (rear 
side). 


5. Slightly widen the sides at the bottom of the housing and pull 
off to the top. Do not damage the microprocessor board. 


6. Assembly is done in reverse order. 


7. Safety seal the rear panel screw after functional check. 


Check 


Electrical safety. 


Infusomat® fm 2.2 gb 


4.6 Handle 


If necessary seal 
additional 

hole with 
adhesive tape 


M6x45 
with serrated lock 


Infusomat® fm 2.2 gb 





Unit Elements 


Designation Ord. No. 
nri >>e*hhh e ebose ne nua 3450 5512 
Oka (20 D65.] 4^ „s2x salê oya û wêd n sê kê wn kayê as 3477 1530 
Exchange 


1. 


Remove battery (see „Battery" > p. 4 - 1). 


2. Dismount cover (see ,, Housing" 3 p. 4 - 4). 

3. Remove the screw from handle/housing. 

4. Assemble new handle. Be sure that the knobs are seated in 
the guides. Seal screw-holes with O-ring. 

Check 


Electrical safety. 


4.7 


Unit Elements 


Microprocessor Board 


Microprocessor board 





Designation 


Microprocessor boards with different software versions can be 
used. The user must be informed about the modified software ver- 
sions (see „Software Update" r> p. 2 - 1). 


The boards with the raw material no. 3810 6221”, 3810 6914", 
3810 6973 and 3810 7228 have different membrane plugs. Please 
pay attention when ordering boards. 


* Raw material no. 3810 6221 and 3810 6914 only as exchange parts. If you wish 
to order new parts please use raw material no. 3810 6973 or 3810 7228. 
Exchange 
1. Remove battery (see ,Battery" 3 p. 4 - 1). 

Dismount cover (see ,, Housing" 3 p. 4 - 4). 

Loosen countersunk screw and bridge. 


Remove both screws of the board. 


2 
3 
4 
5. Carefully swivel out microprocessor board from the guide. 
6. Pull off plugs E1 to E5 and disconnect the flexible cable. 
7. Exchange the microprocessor board. 

8. Assembly is done in reverse order. 


Do not damage the optical components. Make sure that the rear 
panel guide and the fm recessed plug are correctly seated. 


Note 


Inform the user if functions have been modified due to the ex- 
change of the board (software IFMAOOxxx / IFMBOOxxx, see soft- 
ware coding), exchange the instructions for use, if necessary. 


Check 


Electrical safety, functional check. 


Infusomat® fm 2.2 gb 


Exchange 


Å Ord.No.:3488 045B 


Table 4 - 1 





Language Group” New Part Exchange 
Ord. No.:3450 6233 |Ord.No.:3488 0607 
Bo Ord. No.:3450 6195 |Ord.No.:3488 0569 


Ord. No.:3450 6209 |Ord.No.:3488 0577 
D Ord. No.:3450 6217 |Ord.No.:3488 0585 
Ord. No.:3450 6225 |Ord.No.:3488 0593 


Table 4 - 2 


Infusomat® fm 2.2 gb 





Unit Elements 


Microprocessor Board 
(raw material no. 3810 6221 replaces raw material no. 3810 
7228) 


Delivery with: 
- Membrane plug (Dupont blue) 
- Software IFMAO0xxx (update to IFMBOOxxx possible). 


Compatible with display board, order no. of display boards (see 
„Display Board" = p. 4 - 25) 
(raw material no. 3810 6205) 


Microprocessor Board 
(raw material no. 3810 6973 replaces raw material no. 3810 
6914) 


Delivery with: 
- Membrane plug (AMP black). 
- Software IFMBO1xxx. 


Compatible with display board, order no. of display boards (see 
„Display Board" = p. 4 - 25) 
(raw material no. 3810 6906, AMP label and flex cable). 


* Language Group: 

A German, French, Dutch, Italian 

B English, Dutch, Spanish, Castellano 

C Danish, Norwegian, Swedish, Finnish 

D Spanish, Portuguese, English (BSI), English, Turkish 
E Czech, Polish 


4.8 


Unit Elements 


Analog Board 


lim 017 


Analog board 





Designation Ord. No. 

Analog board new part... kk kk RR RR KK KS 3450 580B 

Analog board exvchange huk kk kk kK KK RRA KS 3488 047B 

Fuse T 1.6 Aon analog board ...................… 3477 3312 
(10 pieces) 

Exchange 


1. Remove battery (see ,,Battery" 3 p. 4 - 1). 
2. Dismount cover (see ,, Housing" 3 p. 4 - 4). 


3. Disassemble the microprocessor board (see „Microprocessor 
Board" r> p. 4 - 6). 


4. Remove screw with the serrated lock washer on the analog 
board. 


5. Disconnect the 3 plugs from motor, rear panel wiring and 
drop/air inline sensor connections. 


6. Exchange the analog board. 
7. Position Pertinax cover over the battery socket. 


8. Assembly is done in reverse order. 


Check 


Electrical safety, functional check. 


Infusomat® fm 2.2 gb 


4.9 Rear Panel 





Infusomat? fm 2.2 gb 


Unit Elements 


Designation Ord. No. 
Rear panel without accessories .................- 3450 5628 
Cover for optical interface ...................... 3477 3164 
Sealing strip for rear panel .....................- 3477 3142 
Exchange 


1. Tools: Special socket spanner M18 
Remove battery (see , Battery" > p. 4 - 1). 


Dismount cover (see , Housing" r> p. 4 - 4). 


p pee I9 


Disassemble microprocessor board (see „Microprocessor 
Board" r> p. 4 - 6) and analog board (see „Analog Board" 
r> p. 4 - 8). 

5. Turn unit to one side. Remove 2 screws with serrated lock 


washers from the bottom side. 


6. Place unit upright. Swivel rear panel aside and remove the 
wires (brown, white, green, yellow) out of the plug (do not 
damage the crimp terminals). 


7. Disassemble all components from the rear panel. 


8. Exchange rear panel. Assembly is done in reverse order. Safety 
lock the transformer screws with Loctite 242 e. 


Check 


Electrical safety, functional check. 


Unit Elements 


4.10 Drop Sensor Socket 


ws [ss 
br+5VTP2 Transmitter Signal TP1 


gn TSE 
Receiver Signal e GND 
TP3 P4 


Shielding or not connected 
internal View on Socket 


WS ge free 


UTS Receiver Signal 
P5, View Component Side 


4- 10 





Designation Ord. No. 
Drop sensor socket incl. cable and plug............ 3450 5687 
Exchange 


1. Remove battery (see ,Battery" © p. 4 - 1). 
Dismount cover (see ,, Housing" = p. 4 - 4). 


Disassemble rear panel (see ,,Rear Panel" c» p. 4 - 9). 


qe m e 


Remove 4 wires (brown, white, green, yellow) out of the plug 
(do not damage the crimp terminals). 


Loosen nut and exchange the drop sensor socket. 
Safety lock nut and socket (Sicomet 50). 


Connect the 4 wires. 


p JN. £p oí 


Make connection to ground (at the fixture of the mains 
transformer). 


9. Assembly is done in reverse order. 


Check 


Electrical safety. 


The connection cable of the drop sensor from serial no. 15626 is shielded. If the 
shielded version is retrofitted check that the drop sensor socket (middle pin) is 
connected with the rear panel. Retrofit if necessary (compare with point 8). 


Infusomat® fm 2.2 gb 


4.11 Rear Panel Board 


Rear panel board 


Fig.: 4 - 10 


Infusomat® fm 2.2 gb 





Unit Elements 


Designation Ord. No. 
DIANET recessed plug (8 pin) .................... 3450 5679 
Staff call recessed plug (5 pin).................…. 3450 5660 


Rear panel board 
with 5 and 8 pin recessed plugs 


with 5V Tllmtertaee,,,,, 3450 5636 
with RS232 interface ................. ses 3450 563A 
Exchange 


Tools: Special socket spanner M18 

1. Remove battery (see Battery" 3 p. 4 - 1). 

Dismount cover (see , Housing" 3 p. 4 - 4). 
Disassemble rear panel (see „Rear Panel" = p. 4 - 9). 
Remove screw on the rear panel board. 

Remove crimp terminals (red/blue). 


Remove nuts from the fm recessed plug (red/blue). 


= OX Sb dris e Se 


Remove coupling ring from the staff call plug and the 12V re- 
cessed plug (socket spanner M18). Exchange rear panel board 
or the DIANET/staff call recessed plug. 


8. Assembly is done in reverse order. 


Supplementary Information 


The DIANET interface (PC connection) is designed as a 5V TTL in- 
terface until unit No. 17147. From unit No. 17148 an RS232 inter- 
face is present. Older units can be retrofitted by exchanging the 
rear panel board. 


Please pay attention to the microprocessor board model when 
retrofitting a rear panel board with an RS232 interface: 


- Modification not necessary with microprocessor board order 
no. 3810 6221 index C. 


- Perform retrofitting with microprocessor board order no. 
3810 6221 until index B (see „3. Modification of the Inter- 
face from 5V TTL to RS232" 3 p. A - 2) or exchange board 
by a current type. 


Check 


Electrical safety, functional check. 


4- 11 


Unit Elements 


4.12 fm Recessed Plug 


Potential 
equalization 


fm recessed plug 


2x M4 each 


Fig.: 4 - 11 


4.13 Potential Equalization Bolt 


4- 12 





Designation Ord. No. 
fm recessed plug (3 pin) .......................… 3477 3177 
Exchange 


1. Remove battery (see ,Battery" © p. 4 - 1). 

2. Dismount cover (see , Housing" > p. 4 - 4). 

3. Disassemble rear panel (see „Rear Panel" œ p. 4 - 9). 
4 


Remove nuts from the fm recessed plug (green/yellow, red, 
blue). 


Remove screw with washer. 


p 


6. Slightly bend recessed plug to the top and pull out to the rear. 


7. Assemble new recessed plug. After assembly the plug must be 
slightly moveable. 


8. Assembly is done in reverse order. 


Check 


Electrical safety. 


Designation Ord. No. 
Potential equalization bolt ....................…… 3477 0550 
Exchange 


1. Remove battery (see , Battery" 3 p. 4 - 1). 

2. Dismount cover (see ,, Housing" 3 p. 4 - 4). 

3. Disassemble rear panel (see „Rear Panel" c» p. 4 - 9). 
4 


Remove nuts with ring spanner. Exchange bolt (see above fig- 
ure). 


Check 


Electrical safety. 


Infusomat® fm 2.2 gb 


4.14 Recessed Mains Plug 


4.15 Mains Transformer 


Mains transformer 


Infusomat® fm 2.2 gb 








Unit Elements 


Designation Ord. No. 
Recessed mains plug without fuse holder.......... 3450 5644 
RI el PME" 3450 5652 
Exchange 


1. Remove battery (see , Battery" > p. 4 - 1). 

Dismount cover (see , Housing" 3 p. 4 - 4). 
Disassemble rear panel (see „Rear Panel" = p. 4 - 9). 
Remove shrink tube from the cable terminals. 


Unsolder the connection cables (L/N/SL). 


Cr. Qu a" e Ui 


Loosen countersunk screws and exchange recessed mains 
plug. 
7. Install new shrink tube. 


8. Solder the cables and insulate with shrink tube. Make sure 
that the terminal assignment is correct. 


9. Assembly is done in reverse order. 


Check 


Electrical safety. 


Designation Ord. No. 
Mains transformer 200-240 V .............00000- 3450 5709 
Mains transformer 100-120 V...................… 3450 5695 
Exchange 


1. Remove battery (see Battery" > p. 4 - 1). 


2. Dismount cover (see ,, Housing" 3 p. 4 - 4). 

3. Disassemble rear panel (see „Rear Panel" œ p. 4 - 9). 

4. Remove shrink tube from the transformer cable terminals and 
unsolder the connection cables. 

5. Loosen screws with serrated lock washers and washer. 


6. Disassemble the mains transformer. 


7. Assemble new mains transformer. Safety lock the 4 screws 
with Loctite 242e! 


8. Install new shrink tube. Solder the connection cables and in- 
sulate with shrink tube. 


9. Assembly is done in reverse order. 


Check 


Electrical safety, functional check. 


4- 13 


Unit Elements 


4.16 Reed Sensor Set 


4321 
Operaterunit 


New 


4321 


Operaterunit 


Old 
Front frame 


Fig.: 4 - 14 


4- 14 





Designation Ord. No. 
Reed sensor set 
Sensor incl. 8 pin socket plug ................ 3450 5814 
Reed sensor incl. crimp terminals ............. 3450 5431 
Socket plug (AMP/100 pin) .................. 3451 3744 


Exchange 

1. Remove battery (see ,Battery" > p. 4 - 1). 

2. Dismount cover (see ,, Housing" © p. 4 - 4). 

3. Disassemble front frame (see „Front Frame" > p. 4 - 15). 
4. Press out Reed sensor set from the holder and exchange. 
5 


Assembly is done in reverse order. Do not squeeze the cables. 


Note 


The Infusomat fm is only equipped with the OIL (Original Infuso- 
mat Line) from serial no. 33202 on. The Reed sensors for OIL and 
OIL-FM are not applicable and are replaced by a bridge from pin 
2a and 2b. 


Check 


Electrical safety, check of pump unit, tube type test in Service Pro- 
gram (only with the tube adapter version). 


Infusomat® fm 2.2 gb 


4.17 Front Frame 


d 


Fig.: 4 - 15 


Front frame 


Infusomat® fm 2.2 gb 


Tube adapter 


Distance sleeve 





Unit Elements 


Designation Ord. No. 
Front frame without flow inhibitor 
and pressure Spring... 3450 5822 
with distance sleeve and countersunk screw 
Circular seal behind front frame ................. 3477 3126 
Flow inhibitor with pressure spring ............... 3477 3258 
Pressure spring for flow inhibitor (5 pcs.) .......... 3477 3266 
Tube adapter with magnet ...َ. a KS 3450 5830 
Magnet for tube adapter `... 3450 5849 
Exchange 


Tools: Pin punch 

1. Remove battery (see Battery" © p. 4 - 1). 

2. Dismount cover (see ,, Housing" x» p. 4 - 4). 

3. Disassemble pump cover (see „Pump Cover" = p. 4 - 3). 
4 


Push the tube adapter in arrow direction and remove (only up 
to serial no. 33 201). 


5. Remove tamper-proof caps by piercing a screwdriver through 
the caps. 

6. Loosen countersunk screws. 

7. Remove front frame by pressing the snap-in pins inwards on 


the outer side of the frame (rear side) and pull off frame to 
the front. 


8. Remove flow inhibitor with pressure spring and Reed sensors 
and assemble in new front frame. 


9. Assembly is done in reverse order. Make sure that the tube 
adapter is correctly assembled and that the cable run of the 
Reed sensors is correct. Snap-in pins manually. 


Note 


Up to serial no. 5878: Exchange complete front frame with flow 
inhibitor. - Up to serial no. 33 201 (from 10/96): Use distance 
sleeve (17.5 mm) and countersunk screw M4x25 when the front 
frame is exchanged. Remove OIL and OIL-FM sensors, plug bridge 
in connector position 2a-2b (see „Reed Sensor Set" © 
p. 4 - 14). 


Check 


Electrical safety, pump unit check. 


4- 15 


Unit Elements 


4.18 Pump Unit 





























4- 16 









































Pump unit 


lim 032 


Designation Ord. No. 


Finger pump (without motor) including pump, ..... 3450 5407 
pump cover, seal membrane and boards 


Finger pump (without motor and board) ........... 3450 9038 
incl. pump, pump cover, 
and seal membrane 


Membrane for pump unit WW KK RR RRA 3450 5733 
Motor with pinion sacar ng kk KK KK KK KK ER REOR A 3450 5741 
Exchange 


1. Remove battery (see , Battery" > p. 4 - 1). 

2. Dismount cover (see ,, Housing" 3 p. 4 - 4). 

3. Disassemble front frame (see „Front Frame" > p. 4 - 15). 
4 


Loosen motor screws, pull off motor connector and remove 
motor from pump unit. 


Note 


Do not loosen the mounting screw on the lower pressure sensor 
board! Otherwise the pump has to be recalibrated with a 4mm 
gauge. 


5. Pull off connector from the microprocessor board. 


6. Press down snap-in pins (from the inside) and tilt pump for- 
ward. Slightly pull the pump unit up and unhinge the bottom 
side. 


7. Tilt to front and pull out the unit. 


8. Assembly is done in reverse order. The ribbon cable must not 
hinder the pump movement. 


9. Enter user data in the EEPROM (see below). 


The complete pump unit (pump without motor, pump cover, mem- 
brane, boards) was calibrated by B.Braun. After a complete ex- 
change of the pump unit the unit data and the user data must be 
entered in the Service Program again as the data memory is on the 
pump board: 


- Serial number 

-  Dianet type no. 

If necessary enter: 

- Drug name 

- Special functions (ON/OFF)* 
- Decimal function (ON/OFF)* 


Infusomat® fm 2.2 gb 


Infusomat? fm 2.2 gb 


Unit Elements 


Delivery rate min./max. ” 
Air alarm” 

Ward identification 
Alarm tone 

User language 

Staff call 


If data is not entered, "Calibration data faulty" may be displayed 


after the unit is switched on again. 


After terminating the Service Program save the data. 


The counters for operating hours, battery operation and pump 


head cycles are reset to zero when the pump unit is exchanged. 


If "Calibration data faulty" is displayed, select the Service Program 


and quit with "Save? Yes". 


Check 


Electrical safety, pump unit check. 


* from software IFMBO0001 / 01001 on 
** from software IFMBO00002 / 01002 on 


4- 17 


Unit Elements 


4.19 Pressure Sensor 


Light barrier 
board 


Screw (1) 








washe ss 2 


NG y j 
NZ, 
Nes 


lim 044 


/ Serrated lock washer 
Screw (2) 


Pressure sensor with 


Do not loosen screw pressure sensor board 


Fig.: 4 - 18 





4- 18 


Test Equipment Ord. No. 

Pressure calibration device ...................... 0770 5018 

Calibration gauge 4 mm (for adjustment after ...... 0770 1489 
exchange of the pump cover) 

Designation Ord. No. 

Pressure sensor (cpl.) 1222s es ENNEN ENEE Ria 3450 7345 


(only delivered when a pressure 
calibration device is available) 


Exchange 


CAUTION 


The pressure sensor unit is mechanically very sensitive. It must not 
be dismounted, which means that the pressure sensor board and 
the spring element must not be detached from the pressure sensor 
support (plastic part). Make sure that the movement of the coil 
core inside the coil is not hindered. 


The pressure sensor consists of the light barrier board with EEP- 
ROM and the pressure sensor board with holder. Coil core and 
bending element are screwed to this holder. Both boards are con- 
nected via a flat cable. 


1. Dismount rear panel (see „Rear Panel" r> p. 4 - 9) or pump 
unit (see „Pump Unit" 3 p. 4 - 16). 


2. Pull off the connection cable to the microprocessor board. 
3. Loosen and remove screw (1) of the light barrier board. 


4. Loosen screw (2) of the pressure sensor and remove it togeth- 
er with plain washer and serrated lock washer. Then remove 
pressure sensor with pressure sensor board and light barrier 
board. 


5. Mount new pressure sensor. 


Note 


Make sure that the pressure sensor slide can smoothly run in the 
guides and that the coil core can move inside the coil. 

The serrated lock washer must be mounted, if necessary retrofit 
(see Fig.: 4 - 18). 


6. Adjusting the pressure sensor 


a) Call the pressure display in the Service Program (see 
,Alignment" © p. 3 - 6). 


b) Loosen hexagon socket screw (M 3x6) on the pressure 
sensor unit (with a 2.5 mm Allen key). 


Infusomat® fm 2.2 gb 


Infusomat? fm 2.2 gb 


Unit Elements 


c) Push the pressure sensor unit with board slightly forward 
or backward. 


d) Tighten the Allen screw. 


e) The new OIL value will be displayed (repeat if necessary 
until the value is increased by 5 to 15 digits). 


f) Return to the initial function with END. - Select SAVE NO. 
(Do not actuate the YES key.) 


g) Switch off unit. 
7. Assembly is done in reverse order. 


8. Check the user data and reenter, if necessary, as data was 
saved in the EEPROM of the exchanged light barrier board. 


9. Calibrate pressure sensor: 


WARNING 


NEVER OPERATE THE UNIT WITH PLUGGED IN SERVICE CONNEC- 
TOR ON THE PATIENT. WHEN THE UNIT WAS OPERATED WITH 
PLUGGED IN SERVICE CONNECTOR IT IS TO BE SWITCHED OFF 
ONCE BEFORE ANY FURTHER USE. 


Note 


To calibrate the pressure sensor a pressure calibration device is re- 
quired. An alignment with infusion lines is not permitted. 


a) Clean the pump front side. 


b) Place the Infusomat? fm without mains lead and drop 
sensor and the service connector plugged in horizontally 
(front facing upwards) in the cellular packing of the cal- 
Ibration device. 


c) Call function 310 in the Service Program (see ,,Pressure 
Control Function 310" c» p. 3 - 6). 


d) Open pump cover. 


e) Fasten holders for the calibration weights under the 
pump cover and let snap in at the locking bow. 


f) Position weight 1 (100 g) for 400 mbar carefully on the 
pressure sensor. The weight must be placed without any 
friction in the holder. 


g) Note and confirm the measured value. 


h) Repeatsteps 7. and 8. with weight 2 (128 g) for 800 mbar 
and weight 3 (166 g) for 1200 mbar. Confirm the meas- 
ured value for 800 mbar. 


i) Remove the holder and do not save data. 


4- 19 


Unit Elements 


j) Call function 500 in the Service Program and confirm 
"Change calibration OIL" with "Yes". 


k) Enter the calibration values via the keyboard and ac- 
knowledge with "Yes". 


I) Quit function in the Service Program and save data. 
m) Register the changed pressure values in the unit book. 


n) Check the electronic pressure control (see ,,Electronic 
Pressure Control (Pressure Sensor)" = p. 8 - 3). 


Check 


Electrical safety, functional check. 


4.20 Air Inline Sensor The analog board must be additionally exchanged for boards 
without an index. There are modified calibration and check values 
for the air inline sensor. Enter the new calibration value for every 
unit in the Service Program under CALIBRATION AIR INLINE SEN- 


SOR. 

Designation Ord. No. 
Air inline sensor with crimp terminals ............ 3450 5750 
Exchange 


1. Remove battery (see ,,Battery" © p. 4 - 1). 


2. Dismount cover (see ,, Housing" 3 p. 4 - 4). 

3. Disassemble front frame (see „Front Frame" > p. 4 - 15). 
4. Press the complete sensor out of the frame. 

5. Remove terminals from the connector. 

6. Assemble cable to the connector (check tight seat). 

7. Assembly is done in reverse order. 

Check 


After exchange of the air inline sensor: 
- Check air value 


- Check water value 





- Check calibration value (alarm threshold), adjust if necessary 


Electrical safety, functional check. 


4 - 20 Infusomat® fm 2.2 gb 


4.21 Alarm Buzzer 


4.22 Air Insulation Board 


Infusomat® fm 2.2 gb 


Unit Elements 


Designation Ord. No. 
Buzzer incl. Seal ring «aces e eer kk kK KK RR RR KK e 3450 5776 
Exchange 


1. Remove battery (see Battery" 3 p. 4 - 1). 


2. Dismount cover (see , Housing" = p. 4 - 4). 


Check 


Electrical safety, functional check. 


Designation Ord. No. 
Air insulation board uk ak RR RR SK KIJ 3450 6187 
Exchange 


1. Remove battery (see ,,Battery" > p. 4 - 1). 
Dismount cover (see , Housing" 3 p. 4 - 4). 


Pull off connector P1 from board. 


ps p N 


Remove screws (2 M 3x16) with distance sleeve (12 mm) 
from the air insulation board, exchange the board. 


5. Assembly is done in reverse order. 


Check 


Electrical safety, function check, air value in the Service Program. 


4 - 21 


Unit Elements 


4.23. Frame with Seal Plate 


4- 22 


Seal plate 


Frame unit 





Designation Ord. No. 
Frame unit with seal plate hh kk 3450 5466 
Exchange 


1. 


2 
5 
4 


SS, i Qe. ØR 


9. 


Remove battery (see „Battery” 3 p. 4 - 1). 
Dismount cover (see , Housing" c» p. 4 - 4). 
Disassemble pump cover (see „Pump Cover" = p. 4 - 3). 


Disassemble the microprocessor board (see ,, Microprocessor 
Board" r> p. 4 - 6), analog board (see „Analog Board" > 
o. 4 - 8) and the rear panel. 


Disassemble front frame (see „Front Frame" = p. 4 - 15). 
Disassemble door frame (see „Door Frame" c p. 4 - 23). 
Disassemble pump unit (see „Pump Unit" c» p. 4 - 16). 


Disassemble air inline sensor (see „Air Inline Sensor" => 
p. 4 - 20). 


Exchange the frame unit with seal plate. 


10. Assembly is done in reverse order. 


Check 


Electrical safety, functional check, pump unit check. 


Infusomat® fm 2.2 gb 


4.24 Door Frame 


Aur 


lis 


5 





Infusomat® fm 2.2 gb 


Unit Elements 


Designation Ord. No. 


Door frame incl. pressure spring and magnet ....... 3450 5598 
without door lock 


Seal for door frame / flexible cable (5 pcs.) ........ 3477 3347 
giu c"—-————— SEERE 3450 5571 
Hinge pin for hinge unit ....................2..… 3450 5580 
Exchange 


1. Remove battery (see , Battery" > p. 4 - 1). 

Dismount cover (see , Housing" ™ p. 4 - 4). 

Loosen the countersunk screw (M 3x6) and bridge. 

Swivel out the microprocessor board. 

Disconnect flexible cable from the display board. 
Disassemble pump cover (see „Pump Cover" c p. 4 - 3). 
Disassemble front frame (see „Front Frame" I p. 4 - 15). 


Remove countersunk screws from hinge. 


e Qe d IY GUI afe e gw 


Disassemble door lock (see „Door Lock" I p. 4 - 2). 


. Pull out the door frame to the front and take care not to dam- 


O 


age the cables. 


11. Disassemble the door hinge pin. Pay attention to correct ori- 
entation during assembly. 


12. Remove tamper-proof caps (6 pieces) on the door frame by 
piercing a screwdriver through the caps. 


13. Loosen screws. 
14. Remove support plate with modules from the door frame. 
15. Exchange the door frame. Do not forget the seal. 


16. Assembly is done in reverse order. 


Check 


Electrical safety, functional check, pump unit check. 


4- 23 


Unit Elements 


4.25 Membrane Keyboard with Support Plate 


Display board 


PT 2.2x8 


LCD module 


lim 022 





4- 24 


Designation Ord. No. 
Membrane keyboard with support plate, .......... 3450 5547 


seal and blind plate 


Exchange 


1. 


Remove battery (see , Battery" x» p. 4 - 1). 


2. Dismount cover (see „Housing" x» p. 4 - 4). 

3. Loosen countersunk screw and bridge. 

4. Swivel out the microprocessor board. 

5. Disconnect flexible cable from the display board. 

6. Disassemble pump cover (see „Pump Cover" © p. 4 - 3). 

7. Disassemble front frame (see „Front Frame" = p. 4 - 15). 

8. Disassemble door lock (see „Door Lock" c» p. 4 - 2). 

9. Remove tamper-proof caps on the door frame by piercing a 
screwdriver through the caps. 

10. Loosen screws (PT 2.5x14). 

11. Remove support plate with modules from the door frame. 
Carefully pull out the flexible cable through the frame. 

12. Remove clamp from the display board. 

13. Remove connector from the membrane keyboard and the ca- 
ble to the LCD module. Remove display board from the sup- 
port plate. 

14. Disassemble the LCD module. 

15. Exchange the membrane keyboard with support plate. 

16. Assembly is done in reverse order. 

Check 


Electrical safety, functional check, pump unit check. 


Infusomat® fm 2.2 gb 


Unit Elements 


4.26 Display Board Designation Ord. No. 
Display board, raw material No.: 3810 6205 
Display board new part (Dupont blue) ......... 3450 555A 
Display board exchange (Dupont blue) ........ 3488 046A 
Display board, raw material No.: 3810 6906 
Display board new part (AMP black) .......... 3450 555B 
Display board exchange (AMP black) .......... 3488 046B 
(see „Microprocessor Board" r> p. 4 - 6) 
Clamp for display board (20 pcs.) ................ 3450 0448 
Exchange 


See membrane keyboard / support plate. 


CAUTION 


Type of the boards depends on the membrane connectors on the 
microprocessor board (see „Microprocessor Board" x» p. 4 - 6). 


Check 


Electrical safety, functional check, pump unit check. 


4.27 LCD Module Designation Ord. No. 
LED MOGUIC emu" 3450 5563 
Exchange 


1. Remove battery (see ,,Battery" 3 p. 4 - 1). 
2. Disassemble door lock (see „Door Lock" c» p. 4 - 2). 


oe 


Remove tamper-proof caps on the door frame by piercing a 
screwdriver through the caps. 


Loosen screws (PT 2.5x14). 
Remove support plate with modules from the door frame. 


Disassemble the LCD module. 


S4 oO QETÊ cub 


Assembly is done in reverse order. Set the contrast. 


Check 


Electrical safety, functional check. 


Infusomat® fm 2.2 gb 4- 25 


Unit Elements 


For your notes: 


4- 26 Infusomat® fm 2.2 gb 


1. Visual Inspection 


Û OK after visual inspection 


Infusomat? fm 2.2 gb 


Checks after Repair 


2. Safety Check 
according to IEC/EN 60 601-1 
or VDE 0751 

Mains voltage — VAC 
Protective conductor 
resistance 
incl. mains 
cable < 0.20 
Insulation resistance 
>> 2 MQ 
Earth leakage current 
<30 uA uA 


Patient leakage current 


< 5 pA — HA 


5 


Depending on the work carried out, carry out the respective check 


blocks (1., 2., 3 and / or 4.). 


3. Functional Inspection 


Switch on unit: 
1 Self-test 
O Control lamps 


Compare with display: 
Û Set delivery rate 


Battery test: 

O Switch mains/battery/ 
mains 

Û Switch on in battery mode 
and check self-test 


Air inline sensor: 

Û 0.1 ml air bubbles no 
alarm 
0.4 ml air bubbles alarm 
Air value 
Water value 
Check calibration value 
and if necessary enter 


Drop sensor: 
Simulate occlusion alarm 
(alarm with closed roller 


clamp) 


Simulate free flow (alarm) 
Staff call 
Alarm suppression 


4. Pump Unit Inspection 


Electronic pressure control: 

O Check alarm with switch- 
off pressure 
low / medium / high 


Mechanical pressure control 
max. 1.6 bar bar 
» 0.7 bar bar 


Safety clamp 

(no free flow) 

Pressure check > 0.4bar 
no free flow 


After start-up according 
to the instructions for 
use: 

Delivery accuracy + 5% 





Observe the procedural instructions (see „Procedural Instruc- 


tions for Inspection" x» p. 8 - 1)! 


Checks after Repair 


For your notes: 


Infusomat® fm 2.2 gb 


Infusomat? fm 2.2 gb 


Maintenance 6 


It is recommended every 2 years. In addition to the technical safe- 


ty inspection, the following assemblies/components are to be 


checked: 


LÊ 
2. 


Check rubber feet and if necessary exchange. 


Check smooth running of the pump cover, lock mechanism 
and door. 


Check smooth running of the flow inhibitor, clean and if nec- 
essary exchange pressure springs. 


Check seal membrane and if necessary exchange. 


Check the drop sensor optics and spring mechanism and 
clean, if necessary. 


Open unit. Internal visual inspection. Clean seal surfaces and 
If necessary exchange sealing strip. 


Check mechanical pressure control and if necessary calibrate. 


Check electronic pressure control and if necessary calibrate 
(see „Pressure Sensor" = p. 4 - 18). 


9. Assemble and seal unit ready for operation. 


D Maintenance 


For your notes: 


6- 2 Infusomat? fm 2.2 gb 


Technical Safety Check TSC 


Index c 


(Master - to be added to the documentation) 


Checklist for Technical Safety Check — Every 24 Months 


Unit: Infusomat? fm 








Manufacturer: B. Braun Melsungen AG 


Observe the Service Manual and the instructions for use. All measured values are to be documented. 
Accessories used should be included in testing. Make exclusive use of calibrated measuring instru- 
ments. 


Unit clean, complete, un- |O Mains voltage V AC Switch on unit: Drop sensor: 


damaged ÛJ Protective conductor resist- |O  Self-test O Simulate occlusion alarm 


Pump sealing diaphragm 
Membrane keyboard, rubber 
feet 

Operating unit, lock mecha- 
nism, pump cover 

Safety clamp (flow inhibi- 
tor) 

Mains cable and mains plug 
connector 

Staff call lead and plug con- 
nector 

Interface lead and plug con- 
nector 

Drop sensor line and plug 
connector 

Check voltage values 

100 -120V T 0.315A 

200 -240V T 0.16A 


ance O Control lamps 
incl. mains 


cable < 0.20 


Compare with display: 
— QO Set delivery rate g 
O Insulation resistance Û Set volume 
>> 2 MQ — QO Set time 
Earth leakage current g 
30 uA uA | Battery test: 
O Patient leakage current O Switch mains/battery/mains 
<5 uA — HA|O Switch on in battery mode 
and check self-test 


3. Accessories Used J 


Staff call lead Air inline sensor: 


Û 0.1 ml air bubbles no alarm 
0.4 ml air bubbles alarm 
Air value max. 91mV 
Water value min. 494mV 
Check threshold value 2 182 
mV and adjust if necessary 





(alarm with closed roller 
clamp) 
Simulate free flow (alarm) 


Electronic pressure control: 


Check alarm with switch- 
off pressure 
low / medium / high 


Mechanical pressure control: 


bar 


bar 


max. 1.6 bar 

» 0.7 bar 

Staff call 

Alarm suppression 
Safety clamp 

(no free flow) 
Pressure check > 0.4bar no 
free flow 

After start-up according to 
the instructions for use: 
Delivery accuracy + 5% 


M625 00 00 85 F04 C 3890 9979 (GB) 
Infusomat? fm 2.2 gb 





Technical Safety Check TSC 


For your notes: 


Infusomat® fm 2.2 gb 


Procedural Instructions for Inspection 8 


1. Visual Inspection 


2. Electrical Safety 


Infusomat® fm 2.2 gb 


Operating unit, lock mechanism, pump cover, seal membrane, 
flow inhibitor. Door lock: easy opening and closing, correct top 
and bottom locking. 


Pump cover must automatically open when the unit door is 
opened. 


Protective conductor resistance 

Protective conductor resistance « 0.2 Q incl. mains lead. 

Measuring points: 

- Potential equalization bolt 

- Bolt for door lock 

- Unit housing: 

a) If the unit is not sealed countersunk screw at the rear of 

the unit. 


b) Ifthe unit is sealed, remove lacquer from one of the holes 
in the foot stands. 


Note 


Do not use the foot stand mounting screws as alternative meas- 
uring points. 


Document largest value. 


Insulation Resistance 
Insulation resistance >> 2 MQ 


Measurement with 500 V between shorted mains connectors and 
potential equalization bolt. 


Earth Leakage Current 
Earth leakage current < 30 pA incl. mains cable. 


Measurement under standard conditions at the protective con- 
ductor of the mains cable. Two measurements (one with reversed 
polarity). Document largest value. 


Patient Leakage Current 
Patient leakage current < 5 A 


Connect cannula to the original Infusomat line (medium NaCl so- 
lution). Vent the set and run for approx. 1 minute. Measure can- 
nula with test equipment e.g. NSP 3000/4000. 


Procedural Instructions for Inspection 


4. Functional Inspection 


WARNING 


NEVER OPERATE THE UNIT WITH PLUGGED IN SERVICE CONNEC- 
TOR ON THE PATIENT. WHEN THE UNIT WAS OPERATED WITH 
PLUGGED IN SERVICE CONNECTOR IT IS TO BE SWITCHED OFF 
ONCE BEFORE ANY FURTHER USE. 


Switch-on Test 


Switch-on test keyboard and display: check correct course. Alarm, 
LED display 000, LCD display VOL TIME etc. display of all pixels, 
brightness, contrast. 


Battery Test 


Switch mains/battery/mains: 

Interrupt mains supply twice in intervals of 1 second. Pay atten- 
tion to the switch-over on the LED display. The unit must not 
switch to malfunction. 


Note 


Running time minimum 30 minutes after charging of 16 hours. 


Air Inline Sensor 
Set rate to 400 ml/h. 


-  |nject 0.1 ml air bubble with a 1 ml syringe. An alarm must 
not be activated. 


- [hen inject 0.4 ml air bubble. An alarm must be activated. 


- Check air inline sensor values in the Service Program "Air In- 
line Sensor Function 300.0", (see ,,Air Inline Sensor Function 
300" = p. 3 - 6) and ,… 520.0" (see „Air Inline Sensor 
Function 510* or 520**" c» p. 3 - 11). Values see TSC-list 
(see „Technical Safety Check TSC" œ p. 7 - 1). 


Drop Sensor 
Set rate to 400 ml/h. 


-  Simulate occlusion: Clamp infusion line in front of pump to 
prevent any drops. Alarm after less than 5 sec. 


-  Simulate free flow: Press bottom part of the drop chamber 
together, to generate a Jet. Immediate alarm. 


Staff Call 


Operation: Pin 3 and 5 connected. Pin 1 and 3 open. 
Alarm: Pin 1 and 3 connected. Pin 3 and 5 open. 


Note 


Reverse connection with IFMA software. 


Infusomat® fm 2.2 gb 


Infusomat? fm 2.2 gb 





Procedural Instructions for Inspection 


Alarm Suppression 
Press the alarm key. The current alarm is suppressed for 2 Minutes. 
General Pressure Measurement 


An electronic manometer should be used for the measurement 
described hereafter. 


If a mechanic manometer is used instead values which are appr. 
100 mbar lower are to be expected. 


Connect manometer to the outlet side of the pump and position 
the manometer approx. to medium height of the Infusomat® fm. 
Proceed as follows: 


- Vent manometer (filled with water) 6100 mm or 160 mm. 


- Measuring range: 2.5 or 4 bar. 


Note 


The results differ according to different measuring methods. 


Use infusion line maximum 50 times for pressure measurement. 


Electronic Pressure Control (Pressure Sensor) 
Set rate to 50 ml/h. Pressure sensor threshold low/medium/high. 


Deliver in an open system via the 3 way stop-cock. Turn stop-cock 
to manometer and build-up pressure. 


Document the alarm threshold. 
Pressure sensor threshold low 250 to 650 mbar 
Pressure sensor threshold medium550 to 950 mbar 


Pressure sensor threshold high 900 to 1300 mbar 


Mechanical Pressure Control 
- Set pressure sensor threshold to mechanical (test plug). 


- Switch off drop control. Build up pressure with 400 ml/h. 
Then measure with 100 ml/h. 


Measurement upper 


pressure value 1.35 to 1.76 bar 
(mechanical measurement, max. 
1.6 bar) 

Measurement lower 

pressure value » 0.95 bar 


(mechanical measurement » 0.7 
bar) 


Procedural Instructions for Inspection 


Mechanical Pressure Calibration 


Check pump pressure. If the deviation is max. + 0.3 bar from set 
range, the pump can be recalibrated. 


Pmax 1-35 to 1.76 bar; Pmin > 0.95 bar. 


Carry out pressure measurement. Calibrate pressure range at the 
set screw with an Allen key 2.5 mm. - After the pressure check the 
mechanical setting is to be switched off! 


After the service connector has been disconnected the electronic 
pressure control is not automatically activated again. 


Flow Inhibitor 


Switch to stop at high pressures. Then open unit door. The pres- 
sure must stay above 0.4 bar. 


Infusomat® fm 2.2 gb 


Procedural Instructions for Inspection 


Delivery Accuracy 

Measuring equipment: 

- 500 ml glass bottle, vented, drop chamber filled 2/3. 
- Graduated cylinder 25 ml, accuracy + 0.4 ml 

- Infusion solution NaCl or distilled water . 


- Check system: no narrow sections or kinks. Delivery rate 250 
ml/h. Run-in time minimum 1 minute. 


Delivery rate determination: 
- Perform measurement at ambient room temperature. 


-  |nsert outlet cannula in graduated cylinder. Simultaneously 
start stop watch and Infusomat fm. Stop when the 25 ml 
mark is reached. A deviation of 3.6 ml is equivalent to approx. 
1 % in the relevant range. Example: 


Set rate: ananunua ahead tie Ee iss EUER 250 ml/h 


0/0 ml/h 


Reference height 
80 + 2 cm 


25 ml + 0.4 ml 


5 min 


5 min 


5 min 





ANN 
TE 
Ju 
TU 
e TD 
^ 
m 





Table 8 - 1 


The inspection can be performed with a standard infusion line. 


A calibration line must be used for calibration! (see „Test Equip- 
ment and Special Tools" x» p. 9 - 1) 


Infusomat? fm 2.2 gb 8- 5 


Procedural Instructions for Inspection 


Alternative Measuring Methods 


Inspection of the delivery rate with a weight measurement. Avoid 
errors due to evaporation! 


Measuring equipment: 


- Scales 
Accuracy 0.1 g: 12 min 
Accuracy 0.01 g: 6-12 min 


Delivery rate determination: 


- Set the delivery rate to 200 ml/h and run unit 2 60 sec. (run- 
in time). 


- Insert the outlet cannula in container and simultaneously 
start stop watch and Infusomat fm. 


- After the time has expired stop Infusomat? fm and stop 
watch. 


- Immediately determine the delivery rate. 


Infusomat® fm 2.2 gb 


Test Equipment and Special Tools ` 9 


Test Equipment and Special Tools 


Infusomat? fm 2.2 gb 


For Repair / for Technical Safety Check (TSC) 


Order No. 
Test equipment case Infusomat fm (complete) ..... 0770 1527 
with: 
Calibration gauge 4 mm (for adjustment after exchange 
of the pump COVER) cui: nac esL KK KK KK ee Ae 0770 1489 
Pin punch 1.8 mm x 160 mm (for hinge pin/ 
disassembly of the pump cover) .............. 0770 1446 
Pin punch 6 mm x 125 mm (for hinge pin/ 
assembly of the pump cover)................. 0770 1454 
Flat tool 100 x 20 mm (for assembly/ 
disassembly of the tube adapter) ............. 0770 1462 
Special socket spanner M18 (for disassembly of 
the recessed plug). à «2s scere eec bn divî 0770 1497 
Manometer 0 - 4 Darcs dios RR KK Ris 0770 1357 
Service connector (red)....................... 0770 0709 
MFC service connector.. u...un anann annann 3450 1215 
AE ek de NPE anien a een 0770 1500 
Pressure calibration device zu dre KK RR KA 0770 5018 
9-1 


Test Equipment and Special Tools 


For your notes: 


Infusomat? fm 2.2 gb 


Unit Elements 


Designation Ord. No. 
Fuse [0.16 A 
for 200-240 V (10 pcs.) ....................… 3477 2847 
Fuse T 0.315 A for 100-120 V ..........0 00 eee eee 3477 0534 
(10 pieces) 
FUSE le c ou ue vacan oe dates diat coo ys he db e 3450 5652 
Battery incl. connector plug, 1.8 Ah / 7.2 V. ........ 3450 6357 
Cover for battery compartment .................. 3450 5504 
Door lock complete with push button.............. 3450 5601 
Spring holder for door lock. ....aannnannnannnnnnn 3450 5440 
Fixture Tor door-OCk «45 siye naya ee Ona bai kê RU ebê 3477 2790 
Housing labelling 
Gef dil ê» sdb d oed bee Bates ê a 3450 5610 
ICICI decade a d aan obsit dar erdt Ee oth ae 3450 5946 
DÛLÊ ETE DTE ea alt 3450 5954 
Ladies êê sa dka ma Mule d Lenin Cea SÛKÊ PA 3450 5962 
ENONSE- Sieb stood a bg des Pd oro x Shiye deh Rep 3450 5970 
SPANISH: tenancies PE 3450 5989 
DANS se hater beso a eet sent HHH HH tar ados 3450 5997 
NOMVEDIGN asocio ae ate ye doe dte sawn ee 3450 6101 
SWEGISN cute dara Seduta mem b breeds wed Ed aes 3450 6110 
FANSI uui aos satu pr acit th idus tectus a aa 3450 6128 
POMUUG UCSC. 2 “cawê dnt td kê êê eds p ded dad 3450 6136 
El BEE 3450 6144 
PONS: 3 seyrê keten bee dni sites aoro Durex nae dide k2 3450 6152 
Castellanos Asê n biren he anû we PRAE 3450 6160 
TUKISI, geneet ele A Dwênê ee Saee ebd 3450 6179 
Foot stand complete with rubber feet............. 3450 5415 
PIANC: vais ea A Ey En Dayê EV VR UM MN 3450 5512 
Ûr WE DOS) A nana we kinê ra turo eer Ue Et 3477 1530 
Analog board new part... kk kk kK RS 3450 580B 
Analog board exchange. s «amen po kk RR KV en 34 88 047B 
Fuse T 1.6 A on analog board...................- 34 77 3312 
(10 pieces) 
Rear panel without accessories .................… 3450 5628 
Cover for optical interface ...................... 3477 3164 
Strip seal for rear panel. kK RR RRA ee 3477 3142 
Drop sensor socket incl. cable and plug............ 3450 5687 


Infusomat? fm 2.2gbNot released 


Spare Parts List | 1() 


Designation Ord. No. 
DIANET recessed plug (8 pin) .................... 3450 5679 
Staff call recessed plug 

(SAD): sca gerne renta dicht acra ee 3450 5660 
Rear panel board 

with 5 and 8 pin recessed plugs 

With 5V TIL In têr ACE a Leona a kulb dalal ada 3450 5636 

With RS232 EE ides Kee dei eo wê 3450 563A 
fm recessed plug (3 pin) ........................ 3477 3177 
Potential equalization bolt .....................… 3477 0550 
Recessed mains plug without fuse holder.......... 3450 5644 
USE eror. gu 3450 5652 
Mains transformer 200-240 V Ju kk > 3450 5709 
Mains transformer 1O0-2120M 3450 5695 
Pump cover with lock cusan erg ae ims etat a 3450 5717 
Blind plug 7.1 mm (10 pcs.) ..................... 3477 3207 
Torsion spring in lever / pump cover (5 pes.) ....... 3477 3363 
Torsion spring for pump cover (5 pcs.) ............ 3477 3355 
Lever (DUMP COVER) ux sas sk ae o aee forct 3477 4092 
Hinge pin for pump Cover «scuro KK RR KK KK os 3477 3967 
Hinge pin for pump cover lever..................- 3450 5725 
Reed sensor set 

Sensor incl. 8 pin socket plug ................ 3450 5814 

Reed sensor incl. crimp terminals............. 3450 5431 

Socket plug (AMP/100 pin) .................. 3451 3744 


Front frame without flow inhibitor and pressure spring ... 3450 
5822 


Circular seal behind front frame ................. 3477 3126 
Flow inhibitor with pressure spring ............... 3477 3258 
Pressure spring for flow inhibitor (b pcs.) .......... 3477 3266 
Tube adapter with magnet ................2.0-0-- 3450 5830 
Magnet for tube adapter J.L kk KK KK K SJ 3450 5849 


Finger pump (without motor) including pump, 
pump cover, seal membrane and 


DOAN CS LE EE 3450 5407 

Finger pump (without motor) including pump, 
pump cover, and seal membrane.............. 3450 5407 
Membrane for pump unit occi uec! Gro RR RR K~ 3450 5733 
Motor With PINON sê weê ky id etie bn ors 3450 5741 
10- 1 


Designation Ord. No. 
Pressure sensor (cmpl.) .......u KK SU 3450 7345 
Air inline sensor with crimp terminals ............ 3450 5750 
Buzzer inek Sêdî ING EEN 3450 5776 
Air insulation board ....... 0... c KK eee ee eens 3450 6187 
Frame unit with seal plate hh... k> 3450 5466 
Door frame incl. pressure spring and magnet without 

sped. TF" —M 3450 5598 
Seal for door frame / flexible cable (5 pcs.) ........ 3477 3347 
MAGE UNIT ., ss a lsa wa nêna e nanna a nena dınê anê ERE 3450 5571 
Hinge pin for hinge unit ........... RR RR eee 3450 5580 
Membrane keyboard with support plate, seal and 

DUNG DIG sosse inae eami o mE ım 3450 5547 
Display board, raw material No.: 3810 6205 

Display board new part (Dupont blue) ......... 3450 555A 

Display board exchange (Dupont blue) ........ 3488 046A 
Display board, raw material No.: 3810 6906 

Display board new part (AMP black) .......... 3450 555B 

Display board exchange (AMP black) .......... 3488 046B 
Clamp for display board (20 pes.) ................ 3450 0448 
FED MOUS: ıu n mûn ê ewan Saa ka hewe bebê wênê 3450 5563 
Software Update 
Update kit IFMBO00002 or IFMBO1002 ............. 3450 6306 
Interface [INE cau wate oboe. ûk û seed lak ra m ws 0871 1658 

(see „Computer Interface" > p. 1 - 9) 
Colours 
Touch-up pen RAL 9002 (white).................. 3450 6977 
Touch-up pen RAL 7031 (grey) ................... 3450 6985 
10- 2 


Spare Parts List 


Designation Ord. No. 
Miscellaneous 

Mounting screw PT 2.5x14 (10 pcs.) .............. 3477 3100 
Mounting screw PT 2.2x8 (10 pcs.) ............... 3477 3118 
Screw PT 3x8 (20 DOS.) nic aac rici de i ei m 3477 3185 
Screw M 4X30 uiuo suc kk kk kK KK KK KK KK KK E Standard part 
Sew M Do Rad Padre Standard part 
Screw M 4x8 cesses kk kk kK KK KK KK nre CN Standard part 
Screw M 6X45 Ee kk kk kk kK KK KK KK KIR ee Standard part 
SSEW M E xua sualê êb nê kk a û bê a kl a alê kla Standard part 
Screw M 3x16 ...... cece kk kk KK KK KK KK KK Kı Standard part 
SOW E EE Standard part 
O-ring 6.07x1.78 Ak k kk kk kk kk KK RR KK een 3477 1530 
Distance sleeve (10 pesl,, KK A 3477 3231 
Plain washer 3.2 Düpcsl,,, 3477 3193 
Split washer B 3:1 aua ua nic omia KK KK KK e bee Standard part 
cca P" E eat ne eae s Standard part 
Washer 6.4 MM" Standard part 
Serrated lock washer | 6.4....................… Standard part 
Serrated lock washer | 3.2................221.… Standard part 
Serrated lock washer V 4.3 JW. kk Standard part 
Rubber feet (20 pc), 3477 3096 
Tamper-proof cap 10mm (50 pcs.)................ 3477 3134 
Label for fuse (5 pcs.) gege dea een 3477 3150 
Clamp for display board (20 pcs.) ................ 3450 0448 
Shrink tUDE:SIZe 48, ea su i e Rc xn RC RC Standard part 
Shrink tube size EE Standard part 


Not released Infusomat? fm 2.2gb 


Spare Parts List 





M4x30 and 


Distance —— 







7¬ PT 2.2x8 


PT 2.5x14 


INF006/HNDesign/ 21.05.1997 6 pes $— — — M4x8°® 
1 in addition split washer B 3.1 6 in addition serrated lock washer | 4.3 
? in addition washer 4.3 and serrated lock washer | 4.3 7 in addition serrated lock washer | 3.2 
3 in addition washer 6.4 and serrated lock washer I 6.4 8 in addition washer 3.2/9 
^ in addition serrated lock washer I 3.2 S=Countersunk screw 


? in addition serrated lock washer V 4.3 and tamper-proof cap 10 mm 


Fig.: 10-1 Exploded drawing 


Infusomat? fm 2.2gbNot released 10- 3 


Accessories 
Designation Ord. No. 
Pole Clamp 
Pole clamp (universal clamp, rotating) ............ 3450 9054 


Universal Clamp (Pole Clamp) 


Fig.: 10-2 Universal clamp (Pole clamp) 


Designation Ord. No. 
Universal clamp, complete `... 3450 5857 
UniversalcldliD: š >5 na cıa SA en Ea Ra ewes mis 3450 8325 
Threaded TO ~e any êna e ipu na wênê e Kane wt Ebd lêê 3450 8333 
Star. handle DO. xos ak ed ke oxi d aco scat se Rus 3450 8384 
GENEE asm aho Send M Ext Rod pea ova rg 3450 8341 
Salt ety NOOK tiet den cee hae hone NOD ES 3450 8368 
Plate PEST A ys seraya darê ê aê êyê Qale bebes 3450 2610 
Connection cap D12/4 (5 pcs.).................…. 3477 4149 
Bellows (5PCS) ex sc ka oe leas be AWA AO eta etes 3477 3274 
Pressure spring Db Des.) duse ie dvi wada inka see 3477 4165 
10 - 4 





Spare Parts List 


Designation Ord. No. 


Universal Clamp 





Fig.: 10-3 Universal clamp 


Designation Ord. No. 
Universal clamp, complete............. not available any more 
WIPE ACEO OOS La actus gode ots ns tne orn aber e ris 34 50 5903 
Safety NOOK -srini roana deste b br do pes 34 50 5865 
(fe leger rte 3 ud aded des eoi Seb etes 34 50 5890 
Rubber COVEN (5:DCS J)» asa yan he edo erepta 34 77 3290 
BellOwS:(5 OCS.) xau EXER te EORR LG tai 34 77 3274 
Connection eap (5 0S) e adt edle tea et ca 34 77 3304 
Pressure spring for pole fixation (5 pcs.) ........... 34 77 3282 


Not released Infusomat? fm 2.29b 


Appendix 


Revision Service Documentation 


Current Information 


Infusomat® fm 2.2 gb 


Å 


Version 2.0 


This manual has been completely revised. The most important 
changes are listed below: 


- Changed manual structure 

- New air value 

- New TSC list 

- New paragraph pressure sensor (exchange / calibration) 
- New test equipment for pressure sensor 

- New spare parts 


- Total list of spare parts 


Version 2.1 


- Changed spare parts numbers 


Version 2.2 

This version was approved by B.Braun on 26.09.2006. 

The most important changes of this version are listed below: 
- Changed pressure sensor values 


- Changes in texts and layout 


1. Noise Reduction 


From unit number 31067 the running noise of the motor was re- 
duced by optimizing the motor drive. This modification is standard 
for analog boards from index D on. 


Retrofitting for old boards: 


Designation Ord. No. 
Retrofit kit for noise reduction...................- 3450 0952 
Note 


See assembly instructions for retrofitting. 


Appendix 


2. EMC: Increase of Interference (e.g. against mobile phones) 
Exchange drop sensor against shielded version. 

Identification of a shielded drop sensor: 

The sensor is either marked with a casting stamp or an index Å. 


Check: Ground contact of the drop sensor socket, if necessary ret- 
rofit (see „Drop Sensor Socket" r> p. 4 - 10). 


If the drop sensor is not shielded: 


Retrofit interference clips 


Designation Ord. No. 
NEE Kili ıs a iur Re der aê lê S kla Aya wê Anê dd hı de xwê a 3450 6969 
Note 


See assembly instructions for retrofitting. 


3. Modification of the Interface from 5V TTL to RS232 


- Connect pin 3B of connector P5 on the soldering side to +5V 























Ausschnitt i i 
Mikroprozessorplatine (IR receiver V4, pin 3). 
= IEEE - Document the modification of the board by the following in- 

Wl Lötseite des 
V6 | 8] 
vs [8 ótbrücke zwischen MP old without index r> new index D, 
va J FEYDE EIUS: MP old index A => new index E, 
vE sæ es END MP old index B = new index F. 








Mark the RS232 interface on the unit rear panel with an 
RS232 sticker. 


Infusomat® fm 2.2 gb