Philips 861290 HeartStart XL+ defibrillator monitor
sense simplicity
For resuscitation teams and rapid response teams, the Philips HeartStart XL+ defibrillator/monitor is
designed with meaningful innovations that help you confidently and effectively respond to patients’
needs throughout the hospital environment.
Hospital cardiac emergencies are often stressful and hectic
events. You want a defibrillator that is dependable and
ready to “grab and go” at a moment’s notice. Building on
the clinical history of Philips defibrillation, the HeartStart
XL+ will inspire confidence in its ability to respond in
critical situations.
The HeartStart XL+ offers advanced and basic life-support
in a compact, light-weight and easy-to-use defibrillator/
monitor. It offers both Manual and AED defibrillation,
synchronized cardioversion, and arrhythmia detection
with ECG monitoring through 3- and 5-Lead sets or
multifunction electrode pads. Options include SpO, and
Non-invasive blood pressure (NBP) monitoring and
Pacing.
The HeartStart XL+ introduces the Patient Category
button. With one simple press of the Patient Category
button you can switch from Adult to Infant/Child patient
category. The device not only quickly and automatically
adjusts the energy for AED defibrillation but it also
automatically adjusts all alarm limits.
AED Mode uses the same SMART Analysis algorithm as
Philips’ industry-leading AEDs. In AED Mode, with the use
of the patient category button, the HeartStart XL+
simplifies the ability to defibrillate different patient types
without the need to calculate a dose or special pads to
reduce energy. AED Mode coaches you with easy to
understand voice prompts and display messages.
In Manual Defibrillation Mode, defibrillation is as easy as 1,
2, 3. The entire process is under your control. After
assessing the ECG you decide if defibrillation,
cardioversion or pacing is necessary, select an energy,
charge and shock. Select cardioversion by simply pressing
the Sync button. When Sync Mode is active, the Sync
button lights up and indications appear on the display.
In optional Pacing Mode, you can perform either Demand
or Fixed mode transcutaneous pacing with a configurable
choice of 20 or 40 msec pulse duration. The device also
allows you to use front panel controls to adjust rate and
output.
The HeartStart XL+ is highly configurable such that you
can configure the energy, number of shocks and CPR time
to meet your hospital defibrillation protocol needs.
Viewing and obtaining data from the HeartStart KLF is
easy. You can view patient parameter Trend Reports on
the display. The Trend Reports and Event Summaries are
also available after an event and are easily retrieved. The
results of all daily, weekly and monthly self-tests and
Operational Checks are stored in internal memory and
available for review.
Data can be copied to a USB Drive for easy transfer to
event review data management software.
Since it is part of a larger Philips Healthcare family of
products, the HeartStart XL+ can help you simplify
inventory management and reduce inventory costs. The
device’s supplies and accessories are compatible with
other Philips defibrillators while the patient monitoring
cables and sensors are the same ones used with Philips
patient monitors.
The HeartStart XL+ is ready. Ready to respond. Ready to
revive. Ready to review. And ready to meet the needs of
your organization.
Features/Upgrades
Standard Features
ST/AR Basic algorithm for arrhythmia detection and
ECG analysis
SMART Analysis algorithm for making shock decisions
in AED Mode
One button push to change from Adult to Infant/Child
patient category, including alarm limits in all modes
ECG monitoring through monitoring electrodes and
defibrillation pads
“Ouick Look” ECG monitoring through erternal
paddles
Synchronized cardioversion with dedicated back-lit
button
Adjustable ECG size and autogain
Manual and AED operation
3-wave color display
Strip chart printer (50 mm)
Individual, adjustable volume of ORS beeper, voice
prompts and alerts
Event Summary
Displayed and printed Trending Reports
Configuration Mode
Service Mode
Operational checks
Automated self-tests with “Ready For Use” indicator
Lithium lon battery with fuel gauge
Optional Features
SpO2 with Fourier Artifact Suppression Technology
(FAST)
Noninvasive Blood Pressure
Noninvasive Pacing
Standard Accessories
Lithium lon Battery with fuel gauge
Hands-free multifunction electrode cable
3-Lead ECG trunk cable and lead set
10 monitoring electrodes
50mm printer paper
Defibrillator Test Plug
Three cable wraps (ECG, pads cable, external paddles)
Documentation CD containing instructions for use,
user training material and application notes
Quick Reference Cards
Product Ordering Information
861290 HeartStart XL+ Defibrillator/Monitor
Option Ordering Information
A01 SpO,Monitoring
A02 SpO, and NBP Monitoring
B01 External Pacing
C01 External Paddles with PCI
C02 External Paddles Sterilizable
C03 USB Data Drive
C04 Accessory Storage System
C05 Extra Lithium lon Battery
C06 Bed Rail/Roll Stand Mounting
C07 ECG Sync Out Cable
C08 5-Lead ECG Cable
LP1 Instructions for Use
LP3 User Training DVD
SM1 Service Manual
wo01 One-year on-site warranty
W08 Three-year bench repair and loaner
warranty
W23 Two-year bench repair and no loaner
warranty
W24 Five-year bench repair and loaner
warranty
WAI Five-year Biomed Alliance
861395 SpO,
861396 NBP
861397 External Pacing
861401 Pads Conversion
861402 External Paddles
861403 Hardware Platform
861404 Software Platform
Supplies
Multifunction Electrode Pads
M3501A
M3504A
M3713A
M3716A
M3717A
M3718A
M3719A
989803166021
989803158211
989803158221
989803139261
Hands-Free Pads Therapy Cables
M3507A
M3508A
05-10200
External Paddles
M3543A
M4745A
M4759A
Internal Switched Paddles
M4741A
M4742A
M4743A
M4744A
Internal Switchless Paddles
M1741A
M1742A
M1743A
M1744A
A
Adult/Child Pads, barrel connector,
AAMI
Infant Pads, barrel connector, AAMI
HeartStart Adult/Child Plus Pads
HeartStart Adult/Child Radiolucent Pads
HeartStart Infant Plus Pads
HeartStart Adult/Child
Radiotransparent/Reduced Skin Irritation
Pads
HeartStart Infant Radiotransparent/
Reduced Skin Irritation Pads
HeartStart Adult/Child Preconnect Pads
HeartStart Adult (1 set)
HeartStart Adult (5 sets)
SMART Pads II Adult/Child/Infant
(for handoff only)
Defibrillator Pads Hands-Free Cable,
barrel style (2.2 m/7 ft.)
Defibrillator Pads Hands-Free Cable,
plug style (2.2 m/7 ft.)
Pads Adapter (use with M3507A)
External Paddles
Sterilizable External Paddles
Paddle Electrode Replacement
7.5cm Switched Internal Paddles
6.0cm Switched Internal Paddles
4.5cm Switched Internal Paddles
2.8cm Switched Internal Paddles
7.5cm Switchless Internal Paddles
6.0cm Switchless Internal Paddles
4.5cm Switchless Internal Paddles
2.8cm Switchless Internal Paddles
3-Lead ECG Cables
M1500A
M1510A
M1605A
M1615A
M1669A
M1671A
M1672A
M1673A
M1674A
M1675A
M1678A
M3526A
M3528A
3-lead ECG Trunk Cable (AAMI)
3-lead ECG Trunk Cable (IEC)
3-lead ECG Set with Snaps (AAMI)
3-lead ECG Set with Snaps (IEC)
3-lead ECG Trunk Cable (AAMI/IEC)
3-lead ECG set grabber (AAMI/ICU)
3-lead ECG set ICU grabber (IEC)
3-lead ECG set with snaps (ICU)
3-lead ECG set with snaps (IEC, ICU)
3-lead ECG set with grabbers (OR)
3-lead ECG set, grabber (IEC, OR)
3-wire lead set and plug, snap (AAMI)
3-wire lead set and plug, snap (IEC)
5-Lead ECG Cables
M1520A
M1530A
M1625A
M1635A
M1644A
M1645A
M1668A
M1949A
M1968A
M1971A
M1973A
M1974A
5-lead ECG Trunk Cable (AAMI)
5-lead ECG Trunk Cable (IEC)
5-lead ECG set with snaps (AAMI)
5-lead ECG set with snaps (IEC)
5-lead ICU snaps (AAMI)
5-lead ICU snaps (IEC)
5-lead ECG Trunk Cable (AAMI/IEC)
5 plus 5 ECG Trunk Cable (AAMI/IEC)
5-lead ICU grabber (AAMI)
5-lead ICU grabber (IEC)
5-lead OR grabber (AAMI)
5-lead OR grabber (IEC)
6-Lead ECG Cables
M1667A
M1680A
M1682A
M1684A
6-lead Trunk Cable (AAMI)
6-lead ICU grabber, limb (AAMI)
6-lead ICU snap, limb (AAMI)
6-lead OR grabber, limb (AAMI)
10-Lead ECG Cables
M1663A
M1665A
M1682A
M3525A
10-lead Trunk Cable (AAMI)
10-lead Trunk Cable (AAMI)
6-lead ICU snap, limb (AAMI)
10-lead ECG patient Trunk Cable, 12-pin
ECG Monitoring Electrodes
M2202A
High-Tack Foam ECG Electrodes — 5
electrodes/pack (60 packs/case)
ECG Sync Out Cables
M1783A
M5526A
12-pin Sync Cable (8 feet)
12-pin Sync Cable (25 feet)
NBP Interconnect Tubing
M1598B
M1599B
Adult Pressure Interconnect Cable
(1.5m)
Adult Pressure Interconnect Cable (3m)
Reusable Blood Pressure Cuffs
40400A
40400B
40401A
40401B
40401C
40401D
40401E
M4552B
M4553B
M4554B
M4555B
M4556B
M4557B
M4558B
M4559B
M1571A
M1572A
M1573A
M1574A
M1575A
M1576A
Reusable Cuff Kit, 3 sizes (Pediatric,
Adult, Large Adult)
Reusable Cuff Kit, 5 sizes (Infant,
Pediatric, Adult, Large Adult, Thigh)
Traditional Reusable Cuff — Infant
Traditional Reusable Cuff — Pediatric
Traditional Reusable Cuff — Adult
Traditional Reusable Cuff — Large Adult
Traditional Reusable Cuff — Thigh
Easy Care Reusable Cuff — Infant
Easy Care Reusable Cuff — Pediatric
Easy Care Reusable Cuff — Small Adult
Easy Care Reusable Cuff — Adult
Easy Care Reusable Cuff — Adult Long
Easy Care Reusable Cuff — Large Adult
Easy Care Reusable Cuff — Large Adult
X-Long
Easy Care Reusable Cuff — Thigh
Multi-Patient Comfort Cuffs — Infant
Multi-Patient Comfort Cuffs — Pediatric
Multi-Patient Comfort Cuffs — Small
Adult
Multi-Patient Comfort Cuffs — Adult
Multi-Patient Comfort Cuffs — Large
Adult
Multi-Patient Comfort Cuffs — Thigh
Disposable Blood Pressure Cuffs
M1874A
M1875A
M1876A
M1877A
M1878A
M1879A
M4572B
M4573B
M4574B
M4575B
M4576B
M4577B
M4578B
M4579B
Disposable Cuff — Infant
Disposable Cuff — Pediatric
Disposable Cuff — Small Adult
Disposable Cuff — Adult
Disposable Cuff — Large Adult
Disposable Cuff — Thigh
Soft Single-Patient Disposable Cuff —
Infant
Soft Single-Patient Disposable Cuff —
Pediatric
Soft Single-Patient Disposable Cuff —
Small Adult
Soft Single-Patient Disposable Cuff —
Adult
Soft Single-Patient Disposable Cuff —
Adult X-Long
Soft Single-Patient Disposable Cuff —
Large Adult
Soft Single-Patient Disposable Cuff —
Large Adult X-Long
Soft Single-Patient Disposable Cuff —
Thigh
M1131A
M1132A
M1133A
M1191A
M1191B
M1191AL
M1191BL
M1191T
M1192A
M1192T
M1194A
M1195A
M1196A
M1196T
M1902B
M1903B
M1904B
M1941A
M1943A
M1943AL
Disposable SpO, Sensor — Adult/
Pediatric Finger
Disposable SpO, Sensor — Pediatric
Disposable SpO, Sensor — Neonatal/
Adult (for adults only - any finger except
thumb)
Reusable SpO, Sensor — Adult finger
Reusable SpO, Sensor — Adult finger
Reusable SpO, Sensor — Adult finger
(3m cable)
Reusable SpO, Sensor — Adult finger
(3m cable)
Reusable SpO, Sensor — Adult finger
(9-pin connector)
Reusable SpO, Sensor — Pediatric/Small
Adult
Reusable SpO, Sensor — Pediatric
Finger (9-pin connector)
Reusable SpO, Sensor — Adult/Pediatric
ear clip
Reusable SpO, Sensor — Infant
Reusable SpO, Sensor — Adult clip
Reusable SpO, Sensor — Adult clip
(9-pin connector)
Disposable SpO, Sensor — Infant digit
(available outside the U.S. only)
Disposable SpO, Sensor — Pediatric
Finger (available outside the U.S. only)
Disposable SpO, Sensor — Adult Finger
(available outside the U.S. only)
SpO, Extension Cable (2m)
Reusable SpO, Sensor Adapter Cable
(1m) — Use with M1903B/M1904B
Reusable SpO, Sensor Adapter Cable
989803167281 HeartStart XL+ Lithium lon Battery
Data Management
989803171261
USB Data Drive
Accessory Pouches
989803171281
989803171291
One therapy and one monitoring
accessory system
Three cable wraps (one each for ECG,
hands-free therapy cable, external
paddles)
Mounting Solutions
989803171701
PH-0050-60
PH-0050-03
Bed Rail/Roll Stand Hook
GCX Roll Stand with basket
(orderable through GCX)
GCX Flush Wall Mount
(orderable through GCX)
Test Loads and Test Plugs
M1781A
M3725A
989803171271
Adapters
Test Load for use with M3507A Pad
Cable
Test Load for use with M3508A Pad
Cable
Defibrillator Test Plug for M3508A
Cable
M4740A
05-10200
Paper
For use with internal switchless paddles
For M3507A. Changes barrel cable to
flat (removable or permanent) for
M37xxA pads
40457C
40457D
50 mm Chemical Thermal Paper, gray
grid (10 rolls)
50 mm Chemical Thermal Paper, gray
grid (50 rolls)
Specifications
Physical Dimensions
Size: inches: 11.6W x 9H x 10.9D
cm: 29.6W x 23H x 27.9D
Weight: 14.7 lbs/6.6 kg (includes one battery, one
new roll of paper, one pads cable).
Incremental weight of external standard
paddles and paddle tray is less than 3 lbs
(1.3 kg).
Standard Within one meter (3 feet) of the device.
Operator
Position:
Power Supply
Rechargeable Lithium lon battery; AC power using a
protectively grounded outlet.
Alarm Tone Maximum - 85 dB(A),
and Voice Minimum - 45 dB(A).
Message
Volume Range:
Alarm Tone Volumes
Imminent Continuous tone alternating between
Shutdown: 1000 and 2100 Hz.
High Tone of 960 Hz lasting 0.5 sec repeated
Priority: every second.
Medium Tone of 480 Hz lasting 1 sec repeated
Priority: every two seconds.
Low Tone of 960 Hz lasting 0.25 sec repeated
Priority: every two seconds.
Alarm Characteristics
High Flashing at 2 Hz with 50% duty cycle (a
Priority: .25-sec flash twice every second).
Medium Flashing at 0.5 Hz with 50% duty cycle (a
Priority: 1-sec flash every other second).
Low Constant on.
Priority:
Defibrillator
Waveform: Biphasic Truncated Exponential.
Waveform parameters adjusted as a
function of patient impedance.
Shock Via multifunction electrode pads or
Delivery: paddles.
Shock Series: | Configurable energy escalation in a
series.
Leads Off Apply 500nA rms (571Hz); 200uA rms
sensing and (32K Hz).
PCI sensing for
pads/paddles:
Charge time: 3 seconds to the recommended adult
energy level (150 Joules) with a new
fully-charged battery installed.
Less than 5 seconds to the selected
energy level (up to 200 Joules) with a
new fully charged battery installed.
Less than 15 seconds to the selected
energy level while connected to AC
power only.
The device powers on in manual
defibrillation mode ready to deliver
shock in less than 8 seconds plus
applicable charge time, assuming an
immediate selection of an energy and
initiation of a charge, even at 90V AC
and after 15 maximum energy
discharges.
The device powers on in AED mode
ready to deliver shock in less than 17
seconds plus applicable charge time.
Patient Minimum: 25 ohm (external
defibrillation); 15 ohm (internal
defibrillation); Maximum: 250 ohm.
Actual functional range may exceed
Impedance
Range:
these values.
Delivered Energy Accuracy
Nominal Delivered Energy
vs. Load Impedance
Energy
1)
2)
3J
4)
5)
6)
7)
8)
9)
10)
15)
20)
30)
50)
70)
100)
120)
150)
170)
200)
Load Impedance (ohms) + 2%
25
1.2
1.7
2.6
3.5
4.3
5.2
6.1
6.9
7.8
8.7
13
17
26
43
61
87
104
130
147
173
50
1.3
2.0
3.0
4.0
5.0
6.0
7.0
8.0
9.0
10
15
20
30
50
70
100
120
150
170
200
75
1.3
2.1
3.1
4.2
5.2
6.3
7.3
8.4
94
10
16
21
31
52
73
105
126
157
178
209
100
1.2
2.0
32
4.3
5.4
6.5
7.6
8.6
9.7
11
16
22
32
54
76
108
130
162
184
216
125
1.1
1.9
3.2
44
5.5
6.6
7.8
8.9
10
11
17
22
33
55
78
111
133
166
188
222
150
1.0
1.7
3.1
4.5
5.6
6.7
7.8
8.9
10
11
17
22
33
56
78
111
134
167
189
223
175
0.9
1.6
Deg)
43
5.4
6.5
7.6
8.7
9.8
11
16
22
33
54
76
108
130
163
184
217
Delivered Energy accuracy is +10% or +1] whichever is greater
for all energy settings.
Smart Biphasic Waveform at 200)
2500
2000
1500
into 25-175 Ohms
1000
500
Voltage (V)
-500
-1000
-1500
Manual Defibrillation
Manual Output
Energy
(Selected):
Controls:
Energy
Selection:
Charge
Control:
Shock
Control:
Synchronized
Control:
Synchronized
Shock Timing:
Indicators:
Armed
Indicators:
1-10, 15, 20, 30, 50, 70, 100, 120, 150,
170, 200 Joules; maximum energy limited
to 50) with internal paddles.
On/Off Therapy knob, Charge, Shock,
Sync, ECG Lead Select, Patient Selection,
Print, Mark Events, Reports, Alarms,
Menu Select, Navigation.
Front panel Therapy knob.
Front panel button; button on external
paddles.
Front panel button; buttons on external
or switched internal paddles.
Front panel Sync button.
Maximum time from R-Wave detected
to shock delivered is 25ms, as measured
with oscilloscope from peak of input
QRS wave to leading edge of
defibrillation discharge into a 50 ohm
test load.
Text prompts, audio alerts, QRS beeper,
battery status, Ready For Use (RFU),
External Power, Sync Mode.
Charging/charged tones, flashing shock
button on front of panel and on external
paddles, energy level indicated on the
display.
0 5 10 15 20
25 —50 —75 100 —125 150 —175
Time (ms)
AED Mode ECG and Arrhythmia Monitoring
AED Energy
Profile:
AED Conrrols:
Text and Voice
Prompts:
Indicators:
Armed
Indicators:
ECG analysis:
Shockable
Rhythms:
Shock
Advisory
Algorithm
Sensitivity:
Shock
Advisory
Algorithm
Specificity:
150 Joules for Adult/50 Joules for Infant/
Child (factory default) nominal into a 50
ohm test load.
On/Off, shock.
Extensive text/audible messages guide
user through a user-configured protocol.
Monitor display messages and prompts,
voice prompts, battery status, RFU,
external power.
Charging/charged tones, flashing shock
button, energy level indicated on the
display.
Evaluates patient ECG and signal quality
to determine if a shock is appropriate
and evaluates connection impedance for
proper defibrillation pad contact.
SMART Analysis is designed to shock
ventricular fibrillation, ventricular flutter
and polymorphic ventricular tachycardia.
It is designed to avoid delivering a shock
for rhythms that are commonly
accompanied by a pulse or rhythms that
would not benefit from an electrical
shock.
Meets AAMI DF39 requirements and
AHA recommendations; Adult:
Ventricular Fibrillation - 90% with lower
confidence limit (LCL) of 87 %,
Polymorphic Ventricular Tachycardia
and Ventricular Flutter - 75% with LCL
of 67 %; Infant/Child: Ventricular
Fibrillation - 90% with LCL of 87 %.
Meets AAMI DF39 requirements and
AHA recommendations; Normal Sinus
Rhythm - 99% with LCL of 97%; Asystole
- 95% with LCL of 92%; Other non-
shockable Rhythms - 95% with LCL of
88%.
Inputs:
Lead Fault:
Pad Fault:
Heart Rate
Display:
Heart Rate/
Arrhythmia
Alarms:
Common
Mode
Rejection:
ECG Size:
ECG
waveforms:
ECG Leads Off
Sensing:
Maximum
T-Wave
amplitude:
Frequency
Response:
Up to 3 ECG waves may be viewed on
the display and up to 2 waves printed
simultaneously. Lead I, Il or Ill is obtained
through the 3-wire ECG cable and
separate monitoring electrodes. With a
5-Lead ECG cable, leads aVR, aVL, aVF
and V can also be obtained. Pads ECG is
obtained through two multifunction
electrode pads.
Messages and dashed lines appear on the
display if an electrode or lead becomes
disconnected.
Dashed line appears on the display if a
pad becomes disconnected.
Digital readout on the display from 16 to
300 bpm (Adult Patient Category) or 16
to 350 bpm (Infant/Child), with an
accuracy of +10% or +5 bpm whichever
is greater.
HR high/low, Asystole, VFIB/V-TACH,
VTACH, Extreme Tachy, Extreme
Brady, PVC rate, Pacer Not Capture,
Pacer Not Pacing.
105 dB for Leads ECG, 96 dB for pads
ECG.
1/4x, 1/2x, 1x, 2x, 4x, auto gain (1x gain
is 10mm/mV on the printed strip).
Displayed at a fixed timebase of 25 mm/
sec (printer) +5%, 20 mm/sec (display)
+10%.
3- and 5-Lead wires apply a 30nA DC
current.
Device rejects up to 80% of R-Wave
amplitude for synchronized
cardioversion; up to 55% of R-Wave
amplitude for demand pacing; up to 34%
of R-Wave amplitude for arrhythmia
analysis.
ECG AC Line Filter - 50 Hz or 60 Hz;
ECG for Display - 0.15-40 Hz, 0.05-40
Hz; ECG for Printer - 0.05-150 Hz
Diagnostic, 0.15-40 Hz, 0.05-40 Hz.
ECG and Arrhythmia Monitoring ECG and Arrhythmia Monitoring
Heart rate
accuracy and
response to
irregular
rhythm:
Heart rate
averaging:
Pace Pulse
Detection
Sensitivity:
ECG Analog
Output
Bandwidth:
ECG Analog
Output Gain:
ECG Analog
Output Delay:
Pacemaker
Pulse
Rejection
Capability:
Heart Rate
Response
Time:
Time to Alarm
for
Tachycardia:
Meets AAMI standard for ventricular
bigeminy (HR = 80 bpm); slow
alternating ventricular bigeminy (HR=60
bpm); rapid alternating ventricular
bigeminy (HR=120 bpm); bidirectional
systoles (HR=90 bpm) as measured after
a 20 sec stabilization time.
For heart rates > 50 bpm, heart rate is
determined by averaging the 12 most
recent R-R intervals. Beats N, P, and V
are included. When heart rate drops
below 50 bpm, the four most recent R-R
intervals are used in the average. Note:
For ventricular tachycardia alarms, which
have a user-definable PVC run length
limit, the heart rate is based on the user-
selected PVC length up to 9 PVCs
maximum.
1 mV for a width of 100 ps; 200 uV for a
500 us width and 200 uV for widths of
500 us to 2 ms.
0.5 to 70 Hz
1v output per 1mV input +10%
Propagation delay time is <35ms from
ECG input to ECG analog output.
Amplitude from + 2 mV to + 700 mV,
width from 0.1 ms to 2.0 ms as per
ANSI/AAMI EC 13: 2002 4.1.4.1.
7 sec (80-120 bpm); 6 sec (80 to 40
bpm)
3 sec for 206 bpm (1 mV, halved
amplitude and double amplitude) and
195 bpm (2 mV, halved amplitude and
double amplitude) as measured following
a normal 80 bpm rate with upper alarm
limit set at 100 and lower alarm limit set
at 60 bpm.
Patient
Isolation
(Defibrillation
Proof):
Other
consideration:
Battery
Type:
Dimensions:
Weight:
Charge time
with device
turned off and
AC power
connected:
Charge time
with device on
and AC power
connected
Life:
Lead ECG, SpO2, NBP and Internal
Paddles: Type CF
Pads/Paddles: Type BF
The HeartStart KLF is suitable for use in
the presence of electrosurgery. Burn
hazard protection is provided via a 1K
current-limiting resistor contained in
each ECG lead wire. Proper lead
placement is important to reduce burn
hazards in the event of a defect in the
electrosurgical equipment.
Rechargeable, Lithium lon; See battery
label for capacity information.
23.6mm (H) x 116mm (W) x 146 mm
(L); 1 in (H) x 4.5 in (W) x 5.7 in (L)
Approximately .68kg (1.5 lbs)
With the temperature between 0-35” C
(32-95° F), less than 3 hours to 100%
capacity; less than 2 hours to 80%
capacity.
Charge time is less than 10 hours.
Depends on frequency and duration of
use. When properly cared for and used
in its intended environment, the useful
life is approximately 3 years. Use outside
those conditions can significantly reduce
battery life.
Capacity:
Battery
Indicators:
Battery
Storage:
With a new fully charged battery, at 20
°C (68 °F), one of the following: At least
3 hours of monitoring (ECG and SpO2
monitored continuously and NBP
sampled every 15 minutes) followed by
20 full-energy charge/shocks; OR at least
two hours of pacing (180ppm at 140mA
with 40 msec pulse width) while
monitoring (ECG and SpO2 monitored
continuously and NBP sampled every 15
minutes) followed by 20 full-energy
charge/shocks; OR at least 175 full-
energy charge/shocks.
Battery gauge on battery, capacity
indicator on display, power indicators on
front of device; flashing RFU indicator,
chirp and Low Battery messages on the
display for low battery condition. When
a low battery message first appears there
is still enough energy for at least 10
minutes of monitoring and 6 maximum
energy discharges.
Storing the battery for extended periods
at temperatures above 40° C (104° F)
reduces battery capacity and degrades
battery life.
Continuous
ECG Strip:
Auto Printing:
Reports:
Speed:
Amplitude
Accuracy:
Paper Size:
The Print key starts and stops the strip.
The printer can be configured to be run
real time or with a 10-second delay. The
strip prints the primary ECG lead and a
second wave with event annotations and
measurements.
The printer can be configured to
automatically print on Mark Events,
Charge, Shock and Alarm.
The following can be printed: Event
Summary (Long or Short), Vital Signs
Trends, Operational Check,
Configuration, Status Log, Device
Information.
25 mm/s with an accuracy of +5%.
5% for offset voltages of + 300 mV at
5Hz.
50 mm (W) x 30 m (L)
Pacing
Waveform:
Current Pulse
Amplitude:
Pulse
Duration:
Rate:
Mode:
Refractory
Period:
Monophasic.
10 mA to 200 mA if the pulse width is
set to 20ms (5 mA increments); accuracy
+10% or +5 mA whichever is greater.
For a 40ms setting, the maximum pacing
current is 140mA.
20 or 40 msec with +10% accuracy.
30 ppm to 180 ppm (10 ppm
increments); accuracy +1.5%
Demand or Fixed.
340 msec (30 to 80ppm); 240 msec (90
to 180 ppm) +10%.
Measurement
Range:
Resolution:
Update Period:
Ambient Light
Sensitivity:
Alarm Range:
SpO, and Pulse
High/Low
Alarm Signal
Generation
Delay:
Response
Time (90 to 80
%):
SpO, and Pulse
Averaging
Time:
Emitted Light
Energy:
Wavelength
Range:
0-100%
1%
1-2 sec typical; maximum of < 30 sec
Interference from fluorescent light is
<2% SpO2 under the following
conditions: 0.3 and 1% perfusion, 50 nA/
mA transmission, 10 to 1000 Ix light
intensity, 50/60Hz power line frequency
+0.5 Hz line frequency.
Low Limit: 50-99% (Adult and Infant/
Child); High Limit: 51-100% (Adult and
Infant/Child)
10 seconds
Average 18.9 seconds, standard
deviation 0.88 seconds
10 sec
<15 mW
500-1000 nm (Information about
wavelength range can be useful to
clinicians, especially those performing
photodynamic therapy.)
Desat Alarm
Signal
Generation
Delay:
Pulse Rate
Measurement
Range:
Pulse Rate
Resolution:
Pulse Rate
Accuracy:
Pulse
Response
Time (90 to
120 bpm):
Pulse Alarm
Range:
Accuracy with sensors (1 standard deviation 70-100%):
+ 2%:
+ 3%:
Systolic
Pressure
Range:
Diastolic
Pressure
Range:
Mean Pressure
Range:
Systolic Alarm
Range High:
Systolic Alarm
Range Low:
Diastolic
Alarm Range
High:
Diastolic
Alarm Range
Low:
20 sec
30-300 bpm
5 bpm
2% or 1 bpm whichever is greater
Average 18.0 seconds, standard
deviation 0.86 seconds.
Low Limit: 30-295 (Adult and Infant/
Child); High Limit: 35-300 (Adult and
Infant/Child)
M1132A, M1133A, M1191A, M1191B,
M1191AL, M1191BL, M1192A.
M1131A, M1191T, M1192T, M1194A,
M1195A, M1196A, M1196T, M1902B,
M1903B, M1904B.
Adult: 40-260 mmHg/5-35 kPa;
Infant/Child: 40-160 mmHg/5-21 kPa
Adult: 20-200 mmHg/2-27 kPa;
Infant/Child: 20-120 mmHg/2-16 kPa
Adult: 26-220 mmHg/3-30 kPa;
Infant/Child: 26-133 mmHg/3-18 kPa
Adult: 45-260 mmHg/6-35 kPa;
Infant/Child: 45-160 mmHg/6-22 kPa
Adult: 40-255 mmHg/5-34 kPa;
Infant/Child: 40-155 mmHg/5-21 kPa
Adult: 25-200 mmHg/3-27 kPa;
Infant/Child: 25-120 mmHg/3-16 kPa
Adult: 20-195 mmHg/2-26 kPa;
Infant/Child: 20-115 mmHg/2-15 kPa
Mean Alarm
Range High:
Mean Alarm
Range Low:
Pulse Rate
Range:
Initial
Pressure:
Maximum
Pressure:
Overpressure
Safety Limits:
Cuff Inflation
Time:
Pressure
Transducer
Accuracy:
Auto Mode
Repetition
Time:
Maximum
Measurement
Time:
Interconnect
Tube Length:
Recommended
Frequency of
Pressure
Transducer
Calibration:
Display
Size:
Type:
Resolution:
Sweep Speed:
Wave Viewing
Time:
Adult: 30-220 mmHg/4-30 kPa;
Infant/Child: 30-135 mmHg/4-18 kPa
Adult: 25-215 mmHg/3-29 kPa;
Infant/Child: 25-130 mmHg/3-17 kPa
30 to 220 bpm
160 mmHg, 21 kPa (adult); 120 mmHg,
16 kPa (Infant/Child);
300 mmHg; 40 kPa
300 mmHg; 40 kPa
75 sec maximum
+3 mmHg over the range 1-300 mmHg/
.1-40 kPa
1, 2.5, 5, 10, 15, 30, 60 or 120 min
120 sec
M1598B Connect tubing 1.5 m (4.92 ft.);
M1599B Connect tubing 3.0 m (9.24 ft.)
Yearly
Approximately 6.5 in (16.5 cm) diagonal
viewing area.
Color TFT LCD
640 x 480 pixels (VGA) with 32
brightness levels per color
20 mm/s nominal (stationary trace;
sweeping erase bar) for ECG and SpO,
5.2 sec
Environmental
Temperature:
Humidity:
Atmospheric
Pressure
Range:
Shock -
operating:
Shock -
non-operating
Water/Solids
Ingress
Resistance:
EMC:
Safety:
Mode of
Operation:
AC Line
Powered:
Battery
Powered:
Other
considerations:
0°C to 45°C (32°F to 113°F) operating;
20°C to 70°C (-4°F to 158°F) storage
Up to 95% relative humidity
Operating and Storage - 1014 mbar to
572 mbar (0 to 15,000 ft; 0 to 4,500 m)
Half-sine waveform, duration <11ms,
acceleration > 15.3 G, 3 shocks per face
Trapezoidal waveform, acceleration
30G, velocity change 7.42 m/s +10% 1
shock per face
Meets Ingress Protection level IP21
Complies with the requirements of
standard EN 60601-1-2:2002
Meets UL 60601-1 (1st edition), EN
60601-2-4:2003, EN 60601-1:1990
Continuous
100-240 VAC, 50 or 60 Hz, 1-0.46A,
Class | Equipment
Minimum 14.4 V, Rechargeable Lithium
lon
The HeartStart XL+ is not suitable for
use in the presence of concentrated
oxygen or a flammable anesthetic
mixture with air, oxygen. or nitrous
oxide. Hazards arising from software
errors were minimized by the product's
compliance with the software
requirements contained in EN 60601-1-
4:1996.
Environmental
Vibration
Operating
Random:
Vibration
Non-
Operating
Swept Sine
Vibration
Non-
Operating
Random:
Frequency: 10-100, PSD (m/s2)2/Hz 1.0
Frequency: 100-200, Slope (dB/octave):
-3.0
Frequency: 200-2000, PSD 0.5
Frequency: 10-57, Amplitude + .15 mm;
Frequency: 57-150, Amplitude 2g
Frequency: 5-500 Hz, PSD: 0.0117 g2/Hz,
Acceleration: ~ 2.41 Grms
Patient Data Storage
Internal Event
Summary:
The Internal Event Summary stores up
to 8 hours of data per event. The
number of events that can be stored is
related to the duration of each individual
event.
USB Device
Correct Drive: Use the USB Drive that accompanied
your device. If it is not available, use a
USB device that is USB 2.0 compatible
and is < 32 gigabytes.
Philips Healthcare is part of
Royal Philips Electronics
How to reach us
www.philips.com/healthcare
[email protected]
Asia
+49 7031 463 2254
Europe, Middle East, and Africa
+49 7031 463 2254
Latin America
+55 11 2125 0744
North America
+1 425 487 7000
800 285 5585 (toll free, US only)
Trademarks and trade names
are those of their respective
owners.
Please visit www.philips.com/heartstart